- Product loss from post-filtration hold-up volume following sterile filtration can significantly reduce yields, especially for high-value biologics and advanced therapies.
- This article reviews recovery strategies, including air purging, sterile gas-assisted recovery, buffer flushing, and single-use (SU) air-displacement approaches.
- These established methods improve product recovery while maintaining closed processing and minimize operational complexity in biopharmaceutical manufacturing.
Introduction
Single-use system (SUS) production line losses due to hold-up volumes are a commonly recognized problem. The impact of the problem is increasing as the cost per unit volume of drugs increases; either in the field of monoclonal antibodies (mAbs) with the trend towards high concentration formulations, or with the emergence of advanced therapeutics such as viral vectors, mRNA, or personalized medicines. Aside from impacting upon material costs in drug production, line losses during development stages before commercialization can also reduce the availability of product needed to achieve critical milestones for new drugs.
Sterile filtration is a unit operation where maximizing recovery of fluid is desired but challenging to accomplish. Workarounds to facilitate improved yields at this process step are implemented with some success but can do so at the cost of system design, ease and security of operation and validation complexity. In this article, we assess known approaches to recover fluid entrained downstream of sterilizing grade filters in closed single-use filtration systems alongside some alternative methods. The alternative methods utilize attributes and capabilities of SUS design and operation helping you to achieve the objective of minimizing critical line losses, with less risk.
Hold-up volume downstream of a sterilizing grade filter in an SUS
After the sterile filtration of drug substance or drug product using a closed, SUS, you will typically be left with a filter capsule and tubing containing residual fluid, termed the ‘hold-up volume’.
Fluid upstream of the filter can usually be easily purged downstream by gas displacement, either from a pressurized gas source or by generating a low gas pressure using a transfer pump such as a peristaltic pump to run a system dry. Fluid in the downstream the “downstream hold-up volume” (Fig 1, 2), is more difficult to recover. This is because a wetted, integral filter does not permit the passage of significant volumes of air at low pressures to displace the downstream hold-up volume and that any access to the sterile side of the filter carries the potential to impact the sterility of the filtered fluid.
The percentage of the overall batch volume that this liquid volume represents (and therefore the potential yield impact if not recovered) will depend upon several factors. These include the choice of the filter you have used and the dimension, configuration and orientation of tubing connecting the filter to any receiving biocontainers.
Fig 1. Downstream hold-up volumes in a typical single-use system for the filtration of a drug product.
Fig 2. Hold-up volumes of a typical filter capsule. (A) Downstream hold-up volume (including connected tubing), (B) media hold-up volume, (C) upstream unrecoverable volume.
Irrespective of the methods used to recover this downstream hold-up volume, it is important that the chosen filter is the result of a thorough selection process supported by process validation studies that demonstrate performance under the proposed process conditions. A favorable outcome of filter selection and sizing is the specification of a compact filter that can be connected to tubing of the shortest length and smallest internal diameter as is practical.
In doing this, you will have minimized the amount of downstream hold-up volume they are challenged to recover after filtration and somewhat reduced the amount of time and effort required to implement their chosen workaround.
Working on the basis that the length of tubing connected downstream of a filter can usually be specified independently of filter sizing, Figure 3 illustrates the potential downstream product recovery benefit, or limitations, afforded by different sterilizing grade filters prior to the implementation of product recovery procedures, based on design and feed capacity.
Fig 3. Example of two units with similar effective filter areas but differing downstream volumes.
Methods of recovering downstream hold-up volume
The following methods show different approaches to the recovery of fluid downstream of a sterilizing grade filter, in a closed SUS. Each achieves an improvement in yield, but includes different levels of design, operational and validation complexity, and ultimately risk.
Before adoption, any recovery procedure must be considered during formal risk assessment and FMEA (failure mode effect analysis).
Method 1: Manual manipulation
By adjusting the relative positions of the biocontainer, filter, and connected tubing, you can redistribute air within the downstream collection vessel to displace the downstream hold-up volume (Fig 4).
Fig 4. Manual manipulation.
Advantages
- No design consideration required.
Considerations to manage CQA risk/ for cGMP implementation
- The highly operator-dependent manual manipulation significantly increases the risk of system damage or leakage.
Method 2: Low pressure air purge
Here, you apply a gas pressure to the upstream side of the filter at a pressure lower than the bubble point (BP) value of the filter when wetted with product.
