Virus filtration is a critical step in biopharmaceutical manufacturing for robust viral clearance through a size-exclusion mechanism that complements other virus safety strategies. This article outlines approaches for filterability testing, protein transmission studies, and virus challenge testing using Pegasus™ SV4 virus removal filter membrane in Minidisc capsules, and provides guidance for process development, validation, and scale-up.
Introduction
Of the available viral clearance (inactivation and removal) strategies, filtration is a robust technique that generally is not susceptible to minor changes in process conditions. The size-exclusion mechanism of filtration complements any other inactivation or removal technique selected, as filtration targets a different property of a virus (physical dimensions). This allows for process designs based on orthogonal methods to achieve a high degree of virus safety. To date, virus filtration has demonstrated a high removal efficacy among virus clearance technologies and has become a well-accepted orthogonal method for the clearance of contaminant and potential contaminant viruses.
Pegasus SV4 virus removal filters can provide a robust, economical solution for clearance of small nonenveloped viruses and large enveloped viruses, offering high throughput and providing constant, stable flow rates in both dilute and complex, concentrated biological fluids without dilution. These outstanding properties provide minimal flow decay over a wide range of process conditions and extended process times. This allows a much-improved degree of process control, as well as control of process costs, to provide maximum virus filtration efficiency and economy. This robust and efficient performance helps to qualify Pegasus SV4 virus removal filters as a ‘platform technology’.
Here, we provide assistance in filter qualification, validation, and process development. We describe best practices for how to conduct filterability testing and virus spike challenges. To utilize the full performance of Pegasus SV4 virus filters, and to maximize process performance, we recommend working inside the design space recommended here, which is based on experience from a number of case studies carried out with a variety of biopharmaceutical manufacturers, and our internal application studies.
This article is separated into three main sections:
1. Filterability and protein transmission trials
Filterability and protein transmission trials are often conducted initially to determine the flow, capacity and protein passage through a virus filter with a nutrient additive or product intermediate feed stream. The results show the potential throughput that a virus filter can provide during processing and can aid in projection of scale-up sizing and process cost modelling.
2. Virus spike challenge testing
Viral clearance validation is required to confirm that suitable virus retention can be achieved in process fluid under desired process conditions.
3. Key virus filtration factors
A good understanding of the critical parameters involved in virus filtration and the typical Pegasus SV4 filter design space is required to ensure successful optimization and robust virus validation.
Minidisc capsule and associated equipment
The Minidisc capsule is a preassembled disposable filter capsule assembly incorporating virus filter membranes. It is an effective and simple-to-use capsule for all small-scale virus filter testing requirements. This article guides you to optimize approaches for filterability, protein transmission testing, and virus challenges of Cytiva membranes in the Minidisc capsule.
Details of how to use Minidisc capsules are contained in the instructions for use. Make sure to follow these instructions closely to ensure successful operation. The following are some key specifications that are relevant to planning experiments and analyzing data generated with the Minidisc capsule (Fig 1, Table 1 and 2).
Fig 1. Laboratory equipment to use. (A) Recommended lab assembly. (B) Pegasus SV4 Minidisc capsules.
Table 1. Recommended equipment details
| Part | Product code | Supplier |
| Novasip™ pressure vessel | C3EP1 | Cytiva |
| Sealkleen™ pressure vessel (optional) | ZLK702G23LHKH4 | Cytiva |
| Adapter 1 in. TC/Male Stäubli connector | GFX0290 (two recommended for inlet and outlet) | Cytiva |
| TC clamp and silicone gasket | - | Cytiva |
| Female Stäubli fitting for inlet pressure tubing (6 mm o.d.) | GFX0236 | Cytiva |
| Tubing1 | - | Lab supplier |
| Locking luer valve and hosebarb adapters as required | - | Lab supplier |
| Graduated cylinder | - | Lab supplier |
| Minidisc capsules with Pegasus SV4 virus removal filter membrane | 10MCFSV4 (pack of 3) | Cytiva |
1 Temporary pressures up to 3.4 barg (50 psig) are acceptable, but the target operating and differential pressures should not exceed 3.1 bar (45 psi) to allow for pressure fluctuations during testing. If tests at operating pressures exceeding 3.1 barg (45 psig) are required, a stainless-steel disc holder and 47 mm membrane discs are available.
