Shop in your local currency and language

Choose another country or region to see specific content for your location and shop online.

United Kingdom
Viral vectors

Novel AAV9 affinity resin for high-performance purification

Oct 13, 2026

Adeno-associated virus serotype 9 (AAV9) is widely utilized in gene therapy due to its broad tissue tropism and strong performance in systemic delivery, particularly for neuromuscular and central nervous system applications. As AAV9-based therapeutics continue to advance, there is an increasing need for manufacturing solutions that enhance process efficiency, scalability, and product quality.

Affinity capture represents a critical step in downstream purification, and the development of AAV9-specific resins offers significant opportunities to streamline process workflows. In this application note we describe the performance of our AAV9 affinity rapid resin, a next-generation affinity chromatography resin specifically engineered for selective capture of recombinant AAV9 (rAAV9) capsids. The resin incorporates a rationally designed ligand with high specificity toward AAV9 with high binding capacity, enabling robust binding and elution under mild processing conditions. It supports direct capture from clarified or concentrated, buffer-exchanged feeds, facilitated by short residence times and low bed heights and is compatible with different loading levels.

When evaluated with typical upstream process materials, the resin demonstrated:

  • High selectivity for rAAV9.
  • Effective removal of host cell impurities.
  • Consistently high product yield across a range of load levels.

In contrast with our Capto™ AVB resin, which lacks AAV9 binding and requires milder cleaning conditions, this resin exhibits enhanced robustness and high alkaline stability even at 0.5 M NaOH, enabling rigorous cleaning-in-place (CIP) and multiple reuse cycles.

Additionally, our AAV9 affinity rapid resin offers:

  • Favorable pressure–flow characteristics.
  • Excellent packing performance.
  • Flexible scalability across different column configurations.

Collectively, these features provide a scalable, process efficient solution for purification of AAV9.

AAV9 affinity rapid resin is a valuable addition to the AAV purification toolbox, supporting the growing demand for AAV9-based gene therapies.

Introduction

Viral vectors have become essential tools for targeted gene delivery and cell-type–specific modification (1). Among the various viral platforms explored for cell and gene therapy, recombinant adeno-associated virus (rAAV) has emerged as the most widely used vector due to its favorable safety profile and broad applicability (2).

Following cell harvest requiring cell lysis, the filtration and purification of AAV vectors become critical downstream steps, as crude harvest lysates contain a complex mixture of host cell proteins, DNA, and process-related impurities (3).

Direct affinity capture from clarified harvest in a single step offers the possibility to skip the first TFF step. This requires high selectivity of the affinity ligand and at the same time ability to process large non-concentrated sample feeds within a typical day shift (4).

Such an approach simplifies the purification workflow, reduces process time, and improves overall yield and consistency.


Results: Resin purification performance


Dynamic binding capacity (DBC)

Three independent lots of AAV9 affinity rapid resin were manufactured and we packed them into 1 mL columns. The dynamic binding capacity (DBC) for purified AAV9 was then evaluated at a short residence time of only 1 min. All three lots showed consistent DBC performance, with values of approximately 8 × 1014 viral particles (VP) per mL of resin (Fig 1).

DBC for pure AAV9

Fig 1. Dynamic binding capacity (Qb10%) for pure AAV9 measured on three independently manufactured lots of AAV9 affinity rapid resin.


Elution yield and product purity for high and low load levels of clarified AAV9 feed

We assessed the consistency of elution yield and product purity across three independently manufactured lots of AAV9 affinity rapid resin at two levels of column utilization, corresponding to 5% and 80% of the DBC at 10% breakthrough (Qb10%).

Each resin lot was packed into a Tricorn™ 5/100 chromatography column (2 mL column; 0.5 cm inner diameter, 10 cm bed height). AAV9 material was generated using the ELEVECTA™ transient cell line with triple-plasmid transfection. After harvest, cells were lysed in a buffer containing Tween 20 and treated with a DNA nuclease to degrade the DNA. The lysate was clarified using a STAX™ capsule containing PDK11 media-grade depth filter. The resulting clarified AAV9 feed was stored at −80°C until further processing.

