Bioprocess filtration

Determination of product wet integrity test limits for filter cartridges

Oct 8, 2026

Purpose

This report describes the procedure for determining integrity test limits for filter cartridges when wetted with process liquids, as performed globally by our validation services group.

Introduction

In many critical liquid processes involving filtration, there is a requirement to achieve the highest possible assurance of filter integrity and removal efficiency. In such applications, installation of integrity testable filters and performance of routine integrity tests by the user is essential to demonstrate that the system is performing to specification. Regulatory requirements also highlight the requirement for integrity testing (1,2).

Under standard conditions, the integrity test is performed with a standard fluid such as water or an alcohol solution, for wetting hydrophilic or hydrophobic filters respectively. In some applications, mixing of the process liquid with a standard wetting fluid must be avoided; product residues or interaction between the liquids may prevent a complete and stable wetting of the filter membrane. In other cases, flushing the product out of the filter can be time consuming and expensive, especially where large volumes of flush fluids such as alcohols are required. In these cases, an integrity test using the process liquid as the wetting agent can overcome these problems.

To determine the appropriate test limits and to assure a proper correlation to the filter validation, it is necessary to understand the influence of the wetting fluid on the forward flow characteristics of the filter system during the integrity test.

Forward flow correlation with bacterial challenge

Filter validation requires challenge testing using a procedure sensitive enough to detect the passage of contaminants of interest. For example, challenge tests of 0.2 μm rated filter elements are performed with Brevundimonas diminuta ATCC 19146, a bacterium which under controlled conditions measures 0.3 to 0.4 μm in diameter and 0.6 to 1.0 μm in length. This is a destructive test, and cannot be performed in a production environment. Therefore, a correlation is made to a non-destructive integrity test in the laboratory, for instance, the forward flow test method (3). From that correlation, filter performance can be verified using the validated forward flow or other integrity test method.

These non-destructive integrity tests are performed by applying regulated air or nitrogen pressure to a completely wetted filter and observing gas flow. The forward flow test quantitatively measures both the diffusive flow of gas through all the pores of a wetted filter, and flow through any open pores, while a “bubble point” type test is qualitative and depends upon visual interpretation or instrument (4) interpretation (algorithm) of gas bubbling through the largest pores of a wetted filter.

We developed the forward flow test for quantitative testing of the integrity of critical filters. The bubble point test was developed as the first nondestructive test for filter integrity. While the bubble point test can be used for both filter cartridges and discs, we recommend the forward flow test for cartridges, as interference from background diffusive flow can affect the bubble point test results. For disc filters, where the forward flow is too small to accurately measure, the bubble point test can be used.

Validation documentation has been published for all sterilizing grade filter cartridges (5-8) as typical examples. For example, Table 1 shows data from publication USTR1029 (‘Validation guide for 0.2 μm Nylon 66 Membrane Cartridges’) which demonstrates the correlation of liquid bacterial challenge data to a forward flow integrity test value (water wet, air as the pressure source) for the 0.2 μm sterilizing grade Ultipor N66 filter cartridge. This filter membrane has been validated as sterilizing grade by challenge testing to blockage with Brevundimonas diminuta ATCC19146.

Therefore, based on the forward flow correlation to bacterial retention, we validated the forward flow limit to be retentive in the reference fluid. Since filter elements with forward flow higher than the set limit are still shown to be retentive, there is a safety factor established for each filter type. In this example, the maximum allowable forward flow limit for the cartridge AB1NRP is 8 mL/min (test pressure 2760 mbar gauge, or 40 psig), and the first non-sterile effluent was observed for a filter cartridge with a forward flow value of 17.6 mL/min, a safety factor in excess of 100%. The amount of safety margin will vary depending on the filter type.