Caution: All the upstream tubing and connectors used should be rated to at least 4.0 bar (58 psi).
Table 2. Key specifications of the Minidisc capsule
| Effective filtration area (EFA) | 9.6 cm2 |
| Maximum operating pressure | 3.1 barg (45 psig)1 |
| Maximum recommended differential pressure | 3.1 bard (45 psid)1 |
1 Temporary pressures up to 3.4 barg (50 psig) are acceptable, but the target operating and differential pressures should not exceed 3.1 bar (45 psi) to allow for pressure fluctuations during testing. If tests at operating pressures exceeding 3.1 barg (45 psig) are required, a stainless-steel disc holder and 47 mm membrane discs are available.
1. Filterability and protein transmission trials
How to run a filterability trial
For optimal results, you should perform a filterability trial using a calibrated balance (accuracy ≤ 0.1 g) to collect and measure the filtrate mass over time. A sufficiently accurate estimation of density for water and simple buffers is 1 g/mL, but you should determine your product feed density if this is believed to be significantly different from water. The temperature should be kept at the same level as the full-scale process where possible, to give the correct product viscosity.
A buffer conditioning step can be used if desired, to reduce the risk of aggregate formation at the water-product interface of the prewetted membrane. A buffer flush of ≥ 3 mL will be sufficient to condition the membrane if water is drained from the upstream of the Minidisc assembly before processing the buffer. If buffer flux needs to be determined, measure the buffer flow rate over 10 min.
We strongly recommend that filterability trials are run to full processing time for the most accurate estimation of performance. If time or product volume constraints are in place, then the longest processing time possible should be used and results forward-predicted using the Vmax model to estimate throughput.
Forward prediction of throughput using Vmax analysis
Vmax is the estimated value of the maximum capacity of a membrane (1), that is, the throughput that would be reached when the membrane is completely plugged, if time and feed quantity were not restricted and the membrane fouls in line with the standard pore constriction model.
Vmax is calculated from a plot of time over throughput (At/V) against time (t) and is the inverse of the gradient, as shown by the linear form of the standard blocking law equation below:
Equation 1 Constant pressure Vmax linear equation
A = filtration area (m2), t = time (h), V = volume (L),
Vmax = estimated maximum throughput capacity (L/m2),
J0 = initial flux (L/m2/h).
The gradient should be determined from the linear portion of the graph only, as indicated in Figure 2. The initial flux should also be calculated from the same linear portion of the graph and is evaluated as the y-intercept of that linear data. This initial flux is not always as accurate as direct flux measurement, but will give a suitable forward prediction in many cases. Equation 1 can then be rearranged to estimate the throughput achieved at a given time:
Equation 2 Forward prediction of throughput at constant pressure
Fig 2. Data analysis for forward prediction.
Figure 2 shows typical data collected during a filterability run, showing three phases of data collection:
I—Startup effects cause inaccurate and variable data due to the low flux decay relative to measurement accuracy and start-time accuracy. The slope can be under or over-predicted and the time for this phase will vary between tests.
II—The linear portion of the graph used to determine Vmax and J0 for forward prediction.
III—End effects, only seen if the feed sample is filtered to completion and flow reduces to zero due to the feed running out.
Caution should be taken when forward-predicting. The following conditions should be adhered to, to minimize estimation errors:
- Forward-predicted throughput no more than twice the measured throughput.
- Forward-predicted throughput < 90% of the calculated Vmax
- Coefficient of determination (R2 value) > 0.95.
The closer the raw data collection time is to the estimation time, the more accurate the estimated throughput will be. Fouling is more complex than a simple constriction mechanism, however, the model is the most appropriate of all the traditional membrane fouling mechanisms (2) for small forward predictions of limited data sets. Using Vmax to forward-predict throughput relies on the assumption that the gradient measured from the At/V vs t plot remains constant up to the estimation time. The Vmax value itself should therefore be quoted as a maximum capacity with caution, since this definition is based on the assumption that the gradient remains constant until complete blockage. This is often a long and potentially very inaccurate extrapolation for high-capacity membranes such as Pegasus SV4 membranes.