Before chromatography, the frozen clarified feed was thawed and filtered through a Supor™ EAV Mini Kleenpak™ capsule. The AAV9 titer in the clarified feed was 6.8 × 1011 VP/mL.

For the 80% Qb10% condition, 1870 mL of clarified AAV9 cell lysate was loaded onto each chromatography column.

In a separate experiment, 117 mL of clarified AAV9 feed was loaded, corresponding to 5% of the resin DBC (Qb10%).

Both loading conditions were evaluated across all three resin lots.

The purification protocol used for the clarified AAV9 cell lysate is summarized in Table 1.

Table 1. Purification protocol

Parameter Settings
Columns with three different AAV9 affinity rapid resin lots Tricorn 5/100, 2 mL, 10 cm bed-height
Flow rate A flow rate of 2 mL/min (one min residence time) was applied for all steps with the exception of Step 6 (cleaning), which was performed at 0.4 mL/min.
Chromatography system ÄKTA pure™ 25
UV detection 280 nm, 10 mm path length
Equilibration buffer 20 mM Tris + 200 mM NaCl pH 7.8
Wash buffer 20 mM Tris + 200 mM NaCl pH 7.8
Elution buffer 100 mM Na-citrate pH 3.0
CIP solution 0.5 M NaOH

Chromatography

Step 1 Equilibration

Step 2 Sample loading





Step 3 Wash

Step 4 Elution

Step 5 Neutralization

Step 6 Cleaning

Step 7 Re-equilibration



5 CV Equilibration buffer

1870 mL (6.8 × 1011 VP/mL) AAV9 clarified feed was loaded (80% of Qb10%) into three columns with different resin lots.

117 mL (6.8 × 1011 VP/mL) AAV9 clarified feed, was loaded (5% of Qb10%) into three columns with different resin lots.

10 CV Wash buffer

10 CV Elution buffer, collection of 1 mL fractions.

5 CV Equilibration buffer

CIP with 0.5 M NaOH for 3 CV and 15 min.

5 CV Equilibration buffer


The elution pool was neutralized to pH 7 with 1 M Tris base pH 9.

Figure 2 and 3 show overlays of chromatograms for 80% and 5% Qb10% loading, respectively, across three resin lots. Only part of the loading phase is shown, and the elution peaks are highlighted. The AAV9 particles are eluted in < 3 CV of elution buffer. Note that the peak height (mAU) between high and low loads (80% and 5% of Qb10% in figures 2 and 3) cannot be directly compared since UV signals (10 mm pathlength UV cell) become saturated and are not linear above 1500 mAU (approximately).

The results show that the three independent manufacturing lots of our AAV9 affinity rapid resin did not differ significantly in chromatographic performance, yield and purity.

Elution peaks (280nm) - 3 lots

Fig 2. Overlay of 280 nm signals for the elution peaks run with three different lots of AAV9 affinity rapid resin in 2 mL columns where the clarified AAV9 feed load was 80% of Qb10% (see Table 1). Arrows represent last phase of loading AAV9 feed, washing and elution, respectively.

A

Overlaid 280nm chromatograms - Full

B

Overlaid 280nm chromatograms - Focus

Fig 3. Overlaid 280nm chromatograms from 2 mL columns packed with three different lots of AAV9 affinity rapid resin, loaded with clarified AAV9 feed corresponding to 5% of Qb10% (see Table 1). The full chromatogram (A) with arrows indicating the loading, wash, elution, neutralization, CIP, and re-equilibration phases. A magnified view of the elution phase is shown (B) to enable a clearer comparison of elution peak profiles across the three resin lots.