Table 1. Comparison of forward flow value with retention of B. diminuta ATCC19146 for Ultipor N66 (0.2 μm rated) sterilizing grade filters

Filter cartridge serial number Forward flow water wet 2.76 barg air (mL/min) Bacterial challenge results 7.6 L/min for 6 h
Challenge 1 to 3.5 × 1012
Final dP = 4.1 barg
Sterile effluent Titer reduction
K0140208 2.3 Yes
K0140208 2.3 Yes
K0130107 3.2 Yes
K0140116 3.4 Yes
K0130056 3.8 Yes
K0140207 4.3 Yes
K0130088 4.6 Yes
K0140031 4.7 Yes
K0140018 4.8 Yes
K0130063 5.5 Yes
113AFB 5.7 Yes
K0140205 6.2 Yes
K1900024 6.6 Yes
NRC4666 7.1 Yes
K0140214 7.6 Yes
K1890121 9.8 Yes
NRC4658 10.0 Yes
K1890138 11.5 Yes
K1890137 13.0 Yes
K1890123 14.3 Yes
K1890107 15.5 Yes
NRC4660 16.2 Yes
K1900002 16.3 Yes
K1900002 16.3 Yes
NRC4667 17.1 Yes
NRC4668 17.6 No 1.0 × 1012
020AFA 19.4 Yes
K1900011 20.6 Yes
K1890132 20.8 Yes
K1900046 21.0 Yes
K1890111 22.0 Yes
025AFB 23.7 Yes
K1890131 23.7 No 2.4 × 107
K1890117 24.6 Yes
K1900013 26.7 Yes
K1900037 27.1 No 9.7 × 108
K1900053 28.0 Yes
090AFD 28.6 No 8.1 × 106
NRC4665 29.7 No 1.2 × 108
111AFD 30.0 No 1.0 × 105
037AFD 31.6 No 1.1 × 106
K1900020 33.3 Yes
K1900051 33.4 Yes
NRC4656 35.7 Yes
102AFD 40.0 No 5.1 × 104
014AFD 40.0 No 1.7 × 106
014ADN 40.0 No 4.7 × 109
NRC4669 45.0 No 3.7 × 1010
078AFE 46.2 Yes
138AFE 47.5 No 2.3 × 105
K1890127 50.3 No 1.0 × 108
NRC4659 52.6 No 8.3 × 109
NRC4664 54.3 No 4.5 × 105
114AFD 54.5 No 1.6 × 103
139AFD 66.7 No 3.0 × 105
102AFE 67.0 No 7.1 × 104
K1890116 67.6 Yes
095AFE 160 No 9.1 × 103
043AFE 200 No 8.0 × 103
060AFE 203 No 9.2 × 104
043AFD 275 No 3.0 × 104
059AFE 400 No 2.0 × 104
084AFE 450 No 4.8 × 105

The fundamental principles of the forward flow integrity test

The forward flow integrity test is a quantitative, non-destructive method to check the integrity and removal rating of membrane filters. This test is based upon the following physical phenomena:

Capillary action
  • If a microporous membrane is wetted with an appropriate liquid, then the liquid is held in the pores by capillary action. As a model for the first approximation of capillary force, the pores might be assumed to be cylindrical capillaries filled with liquid. If there is a differential pressure across the membrane, the balance of forces, that is, the pressure force and the capillary force, can be written as:

Equation 1

With the capillary diameter d, the surface tension y, the wetting angle θ, and the differential pressure (p2 - p1).

Equation 1 includes a form factor (K) that takes into account the geometric differences between the capillaries and the real pores.

Bubble point
  • At pressures less than the bubble point, liquid is retained within the filter pore by capillary force.
  • Liquid is expelled from the filter pores when the pressure exceeds the bubble point (Fig 1).

Fig 1. Bubble point pressure on a wetted membrane.


  • From Equation 1, the critical differential pressure, dpcrit or (p2 - p1)crit that is necessary to force the liquid out of the capillaries can be calculated. This is the so-called bubble point pressure.
  • Equation 1 can be rearranged, and an expression is thus derived including (p2 - p1)crit:

Equation 2

This leads to the LaPlace equation, which defines the pressure at which liquid is displaced from a capillary. From Equation 2, it is shown here to include the form factor K, to take account of the geometric differences between circular capillaries and the real pores in a filter membrane.