In general, Vmax values for our virus removal filters are very high and exceed the throughput that can be reached in typical processing times. Most of the time membrane performance (batch area requirement) is either independent or weakly dependent on Vmax and batch area requirement is governed by the processing time and membrane initial flux. For high Vmax values, we do not recommend performance comparisons using Vmax, whereas for cases when the membrane is plugged Vmax can potentially be quoted with caution as highlighted above.
Contact us for more advanced fouling analysis if you believe that the model does not fit the raw data.
Typical filterability results
For filterability testing, we used proprietary monoclonal antibody (mAb) solutions and processed a commercially available human gamma globulin (hIgG) solution to represent plasma-derived product applications.
Figure 3A shows a typical flux profile for a Pegasus SV4 membrane challenged with a mAb solution. The fluxes remain constant throughout the experiment for both operating pressures. Increasing the operating pressure for Pegasus SV4 membranes to 3.1 bar (45.0 psi, 0.031 MPa) yields higher flux without any negative impact on flux decay. We therefore recommend using an operating pressure of 3.1 bar to achieve the maximum flux performance. Results are typical for multiple mAb solutions tested, up to 25 g/L.
Figure 3B demonstrates the robust nature of the Pegasus SV4 membrane tested in plasma-derived protein solutions, again showing that steady fluxes can be achieved at high operating pressures. The flux decay rates per unit mass are comparable to customer testing carried out with high purity IVIG solutions.
Due to complexities including process impurity levels, buffer conditions, and donor profiles, results from plasma sources vary significantly from product to product and therefore direct comparisons must be made using your own specific feed solutions.
Fig 3. Increasing performance at higher operating pressure. Typical flux profiles for Pegasus grade SV4 membrane filterability tests at 2.1 bar (30 psi, 0.021 MPa) or 3.1 bar (45 psi, 0.031 MPa) with (A) 10 g/L (1%) hIgG, (B) up to 25 g/L (2.5%) mAb.
Scale-up
Table 3 shows the relative effective filtration areas for scaling calculations.
Table 3. Pegasus SV4 filter scaling calculations
| 47 mm disc (used with FTK200) effective filter area (m2) | Minidisc capsule effective filter area (m2) | 10 in. cartridges AB1USV4 effective filter area (m2) |
| 1.11 × 10-3 | 9.6 × 10-4 | 2.25 |
Constant flow operation
We recommend that small-scale filterability tests and virus validation studies are carried out at constant pressure. Difficulties in maintaining a constant flow and accurately measuring an increasing test pressure will typically generate more experimental noise than a constant pressure test and the results will be less reliable. Therefore, constant pressure testing will always be preferable.
The flow decay for Pegasus SV4 membrane is slow and steady and therefore there is no difference seen between constant flow and constant pressure testing. Permeability (L/m2/h per unit pressure) decays relative to the product throughput and is independent of the pressure applied. Therefore, using constant flow (or stepwise increases in the flow rate) to eliminate fouling due to high initial flow, as may be required in certain sterile filter applications, is not necessary for Pegasus SV4 virus filtration. What may appear to be a lower flux decline will actually be an equivalent permeability decline.
If required, the key to successful constant flow operation is a pump that is capable of supplying the required flow rates accurately up to the maximum test pressure, without pump slippage or pulsing of flow/pressure.
Protein transmission
Protein transmission studies can be conducted concurrent with filterability testing by collecting samples of feed solution before and after filtration and subjecting them to protein assays.
Where the target protein is the major protein species, either a generic protein assay or a target protein-specific assay can be applied. Other assays may be employed on the filtrate to assess conformation, biological or enzymatic activity, as appropriate.
Typical protein transmission is > 95%, although the exact level will depend on the product concentration, quality, stability and process throughput. Take care not to underestimate protein transmission from the Minidisc capsule by considering the hold-up volumes of the Minidisc capsule as described in the following section.
Accounting for minidisc capsule hold-up volumes
During normal operation there will be a small volume of liquid within the Minidisc capsule that remains in the system when switching from one process fluid to the other. The total hold-up volume is 2.1 mL. The downstream hold-up volume is 0.7 mL.