We assessed AAV9 VP yield using Progen AAV9 ELISA kit for both the high-load (80% Qb10%) and low-load (5% Qb10%) conditions, and the results are summarized in Table 2. VP yields remained above 80% even at low load, demonstrating that this affinity capture step performs well across a broad range of column load levels. This flexibility may reduce the need to wait for clarified feed titer results before loading. They may also enable the use of oversized columns that can be operated at higher flow rates to substantially shorten loading time.

Table 2. AAV9 eluate yields for high-load (80% Qb10%) and low-load (5% Qb10%) clarified AAV9 feed processed on 2 mL AAV9 affinity rapid resin columns packed with three different resin lots

Experiment Volume clarified feed loaded, 6.8 × 1011 VP/mL (mL) Total VP loaded Total VP in eluate Eluate VP yield (%) CV*
(%)
Lot 1 high load 1870 1.35 × 1015 1.23 × 1015 91 1.3
Lot 2 high load 1870 1.35 × 1015 1.23 × 1015 91 1.1
Lot 3 high load 1870 1.35 × 1015 1.21 × 1015 89 4.9
Lot 1 low load 117 8.43 × 1013 7.39 × 1013 88 2.9
Lot 2 low load 117 8.43 × 1013 7.05 × 1013 84 2.9
Lot 3 low load 117 8.43 × 1013 7.03 × 1013 83 0.8

*CV for duplicate analysis

Figure 4 shows the purity of eluates from the high and low load runs on the three different lots as assessed by SDS-PAGE. Elution fractions were pre-labeled with fluorescence Cy5 (Amersham™ QuickStain™ kit, LOD 50 pg) before electrophoretic separation. To improve detection of low-abundance impurities, we used Amersham™ ImageQuant™ 800 system to scan the gel on high-intensity settings. The results indicate very high eluate purity, with AAV capsid proteins VP1, VP2, and VP3 clearly visible and no detectable protein impurity bands (at high contrast), at both high and low AAV9 affinity rapid resin load conditions.

SDS-PAGE analysis of eluate samples

Fig 4. SDS-PAGE analysis of Cy5-prelabeled eluate samples (LOD 50 pg) obtained from high- and low-load clarified AAV9 feed runs using three different lots of AAV9 affinity rapid resin. Equal sample volumes were loaded in each lane.

Table 3 summarizes host cell protein (HCP) levels, DNA reduction, and ligand leakage. HCP was measured using the Gyrolab Bioaffy 4000 HC Assay Toolbox and the Gyrolab HEK293 HCP Type CL reagent kit from Gyros Protein Technologies. HCP concentrations were higher at 80% Qb10% loading than at 5% Qb10%, reflecting the greater HCP challenge applied to the column. However, the amount of HCP per viral particle was generally lower at the higher load, and a similar trend was observed for ligand leakage. DNA levels in the eluates, measured using the Quant-iT PicoGreen dsDNA Assay Kit, were below the assay quantification limit, indicating efficient removal. Ligand leakage was quantified by ELISA using MonoRab VHH antibodies from GenScript. The leaked ligand was < 11 ng/mL in the eluates. These already low levels of impurities and leached ligand are expected to be further reduced by the downstream steps that follow after affinity capture; polishing, TFF2, and sterile filtration.

Table 3. Eluate concentrations and levels per 1 × 1014 VP of host cell protein, total DNA, and leached ligand after processing clarified feed at different load levels on three lots of AAV9 affinity rapid resin

Experiment HCP conc
(ng/mL)
HCP
(ng/ 1 × 1014 VP)
Log DNA reduction Ligand content
(ng/mL)
Leaked ligand
(ng ligand / 1 × 1014 VP)
Lot 1 80% of Qb10% 227 129 > 3 8.5 4.8
Lot 2 80% of Qb10% 208 121 > 3 6.7 3.8
Lot 3 80% of Qb10% 154 88 > 3 10.9 6.3
Lot 1 5% of Qb10% 12 118 > 5 3.7 34.9
Lot 2 5% of Qb10% 32 300 > 5 3.0 29.8
Lot 3 5% of Qb10% 34 338 > 5 2.5 25.2