Equation 3

Where:

P = Bubble point pressure [(p2 - p1)crit]
K = Form factor, constant for a specific filter medium
ϒ = Surface tension of the wetting liquid
θ = Wetting contact angle
d = Pore diameter

Diffusion

  • If the differential pressure does not exceed the bubble point value for all pores in the membrane, the liquid contained in a wetted filter membrane behaves as a layer whose thickness is equal to that of the filter membrane. Under the influence of an increased gas pressure applied to one side of the membrane, gas dissolves in the liquid and diffuses through it, being released at the liquid interface with air at lower pressure on the other side of the membrane (Fig 2).

Fig 2. Diffusional gas flow in the forward flow test.


Gas at higher pressure goes into solution in the wetting liquid, and a diffusion gradient is established through the liquid film in the filter membrane.

At the region of lower pressure, the concentration of gas in the wetting liquid exceeds the equilibrium for prevailing pressure; gas comes out of solution and the gas flow can be measured.

This may be approximated by a simplified version of Fick’s law:

Equation 4

In Equation 4, n is the area specific molar flow, D is the diffusion constant for the gas-liquid system, c2 and c1 are the molar gas concentrations and L is the length of the diffusional path. As the value of L cannot be measured, the thickness of the membrane is substituted. The membrane specific correction factor K takes into account the void fraction of the membrane and the real length of the diffusion path.

As the solubility of any test gas is rather low in a wetting liquid, the difference of the molar concentration (c2 - c1) can be expressed by Henry’s law:

Equation 5

with HC being the Henry’s law coefficient for this liquid-gas system at a given temperature.

To calculate molar gas flow for a filter of area AF, the molar flow Ṅ might be written as:

Equation 6

And the standard volume flow V̇S as:

Equation 7

with M̃ the molar mass and ρS the density of the test gas in the standard state.

Therefore, V̇S (standard volume flow) can be expressed by using Equation 4, 5, 6, and 7, as:

Equation 8

This standard diffusive flow which results from a specified differential pressure, and the bulk open pore flow through larger non-wetted pores can be correlated with the retentive capability of the filter membrane for microbial contamination as presented in the filter validation guides (5-8).

Forward flow spectrum

  • For a range of increasing applied pressures, the rate of gas flow (sum of diffusive and open pore flow) can be measured. Figure 3 illustrates gas flow through a wetted filter element plotted as the gas flow rate per unit pressure versus differential pressure across the membrane.

Fig 3. Illustration of a forward flow spectrum (KL curve).


This is described as a forward flow spectrum, or KL curve.

  • The portion of the curve where, at low pressures, the flow rate per unit pressure remains constant with increasing pressure is the region of predominantly diffusive flow (Fig 4A and 4B).
  • As the pressure increases, the flow rate per unit pressure is no longer constant, and there begins a transition from diffusive flow to bulk open pore flow through the membrane. During the transition between diffusive flow and bulk flow the filter will exhibit an increasing flow rate per unit pressure as, with increasing pressure, pores are emptied of wetting liquid, resulting in bulk open pore flow (Fig 4C).
  • As the pressure is further increased, the flow rate per unit pressure increases dramatically. The portion of the curve is dominated by bulk open pore flow (Fig 4D).

Fig 4. Development of a KL curve.


Note: Q/p represents the normalized flow (flow/pressure). Graphing this way allows for a more defined visual of the various portions of the curve.

Fig 5. Comparison of forward flow spectra.


To develop product specific integrity test limits, the forward flow spectra is performed using a standard wetting fluid and the customer process fluid using the same test filter. In this example, the process liquid exhibits a higher rate of diffusion and a lower KL value than the standard wetting liquid (water) (Fig 5).

Equation 8 shows the permeation V̇S (standard volume flow) or gas flow rate, is directly proportional to the differential pressure (p2 - p1) across the wetted membrane, and is dependent on both pore geometry and membrane thickness. Thus, is constant in the diffusive flow region (Equation 9), as described above and shown in Figure 5. KL is defined as the pressure at the intercept of the two straight regions of this curve. The test pressure for a forward flow test must be less than the KL value (or bubble point) of the membrane, and is set in the area of the curve where stable normalized flow is measured.