Water or buffer will make up 2.1 mL of the initial filtrate collected from the filter. For virus challenges the effect on retention will be negligible, with an error of 0.01 in the log reduction value (LRV) at the recommended aliquot volume of 100 mL. For any small grab sample taken directly from the capsule and not the pooled filtrate, the recovery will be accurate if taken after the first 2.1 mL of filtrate. Again, the throughput at which these retentions and recoveries are quoted should be adjusted to account for hold-up volume, and this is detailed in Table 4.
For protein transmission the hold-up volumes will have a greater impact, particularly at lower volumes. Grab samples taken after the first 5 mL of filtrate will provide accurate assessments of the protein transmission, but for total pool measurements and full mass balance the total protein input must be based on a value that is. Alternatively, an FTK200 disc holder and 47 mm disc can be used for more accurate transmission assessments.
Table 4. Throughputs at various filtrate volumes
| Volume collected (mL) | Throughput (L/m2) | Adjusted throughput (L/m2) to account for hold-up volumes (for protein transmission/viral retention) |
| 1.0 | 1.0 | 0.0 |
| 2.1 | 2.2 | 0.0 |
| 10.0 | 10.4 | 8.2 |
| 25.0 | 26.0 | 23.9 |
| 50.0 | 52.1 | 49.9 |
| 100.0 | 104.2 | 102.0 |
| 150.0 | 156.3 | 154.1 |
| 200.0 | 208.3 | 206.1 |
| 300.0 | 312.5 | 310.3 |
| x | x/0.96 | (x – 2.1)/0.96 |
Calculating flux
When aliquots are taken, the flux (L/m2/h) for that aliquot is simply the total unadjusted throughput (L/m2) divided by the time (h). If continuous data is collected using a balance, then the flux can be charted throughout the experiment. Many different options for calculating flux from continuous data exist with varying complexity. One solution we recommend is that the flux at a given data point should be calculated by the slope of the throughput and time data up to 5 min either side of the data point. Calculating the instantaneous flux between every time point collected can lead to significant variation in the calculated flux due to the discrete nature of the filtrate drops, especially when collecting data over small time intervals.
Contact us if you require further advice on flux and throughput analysis or any other aspect of Minidisc operation.
2. Virus spike challenge testing
The validation of virus filtration processes requires special attention, as both the filter manufacturer and you serve vital roles. The filter manufacturer has responsibility for ensuring that each filter will perform to the same specification. You as the user must demonstrate that the selected filter satisfies the needs of your process.
Important factors to consider in designing product-specific viral filter retention validation studies:
Proper design of the product-specific validation study is critical to ensure success of the study. Usually, the retention study follows a scaled down version of the full-scale process. Some of the important factors that need to be addressed in the study design are:
- The choice of spike viruses (models).
- Target reduction factor.
- Spike virus titer.
- Test feedstock comparable to process feedstock with respect to, for example, concentration, temperature, chemistry, etc.
- Equivalence of scale-down filter to process scale filter.
- Same volume to filter area ratio for the test and process filter.
- Inclusion of proper study controls.
- Inclusion of measures aimed at removing virus aggregates potentially present in the spiked challenge solution (e.g., spike prefiltration).
How to run a virus spike challenge test
Contact your local Cytiva representative for detailed instructions of the steps required for successful operation of virus spike challenges, specifically the pre- and post-use testing required.
As with filterability studies, we recommend to run virus spike challenges in constant pressure operation. Difficulties in maintaining a constant flow and accurately measuring the test pressure will typically generate more experimental noise and variability compared to a constant pressure test, and the results may not be representative of the large-scale process filter. Therefore, constant pressure testing will always be preferable, even where a pump is to be used at process scale, with the pressure limits set by the expected full scale pump performance. A constant pressure will also always be required to run post-use installation checks to verify the correct operation of the challenge and release the samples for viral assay.
General protocol recommendations
It is best practice to minimize the amount of non-viral contaminants added to the product in spike studies to keep maximum equivalency between viral validation and production-scale feedstreams. Therefore, excessive spiking, which also increases virus preparation-derived contaminants, is not optimal. We recommend that virus spikes should be designed on the basis of required input titer rather than a particular spike percentage. Our recommended approach is to use a spike level that achieves a 106 pfu/mL input titer (or another appropriate target titer based on your requirements).