Elution yield across high and low load levels using concentrated, buffer-exchanged (TFF) AAV9 feed

The same purification protocol was applied to concentrated, buffer-exchanged feed using a 2 mL AAV9 affinity rapid resin column at a residence time of one min. The clarified feed was processed by tangential flow filtration (TFF), resulting in a 10-fold concentration and 5-fold diafiltration into 20 mM Tris, 200 mM NaCl, pH 7.8. A hollow-fiber cartridge (UFP-300-C-6A, 300 NMWCO) was used with the ÄKTA Flux™ 6 filtration system. The final VP titer of the TFF-treated feed was 5.7 × 1012 VP/mL.

For the high-load condition (80% of Qb10%), 161 mL of TFF-treated feed was loaded onto the 2 mL AAV9 affinity rapid resin column, whereas 10 mL was loaded for the low-load condition (5% of Qb10%). Figures 5A and 5B show the corresponding chromatograms obtained with TFF-treated feed.

A

UV 280 nm chromatograms - 80%

B

UV 280 nm chromatograms - 5%

Fig 5. UV 280 nm chromatograms obtained using TFF-treated AAV9 feed loaded at 80% (A) and 5% (B) of Qb10%, respectively. Arrows indicate the loading, wash, elution, neutralization, CIP, and re-equilibration phases.

Table 4. AAV9 eluate yields for TFF-treated feed loaded at high (80% Qb10%) and low (5% Qb10%) levels on a 2 mL AAV9 affinity rapid resin column (see run protocol in Table 1)

Experiment Elution yield AAV9
(VP %)
CV*
(%)
High load with TFF feed 104 14
Low load with TFF feed 87 2.7

*CV for duplicate analysis

Figure 6 shows fluorescence SDS PAGE purity analysis of the elution peak from a run with TFF. High intensity was used on the Amersham ImageQuant™ 800 system, and high contrast to enhance detection and visibility of any low-abundant protein impurities.

SDS PAGE analysis 80% & 5%

Fig 6. SDS PAGE analysis with fluorescence Cy5 prelabeling of eluate samples from high and low load levels of AAV9 TFF feed. The same volume from the different eluate samples was applied to each lane.


Caustic stability enables resin reuse with maintained performance

Our AAV9 affinity rapid resin is designed for cleaning and regeneration with 0.5 M sodium hydroxide, which also reduces the risk of microbial contamination.

In this study, we used the clarified AAV9 feed, as previously described.

In a 1 mL AAV9 affinity rapid resin column, 1 mL was subjected to 21 purification cycles with clarified AAV9 feed, including 20 CIP cycles. The chromatographic parameters used in each cycle are summarized in Table 5.

Table 5. Protocol for the 0.5 M NaOH CIP cycling study

Parameter Details
Column AAV9 affinity rapid resin column, 1 mL
Flow rate A flow rate of 1 mL/min (one min residence time) was used for all steps except Step 6 (cleaning), which was performed at 0.2 mL/min.
Chromatography system ÄKTA Pure™ 25
UV detection 280 nm, 10 mm path length
Equilibration buffer 20 mM Tris + 200 mM NaCl pH 7.8
Wash buffer 20 mM Tris + 200 mM NaCl pH 7.8
Elution buffer 100 mM Na-citrate pH 3.0
CIP solution 0.5 M NaOH

Chromatography

Step 1 Equilibration

Step 2 Sample loading

Step 3 Wash

Step 4 Elution

Step 5 Neutralization

Step 6 Cleaning

Step 7 Re-equilibration



5 CV Equilibration buffer

380 mL (6.8 × 1011 VP/mL) clarified AAV9 feed, 40% of Qb10%.

10 CV Wash buffer

10 CV Elution buffer, collection of 1 mL fractions.

5 CV Equilibration buffer

CIP with 0.5 M NaOH for 3 CV and 15 min.