Equation 9


Determination of forward flow integrity test limits

  • To determine the integrity test limits for a process liquid, it is necessary to measure those physical properties of the liquid which influence the test pressure and the rate of diffusive flow. Basic data measured are the surface tension, to ensure that account is taken of surface-active excipients in the process liquid (e.g., preservatives) and viscosity, to ensure that appropriate stabilization times are used during the test.
  • Fundamental to the determination of forward flow integrity test limits is knowledge of the diffusive flow characteristics of the filter system. Direct comparison of diffusive flow spectra (Fig 5) defines the wetting and forward flow behavior for the filter membrane wet with the process liquid.
  • Stable wetting of the filter membrane is essential for meaningful forward flow measurements and is revealed in a KL curve as a straight, horizontal line in the region of diffusive flow (Fig 3, 4A, and 4B). A sharp transition from predominantly diffusive flow to bulk open pore flow (Fig 3 and 4C) is necessary to permit accurate determination of a sensitive test pressure for use with the process liquid. The pressure at which this transition occurs is dependent upon the surface tension of the wetting liquid.
  • By comparative analysis, the data obtained (Fig 5) permit calculation of forward flow integrity test values for the process liquid that are directly correlated to the standard reference liquid (water or an alcohol solution, as appropriate to hydrophilic or hydrophobic filters).
  • This analysis consists of two factors that are used to calculate the product specific test parameters: the test pressure factor (TPF) and the forward flow factor (FFF).
  • The KL point of the membrane is calculated using linear regression to find where the intersection of the two straight portions of the curve intersect the x axis (pressure). The test pressure factor will be either the ratio of the surface tensions (product to reference fluid) or the calculated KL points (product to reference fluid), depending on the product composition and the shape of the product wet forward flow curve.
  • The normalized flow (mL/min/pressure unit) is calculated at each applied gas pressure of the forward flow spectra by dividing the gas flow by the applied pressure. A stable forward flow region is selected.
  • A forward flow factor is then calculated by taking the ratio of the average of the normalized flows (flows per unit pressure) measured in the product fluid to the average in the reference fluid.
  • The product-wet test pressure and minimum expected bubble point are then calculated by multiplying the reference test pressure or minimum expected bubble point by the appropriate test pressure factor.
  • The product-wet forward flow limit is calculated at the product wet test pressure by multiplying the forward flow limit for the reference fluid by the forward flow factor, and multiplying by the test pressure factor.
  • The difference between the standard test limits and the test limits determined for the process liquid reflect the differences between the two liquids. They may also be affected by the use of a different test gas or temperature.

Use of a single test filter

  • Our approach to development of product-specific integrity test limits is based on the methodology presented in PDA 26 (9), but it uses a single filter cartridge, instead of multiple filter disc equivalents (as described in PDA 26). Also, our methodology allows for determination of both forward flow and bubble point (BP) parameters using the same set of test data. We have successfully used this method for more than 30 yr.
  • Although a single filter is used, this method of integrity test limit-development is based on multiple measurements. By collecting data at many test pressures, a region of stable normalized flow (flow per unit pressure) can be established for both the reference and product wet tests. As described above, these ratios are then used to calculate the product's wet forward flow limit.
  • PDA 26 Section 7.2.2 states “The diffusive/forward flow ratio depends on the diffusion constant and the solubility coefficient of the test gas in these fluids, not the fluid’s interaction with the filter. Therefore, for the diffusive/forward flow test, one filter (or more) can be used and potential variability in test results can be addressed by averaging several replicate tests in the reference fluid (e.g., water) followed by replicate tests on the product wetted filter.” This is true when the same test filter is used for both the product wet and reference forward flow curves. Considering that the forward flow ratio is independent of the fluid’s interaction with the filter, it can be concluded that one test filter is sufficient to develop product-specific forward flow test limits.
  • Therefore, our methodology using a single pleated filter, versus multiple bench scale filter disc equivalents, is justified when deriving product specific integrity test limits. This methodology allows for determination of both forward flow and bubble point (BP) parameters in the same test to meet customer requirements.