High spike percentages (> 1%) can be necessary, for example due to low stock titers. In these specific cases the robustness to fouling of Pegasus SV4 allows the use of these spike percentages without additional flux decay impacting on the throughput that can be validated. The spike level should always be minimized to maintain equivalency and control any additional contaminants, but spikes of up to 5% can generate acceptable throughputs where required. Discuss this further with your virus test laboratory for additional options or contact your local Cytiva representative for assistance.
Filterability and protein transmission trial results should guide the target throughput for virus filtration. For initial virus spike challenges (or bacteriophage studies) the filtrate should be collected in at least two aliquots. The recommended maximum aliquot volume is 100 mL (104 L/m2). Once retention data has been established then aliquot volumes can be increased based on assessment of the data with respect to target retention. Contact your local Cytiva representative for more detailed discussion of aliquot plans.
Collection of aliquots should be in individual graduated sterile containers. The time taken to collect each aliquot should be recorded to calculate the flux.
If a product recovery buffer flush sample is required, we recommend priming the upstream volume with buffer and flushing through 3 mL of buffer, or another appropriate amount as determined by protein transmission studies.
Virus clearance is measured by the log titer reduction (LTR) or LRV, which is the base-10 logarithm of the ratio of the total virus input and total virus measured in all filtrate aliquots. For an individual aliquot or grab sample this simplifies to the ratio of feed concentration (Cfeed) to filtrate concentration (Cfiltrate).
Equation 3 LTR or LRV
Typical virus spike challenge results
Table 5. Example virus spike challenge results for Pegasus SV4 virus filters
| Carrier fluid | Virus spike reduction |
| Plasma derivative against hemophilia | > 4 log PPV |
| Mammalian cell cultured derived protein | > 4.8 log PPV |
| Post-cation exchange monoclonal antibody | > 5.5 log PP7 |
Table 5 shows details of the typical virus removal performance that can be expected from Pegasus SV4 membrane in a variety of protein feed solutions. Figure 4 shows typical performance of Pegasus SV4 membrane in a 1 g/L BSA solution as per the PDA guidelines (3) for bacteriophage model parvovirus (PP7) and porcine parvovirus (PPV). Live virus testing was carried out at an independent virus validation test laboratory. Filtration was carried out at 3.1 bar (45.0 psi, 0.031 MPa) to achieve the maximum flux performance and demonstrate that Pegasus SV4 membrane is capable of robust viral clearance under these conditions.
Fig 4. Live virus (PPV, n = 4) and model virus (PP7, n = 9) retention performance of Pegasus Grade SV4 virus filter membrane in 1 g/L BSA at 3.1 bar (45 psi, 0.031 MPa). Live virus testing was carried out at an independent virus validation test laboratory.
3. Key virus filtration factors
There are several product and process parameters that may affect microbial, including virus, retention by filtration (Table 6).
Table 6. Product and process parameters for virus filtration
| Product parameters | Process parameters |
| Protein concentration | Batch size |
| Amount of aggregates | Temperature |
| Amount of contaminants | Process (filtration) time |
| pH | Pressure differential or flow rate |
| Viscosity | Throughput (volume to filter area ratio) |
| Ionic strength | Prefiltration |
Details of these parameters are outlined in PDA Technical Report 41-08 ‘Virus Filtration’ (4). General guidance for all filters is addressed in detail in PDA Technical Report 26-08 ‘Sterilizing Filtration of Liquids’ (5) and much of this guidance can be applied to virus filters.
These parameters should be considered when running filterability optimization studies and designing viral clearance validation tests for virus filters. Specific recommendations for Pegasus SV4 virus filters are detailed in the following sections.
Operating differential pressure
As shown in Figure 3, increasing operating differential pressure increases the flux of Pegasus SV4 membrane and this is maintained across the course of the test. Testing with polyclonal human IgG solutions having different fouling levels at a variety of different pressures has demonstrated no significant change in the level of fouling of the Pegasus SV4 filter membrane (as measured by Vmax) from 2.1 bard (30 psid) to 3.1 bard (45 psid) operating differential pressure.