5 CV Equilibration buffer


Figure 7 shows overlaid chromatograms from the first, fifth, 10th, 15th, and 20th purification cycle performed on the same AAV9 affinity rapid resin column. The chromatograms were nearly identical across runs. For practical reasons the load (40% Qb10%) was approximately half of a load using full capacity (80% Qb10%), to reduce duration of the full 23 run cycle study. The ligand was exposed to repeated run protocols and harsh CIP condition cycles but also exposed to a lower cell lysate load level. Despite this, there was no build up in column pressure with increasing cycling.

Chromatograms overlay - cleaning cycle

Fig 7. Overlaid UV 280 nm chromatograms from the first, fifth, 10th, and 20th purification cycle performed on the same AAV9 affinity rapid resin column using clarified AAV9 feed. Arrows indicate the loading, wash, elution, neutralization, CIP, and re-equilibration phases.

Figure 8 shows the remaining DBC at Qb10% measured with pure AAV9 after 1, 10, and 20 purification cycles. The results indicate that AAV9 affinity rapid resin retains 94% of its original binding capacity after 20 purification cycles, with 0.5 M NaOH CIP performed between cycles.

Remaining AAV9 DBC values

Fig 8. Remaining AAV9 DBC values at Qb10% for AAV9 affinity rapid resin column (1 mL) compared to first run after 10 and 20 purification run cycles with clarified AAV9 feed.

AAV9 VP recovery in the flowthrough and eluate fractions from cycles 1, 10, and 20 were determined using a Progen AAV9 ELISA kit. In all cases, the flowthrough contained low levels of viral particles, whereas the eluate showed an AAV9 yield of approximately 100%.

Table 6. AAV9 VP elution yields at different purification run cycles

  Cycle 1 Cycle 10 Cycle 20
Sample Start        Flow-through Eluate Start        Flow-through Eluate Start        Flow-through Eluate
Volume (mL) 380 380 4 380 380 4 380 380 4
Concentration (VP/mL) 7.3 × 1011 1.1 × 109 6.4 × 1013 6.8 × 1011 6.6 × 108 5.8 × 1013 7.1 × 1011 7.3 × 108 6.0 × 1013
Amount VP 2.8 × 1014 2.1 × 1011 3.0 × 1014 2.6 × 1014 2.5 × 1011 2.7 × 1014 2.7 × 1014 2.8 × 1011 2.8 ×1014
CV between duplicates (%) 4.9 1.3 6.6 4.7 5.6 4.0 4.2 1.3 3.5
VP yield (%) - 0.2 107 - 0.1 106 - 0.1 104

Ligand leakage in the eluates was 3.7 ng/mL after the first run and decreased further to 0.5 ng/mL in the eluate after 10 and 20 run cycles with clarified AAV9 feed (Fig 9). Leached ligand was quantified by ELISA using GenScript MonoRab™ Rabbit Anti-Humanized antibody (#A02165-200) for capture and HRP-conjugated MonoRab™ Rabbit Anti-Camelid VHH Cocktail (#A02016-200) for detection. The ligand is available under a non-transferable limited license for analytical use in ligand leakage qualification; contact us for further information.

As shown in Figure 9, ligand levels in the AAV9 eluate remained very low throughout the study and declined progressively over successive cycles.

Levels of leached ligand

Fig 9. Levels of leached ligand in the eluate from different purification cycles analyzed by an ELISA assay.


Easy process development and scale up for GMP production

The resin is readily scalable from milliliter-scale columns to multi-liter formats, supporting bioreactor volumes from a few liters up to 2000 L. Once the dynamic binding capacity of AAV9 affinity rapid resin has been established for AAV9 capsids, our recommended run protocol gives high performance. As such, further optimization of wash and elution buffers or additional high-salt wash buffers or acidic strip steps for impurity removal (before the CIP step) are typically not needed.

A single buffer system can be applied for both column equilibration and washing prior to elution. Elution is performed under mild acidic conditions (pH 3.0). This streamlined approach simplifies process development and facilitates scale-up, minimizing the number of critical parameters that must be controlled during scale-up.