Conducting an integrity test under process conditions

  • In order to maintain system sterility, the measurement of forward flow or bubble point must be performed on the upstream side of the filter. The measurement of forward flow in presterilized process systems can be performed on the upstream side of the filter, for example by means of direct flow measurements technique used by Palltronic Flowstar instruments (10).

Requirements

All product wet integrity test limits are determined and issued by our validation services department. Global procedures are employed to ensure that consistent test results are obtained. All validation reports include a detailed procedure, test results, and analysis.
Our validation services departments establish on request integrity test limits for filter cartridges wet with process liquid, based on the following information:

  • Composition of the process liquid.
  • Material safety data sheet for the process liquid or its components, to satisfy local regulations for the control of hazardous substances.
  • Process conditions appropriate to the integrity test data:
    • Temperature at which the test will be conducted.
    • Test gas (air or nitrogen).
    • Filter cartridge and housing part numbers.
    • Additional system volume upstream of the filter assembly (pressure decay test limits).
  • A small volume of the process liquid may be required for laboratory testing (typically 2 L), depending on the nature of the liquid and the size of the filter.
  • Diffusive flow spectra for the determination of integrity test limits are generated using a pleated filter cartridge. The relatively large area of filter membrane provides greater accuracy in laboratory measurements. When making these measurements, conditions representing the actual process are maintained (e.g., temperature, pressure source).

    Summary

    Product wet integrity test limits can be provided for most product wet filters. These limits will be correlated to the reference fluid limits, which in turn, are correlated to bacterial retention as described in the validation guide for the particular filter type.

    The product wet integrity limits are calculated by performing a forward flow curve using a reference fluid, followed by a product wet forward flow curve using the same test filter. Forward flow measurements are taken at increasing pressure steps until bulk air flow occurs through the filter. The ratio of the KL points, the surface tension ratio, and the stable region of flow are all used to determine the product wet test parameters.

    When a product wet integrity test is performed using these parameters, a passing test result indicates that the filter is integral.

    References

    • EU Guidelines to Good Manufacturing Practice Medicinal Products for Human and Veterinary Use, Annex 1: Manufacture of Sterile Medicinal Products. Paragraph 113.
    • U.S. Department of Health and Human Services Food and Drug Administration. Guidance for Industry: Sterile Drug Products Produced by Aseptic Processing: Good Manufacturing Practice. September 2004.
    • Center for Drugs and Biologics and Office of Regulatory Affairs, Food and Drug Administration. Guideline on Sterile Products Produced by Aseptic Processing. Sterilizing Operations-Filtration. Rockville, MD, USA, 1987.
    • Pall Corporation. Instruction For Use: Palltronic Flowstar IV Instrument. USTR2594(2).
    • Pall Corporation. Validation Guide for Pall 0.2 μm Nylon 66 Membrane Cartridges. Scientific and Technical Report USTR1029.
    • Pall Corporation. Validation Guide for Pall NT Grade 0.1 μm Microbially rated Nylon66 Membrane Cartridges. Scientific and Technical Report USTR1389.
    • Pall Corporation. Validation Guide for Pall 0.2 μm Fluorodyne™ Membrane Cartridges. Scientific and Technical Report USTR1292.
    • Pall Corporation. Validation Guide for Pall Emflon™ II Membrane Cartridges for Air and Gas Filtration. Scientific and Technical Report USTR1132.
    • Parenteral Drug Association. Sterilizing Filtration of Liquids. PDA Technical Report No. 26. PDA J. Pharmaceutical Science and Technology. 1998;52(3 Supplement):1-31.
    • Pall Corporation. Validation Guide for the Palltronic Flowstar Filter Integrity Test Instrument. USTR2710.


    CY61798-11Sep-26-AN