Typical bacteriophage clearance by Pegasus SV4 filter membrane in a 1 g/L BSA solution, as per the PDA recommendations (3) is > 4 log and consistent from 2.1 bard (30 psid) to 3.1 bard (45 psid) operating differential pressures. This demonstrates that with Pegasus SV4 filter membrane, the optimum filterability performance seen at higher pressures does not impact retention performance.
For operating pressures above 3.1 barg (45 psig) the FTK200 stainless-steel disc holder and 47 mm Pegasus SV4 discs are available.
Throughput/processing time
Under many process conditions, other virus filters characterized by high initial flow rates will display rapid decay in flow and become less economical over time compared to a fouling resistant constant flow filter such as the Pegasus SV4 filter. Hence, the most economical approach for virus filtration is to allow for longer processing times using a fouling resistant filter and therefore achieve higher throughputs with a minimized cost per batch.
Viral filtration validation testing must be run to at least the expected maximum process throughput (volume to filter area ratio). This will correspond to the expected maximum process time. Filterability studies should also be run to the maximum throughput, although initial scouting studies can use smaller volumes and forward predict performance. This is important due to factors related to process throughput and time such as product stability over the processing time and changes in performance at the higher loading levels present during extended processing.
Temperature and viscosity
Higher processing temperatures can reduce product viscosities and thereby increase filtration flux. Lower temperatures tend to increase viscosities and reduce filtration flux rates.
pH and ionic strength
Ionic strength and pH may have effects on processing parameters like filtration flux rates and total throughput, but also can affect properties of the spiked viruses in the carrier fluid. Therefore, careful control of pH is required during all virus filter testing.
No specific recommendations are given for pH and ionic strength when using Pegasus SV4 filter membrane, as optimal conditions may vary for different products. Other buffer components, for example, stabilizers, excipients, etc., can also impact the overall filterability performance. In general, extremes of pH (< 4, > 8) and high ionic strength (> 1 M) should be avoided unless there is evidence of product stability at these conditions.
Product aggregation
Product aggregation can be caused by a variety of factors such as extremes of temperature, ionic strength, and pH (also at pH ≈ pI). Not only percentage aggregate content, but also aggregate size distribution may impact virus filter performance. Process steps including virus inactivation and freeze thawing can also introduce aggregation. Some products may also aggregate over time due to intrinsic instability.
An important benefit of Pegasus SV4 filter membrane is its high resistance to fouling for a range of aggregates, which results in outstanding throughput capacity in both dilute and complex, concentrated biological fluids. This enables maximum virus filtration economy and efficiency. General precautions present during product development to maintain product stability are typically sufficient to ensure that the Pegasus SV4 filter can process the aggregate burden in product feedstreams with low flux decay. Where significant flux decay is seen, we recommend the use of prefiltration to improve the overall process performance (see the Prefiltration section for details).
Protein concentration
Pegasus SV4 filters have proven to achieve stable flows over a wide range of process conditions, including different protein concentrations, because of robustness of flux and resistance to fouling.
For all protein solutions, an optimum concentration will exist where a given mass can be processed with the minimum amount of filter area. It will be a balance between three effects:
- Reduced flow at higher concentrations due to increased viscosity.
- A decrease in capacity at higher concentrations.
- Reduction in process volume at higher concentrations.
As well as minimizing costs, operating at or close to this optimum is also preferable since variations in batch concentration will have lower impact on performance. This is especially true for a robust, fouling-resistant virus filter such as the Pegasus SV4 filter, which typically has a relatively wide and flat optimum design space.
Optimum protein concentrations for Pegasus SV4 Minidisc capsules are typically > 30 g/L and performance is stable around these optima. Therefore, typical variations in process concentrations are not likely to impact on Pegasus SV4 performance but should still be considered in robustness studies. In general, it is not necessary or recommended to dilute or select a process position for virus filtration with a lower concentration. However, for certain extremely high fouling feeds or products with unusual viscosity trends, this may be necessary and is more likely to be beneficial where undiluted concentrations exceed 50 g/L. Where concentrations are low (< 20 g/L) Pegasus SV4 performance will still be strong, however, where there are process positioning options, it is likely that performance will be improved by operating at higher concentrations.