In this study, scale-up performance was evaluated by comparing a 2 mL column with a 150 mL column using TFF-processed AAV9 feed. AAV9 material was produced at the 200 L scale using ELEVECTA™ transient cell line with triple-plasmid transfection in a Xcellerex™ XDR-200 bioreactor. After harvest, cells were lysed in a buffer containing Tween 20 and treated with a DNA nuclease to degrade nucleic acid. The lysate was clarified using four 1 m2 STAX™ capsules containing PDK11 media-grade depth filter and Supor EKV sterilizing grade filter. The clarified AAV9-containing lysate was then processed by TFF using the same hollow-fiber cartridge described above. The TFF-treated AAV9 feed was stored at −80°C before chromatography and thawed prior to use.

AAV9 affinity rapid resin was packed into a Tricorn 5/100 chromatography column with a bed height of 10.1 cm and a bed volume of 2.0 mL. A larger HiScale™ 50 chromatography column (50 mm inner diameter) was also packed, with a bed height of 7.8 cm and a bed volume of 150 mL.

To assess scale-up performance, purification on the 2 mL and 150 mL columns was compared using proportionally scaled process parameters, as summarized in Table 7. The 2 mL column was run using a separate small-scale TFF batch prepared from the same cell line and with the same clarification procedure as the material produced at 200 L (used for the 150 mL column). However, the impurity profiles of the two TFF feeds differed; most notably, the large-scale feed contained approximately threefold higher host cell protein levels per mL than the small-scale material. We introduced minor process adjustments, with the main difference being that the 2 mL column used a 10 CV wash before elution, whereas the large-scale run used a 20 CV wash. Immediately after chromatography we neutralized the AAV9 eluates to pH 5.0 with 1 M Tris to pH 9.

Table 7. Run protocols for the purification of TFF AAV9 feed using 2 mL column and 150 mL column loaded at ~ 80% of QB10%

Parameter 2 mL column 150 mL column
System ÄKTA pure 25 ÄKTA Pilot 600
UV detection 280 nm, 10 mm path length 280 nm, 2 mm path length
Column volume 2.0 mL 150 mL
Residence time (min) 1 1
Flow rate (mL/min) 2 150
Equilibration buffer and wash buffer 20 mM Tris + 200 mM NaCl pH 7.8 20 mM Tris + 200 mM NaCl pH 7.8
Elution buffer 100 mM Na-citrate pH 3.0 100 mM Na-citrate pH 3.0
CIP buffer 0.5 M NaOH 0.5 M NaOH
TFF AAV9-(GFP) feed AAV9 titer 5.4 × 1012 VP/mL
HCP 16761 ng/mL
AAV9 titer 8.6 × 1012 VP/mL
HCP 46030 ng/mL
Load 80% of QB10% 161 mL 8400 mL
VP load per mL resin 5.39 × 1012 × 161 / 2.0 =
4.3 × 1014 VP/mL resin load
8.6 × 1012 × 8400 / 150 =
4.8 × 1014 VP/mL resin load

Chromatography

Step 1 Equilibration

Step 2 Sample loading


Step 3 Wash

Step 4 Elution


Step 5 Neutralization

Step 6 Cleaning


Step 7 Re-equilibration



5 CV Equilibration buffer

161 mL (5.4 × 1012 VP/mL) AAV9 TFF feed, (80% of Qb10%).

10 CV Wash buffer

10 CV Elution buffer, fractionation in 1 mL fractions.

5 CV Equilibration buffer

CIP with 0.5 M NaOH for 3 CV and 15 minutes. 0.4 mL/min.

5 CV Equilibration buffer

5 CV Equilibration buffer

8400 mL (8.6 × 1012 VP/mL) AAV9 TFF feed, (80% of Qb10%).