Higher concentrations are possible but performance may be better at lower concentration process intermediates in this range or after dilution.
Prefiltration
Prefiltration requirements will vary from feed to feed based on the presence of various sizes of aggregates or contaminants. Pegasus SV4 filters are capable of performing without any prefiltration beyond upstream sterilizing filtration (0.2 or 0.1 μm membrane) that may be already built into the purification process. This has been demonstrated in many mAb and plasma protein tests and highlights the robustness to flow decay of the Pegasus SV4 membrane. Where an existing sterilizing grade prefilter is not in place, we recommend Fluorodyne™ II DJL filters (0.1 μm rated). Where required, there are several other sterilizing grade filter options (Table 5).
Each process and its associated contaminant profile are different and where high levels of particulate contaminants are present, protection of Pegasus SV4 filters by membrane prefiltration may be required. If flow decays are substantially faster than the typical performance seen in Figure 3, an additional prefilter should be considered. Where flux decay is < 20% the process is unlikely to benefit from an additional prefilter. Where flux decay is > 50% a prefilter is likely to make the process more economical.
A range of prefilter options capable of removing finer aggregates are available as listed below (Table 7) and should be discussed with your Cytiva representative for recommendations.
Table 7. Prefilter recommendations
| Market | Process fluid | First prefilter | Second prefilter (optional)1 | Final filter |
| General market | Typical combination for many fluids | Fluorodyne II DJL Fluorodyne EX EDT |
Not required | Pegasus SV4 Ultipor VF DV20 |
| Biotech market | Low fouling fluids such as mAbs after purification by chromatography | Supor™ ECV Supor EKV Supor EBV Ultipor N66 NF |
Not required | Pegasus SV4 Ultipor VF DV20 |
1 For fine particles or aggregates < 0.1 μm a second prefilter might make the process more economical.
Summary of recommended design space using Pegasus SV4 filters
Please note that the following design space specifications are a guideline for optimal performance of Pegasus SV4 filters (Table 8). Prior knowledge and understanding of the particular feed to be tested should also be applied and filterability studies are recommended before virus spiking to confirm performance and reproducibility.
Table 8. Design space recommendations when using Pegasus SV4 filters
| Differential pressure | 3.1 bar (45 psi, 0.031 MPa) |
| pH and ionic strength | • Highly product specific optima. • General recommended ranges: - pH 4 to 8. - < 1 M ionic strength • More extreme values can be tested if the product remains stable under those conditions. |
| Protein concentration | • Optimum performance at higher concentrations (20 to 50 g/L). • < 50 g/L may require further studies for high fouling and highly viscous solutions as a lower concentration may provide improved performance. |
| Prefiltration | • 0.2 μm sterilizing grade filter as a minimum. • Flux decay < 20% no additional prefilter required. • Flux decay 20% to 50% additional prefilter recommended. • Flux decay > 50% additional prefilter required (see Table 5 for prefilter options). |
| Spike titer | 106 pfu/mL |
| Spike concentration | Minimum required to generate target spike titer: • ≤ 1% = minimal additional flux decay. • 1% to 5% = acceptable additional flux decay if necessary. |
| Virus challenge aliquot throughputs | • Minimum two aliquots. • ≤ 111 mL (100 L/m2) per aliquot for initial tests. |
REFERENCES
- Badmington F, Wilkins R, Payne M, Honig ES. Vmax testing for practical microfiltration train scale-up in biopharmaceutical processing. Pharmaceutical Technol. 1995;19:64–76.
- Hermia J. Constant pressure blocking filtration laws - Application to power-law non-Newtonian fluids. Trans Inst Chem Eng. 1982;60:183–187.
- Lute S, Riordan W, Pease LF 3rd, et al. A consensus rating method for small virus-retentive filters. I. Method development. PDA J Pharm Sci Technol. 2008;62(5):318–333.
- Parenteral Drug Association (PDA). Virus Filtration. Technical Report No. 41. PDA J Pharm Sci Technol. 2008;62.
- Parenteral Drug Association (PDA). Sterilizing Filtration of Liquids. Technical Report No. 26. PDA J Pharm Sci Technol. 2008;62.