20 CV Wash buffer

10 CV Elution buffer, peak fractionation


5 CV Equilibration buffer

CIP with 0.5 M NaOH for 3 CV and 15 min. 30 mL/min.

5 CV Equilibration buffer


Figure 10 highlights the chromatograms from the purification runs on the 2 and 150 mL columns packed with AAV9 affinity rapid resin, respectively. The column pressure for the 150 mL column during feed loading at 1 min residence time did not exceed 0.25 bar.

A

Chromatograms - Purification - 2mL

B

Chromatograms - Purification - 150mL

Fig 10. Chromatograms from AAV9 TFF feed purification runs using a 2 mL (A) or a 150 mL (B) AAV9 affinity rapid resin columns loaded at 80% of Qb10%. Arrows indicate the loading, wash, elution, neutralization, CIP, and re-equilibration phases.

Both scales showed high AAV9 viral particle elution yields, as determined using the Progen AAV9 ELISA kit (Table 8). The difference in elution yield between the two scales could be due to variation in ELISA assay and small differences in peak collection.

At both scales, AAV9 detected in the flowthrough accounted for only 0.1% of the total loaded viral particles.

Table 8. AAV9 yield for the 2 mL column and 150 mL column after loading TFF AAV9 feed at 80% of QB10%

  2 mL column 150 mL column
Sample Start Flowthrough Eluate Start Flowthrough Eluate
Volume (mL) 116 116 5 8400 8400 245
Concentration (VP/mL) 4.8 × 1012 3.8 × 109 1.4 × 1014 8.6 × 1012 1.3 × 1010 2.6 ×1014
Amount VP 7.8 × 1014 6.1 × 1011 8.1 × 1014 7.2 × 1016 1.0 × 1014 6.4 × 1016
Yield (% VP) - 0.08 104 - 0.14 89

Eluate purity for the 2 mL and 150 mL column purifications was assessed by SDS-PAGE with Cy5 fluorescent prelabeling (Fig 11). VP band intensity was higher in the large-scale purification because the same sample volume was loaded in each lane and the large-scale elution pool was narrower and more concentrated than the small-scale pool. However, both eluates showed strong VP1, VP2, and VP3 bands with no visible protein impurity bands, indicating high purity at both scales.

SDS PAGE analysis - 2mL & 150mL

Fig 11. SDS-PAGE analysis of Cy5-prelabeled eluates from the 2 mL and 150 mL column purifications with AAV9 TFF feed.

HCP reduction was also compared between the small- and large-scale purifications. As shown in Table 9, the resin demonstrated comparable HCP clearance at both scales.

Table 9. TFF AAV9 feed details and HCP removal comparison between AAV9 affinity rapid resin 2 mL and 150 mL column purifications

  2 mL column 150 mL column
Volume feed loaded (mL) 161 8400
Total AAV9 VP (at start) 7.8 × 1014 7.2 × 1016
Total AAV9 VP (in eluate) 8.1 × 1014 6.4 × 1016
HCP concentration (in feed) 16800 ng/mL 46000 ng/mL
Total HCP loaded 2 700 000 ng 386 000 000 ng
HCP concentration (in eluate) 257 ng/mL 329 ng/mL
Total HCP (in eluate) 1285 ng 80 600 ng
ng HCP per 1 × 1014 VP (in feed) 2 700 000 / 7.8 = 346 000 386 000 000 / 720 = 536 000
ng HCP per 1 × 1014 VP (in eluate) 1285 / 8.1 = 158 80 600 / 640 = 125
Log reduction HCP* Log (346 000 / 158) = 3.3 Log (536 000 / 125) = 3.6

*Based on ng HCP per 1 × 1014 viral particles



Discussion

The results presented in this report demonstrate that our AAV9 affinity rapid resin provides a robust and efficient capture step for AAV9 purification. A key finding was the high dynamic binding capacity achieved at a short residence time of one min, an important parameter for manufacturing settings where productivity and throughput are critical. The ability to bind AAV9 efficiently without prior concentration of clarified feed simplifies the downstream process and may reduce the need for additional unit operations.

Importantly we observed consistency across independently manufactured resin lots. This was seen as similar chromatographic profiles, comparable elution behavior, and closely aligned viral particle yields, indicating that the resin manufacturing process is reproducible and that the product performs reliably from lot to lot. This is a critical attribute for process development and manufacturing transfer, where lot-to-lot variability can introduce risk to process consistency and product quality.

The purification results further showed that the resin combined high yield with efficient impurity removal for both clarified and TFF feeds. Across both low and high DBC utilization of the resin, AAV9 recovery remained high and SDS-PAGE analysis indicated highly pure eluates. This supports flexibility in load levels, reducing the need to await analysis results before loading and enables the option to use oversized columns to dramatically shorten loading time. Together, these findings indicate that the resin can support a simplified capture strategy with high productivity while still delivering product streams with favorable purity characteristics.

The cycling study also demonstrated that the resin tolerates repeated cleaning with 0.5 M NaOH while maintaining chromatographic performance, binding capacity, yield and purity. This is particularly relevant for process development but also for manufacturing applications, where resin reuse contributes to process economy and operational robustness. The low and decreasing ligand leakage over repeated cycles further supports the stability of the ligand–matrix construct under the applied cleaning conditions.

With regard to scale-up our study showed that the purification concept transferred well from small to larger column formats. Despite differences in feed composition, including higher host cell protein content in the large-scale TFF material, both scales achieved high AAV9 yields and similar impurity clearance per viral particle. These data suggest that the resin and associated process conditions are sufficiently robust to support scale-up with limited re-optimization, which is an important advantage in AAV process development.

Conclusion

Cytiva AAV9 affinity rapid resin delivers a robust and scalable capture solution, combining high productivity, strong recovery, operational flexibility, and reproducible purification performance across relevant process conditions.

  • Binding capacity > 5 × 1014 VP/mL resin at one min residence time.
  • No clarified feed concentration required before chromatography.
  • Elution yields > 85%, with highly pure eluates.
  • Consistent performance across load levels, resin lots, feed types, and scales.
  • Efficient host cell impurity and DNA reduction, with DNA below assay quantification limit.
  • Stable after ≥ 20 CIP cycles with 0.5 M NaOH for 15 min.

Finally, comparable purification performance at small and larger scale indicates that the platform is well suited for straightforward process development and scale-up for AAV9 manufacturing.


References

  1. De Haan P, Van Diemen FR, Toscano, MG. Viral gene delivery vectors: the next generation medicines for immune-related diseases. Hum Vaccin Immunother, 2021; 17(1), 14–21. https://doi.org/10.1080/21645515.2020.17579892.
  2. Wang JH, Gessler DJ, Zhan W, et al. Adeno-associated virus as a delivery vector for gene therapy of human diseases. Sig Transduct Target Ther 2024; 9, 78. https://doi.org/10.1038/s41392-024-01780-w
  3. Wang D, Tai PWL, Gao G. Adeno-associated virus vector as a platform for gene therapy delivery. Nat Rev Drug Discov, 2019; 18, 358–378. https://doi.org/10.1038/s41573-019-0012-9
  4. Mietzsch M, Smith J, Yu J, et al. Characterization of AAV-Specific Affinity Ligands: Consequences for Vector Purification and Development Strategies. Mol. Ther. - Methods Clin. Dev. 2020; 19, 362-373


Ordering information

Product Size Product code
AAV9 affinity rapid resin 25 mL 17542601
AAV9 affinity rapid resin 100 mL 17542602
AAV9 affinity rapid resin 1 L 17542603
AAV9 affinity rapid resin column, 1 mL 5 × 1 mL 17542652
AAV9 affinity rapid resin column, 4.7 mL 4.7 mL 17542615

CY61075

Test the AAV9 affinity rapid resin
Connect with our team to organize a trial of our AAV9 affinity rapid resin.

Did you find what you were looking for?