At this pressure, two separate mechanisms facilitate product recovery.
- Direct upstream recovery is achieved by forcing residual fluid in the upstream side of the capsule through the filter membrane and through the compression of the filter layers.
- Downstream fluid recovery is achieved as gas slowly diffuses through the wetted filter membrane to displace any fluid in the downstream volume in the system.
The rate of diffusion is related to the pressure applied to the filter, the area of filter membrane in the unit, the pore size and the diffusion characteristics of the fluid wetting the filter membrane. Common filter designs typically dictate the greater the amount of filtration area, the greater the downstream volume in the filter’s core.
Note: This procedure tends to be slow due to the low rate of gas diffusion through the filter in proportion to the volume in any tubing downstream of the filter.
This operation can however be started during an in-situ post-use filter integrity test when using a diffusive flow (forward flow) test method. A time greater than the natural test time may be required for most filter/tubing combinations; however, the filter integrity test instrument provides a carefully controlled way to apply a defined pressure.
Fig 5. Air purge below BP pressure of the filter membrane.
Advantages
- Uses a defined parameter that can easily be validated to facilitate recovery. May be incorporated into an in situ, post-use filter integrity test.
Considerations to manage CQA risk/ for cGMP implementation
- Potentially long recovery times.
- Process specific validation of the filter for bacterial retention at a pressure above that employed during the filtration period may be required.
- Viscous feeds may not permit sufficiently high diffusive flow to evacuate downstream fluid.
Method 3: High pressure air purge
Here, you apply a gas pressure above the BP value of the filter when wetted with the product to the upstream side of the filter to ‘blow down’ the system.
At higher pressure, the filter membrane pore structure is de-wetted and the bulk gas flows through the largest pores of the filter membrane to rapidly displace any fluid entrained in the filter membrane in the downstream hold-up volume.
This achieves recovery more rapidly than an air purge below BP; however, significant risk of pressure-related damage to downstream components exists without adequate control.
Fig 6. Blown down at above minimum BP pressure for the filter membrane.
Advantages
- Rapidly displaces fluid in the downstream hold-up volume, including some entrained in the filter membrane.
Considerations to manage CQA risk/ for cGMP implementation
- Variability in filter membrane BP may make selecting a pressure, that will always work, more complex.
- Process-specific validation of filter for bacterial retention at the BP pressure should be considered in the risk assessment.
- An SUS would need to be carefully designed to support the high pressure/gas flow downstream whilst not placing batch at risk of loss due to system damage/leak, or product exposure to high gas pressure and foaming.
- Typical single-use (SU) components—such as tubing, biocontainers, and junctions—may not withstand these pressures.
Method 4: Sterile gas-assisted recovery
Here, you apply a low gas pressure (< 50 mbar) through a hydrophobic sterilizing grade filter, connected to the downstream side of the final filter.
In this arrangement, sterile gas can be introduced into the downstream hold-up volume of the filter to displace any entrained fluid.
Fig 7. Sterile gas-assisted recovery method.
Advantages
- Gentle, low risk to SU components.
Considerations to manage CQA risk/ for cGMP implementation
- The downstream gas filter enabling the recovery is in a critical position (downstream of the sterilizing filter). The integrity of this gas filter and its connection to the downstream product flowpath are therefore also critical for the sterility of the downstream flowpath. The integrity of this critical hydrophobic gas filter typically cannot be verified in situ before use and must be confirmed post-use.
- This gas filter and its connection to the downstream flowpath are adding complexity to the system design, validation and operation.
- Post-use testing of the gas filter in addition to sterilizing liquid filters adds operational complexity.
Method 5: Buffer flush
Here, you would pass a measured volume of buffer through the filter after filtration is complete. This hydraulic flush pushes entrained fluid through the filter membrane and recovers the downstream volumes to the collection containers.
Fig 8. Buffer flush.
Advantages
- Can recover a substantial amount of drug entrained in filter membrane and downstream volumes.
- Additional bacterial retention validation may not be needed if the flow rate is controlled to below the process flow rate.
Considerations to manage CQA risk/ for cGMP implementation
- Flush volumes need careful optimization to maximize recovery of each drug/filter installation while avoiding any undesirable product dilution.
Alternative techniques for fluid recovery in SUS
To facilitate recovery of downstream hold-up volume, the following techniques utilize sterile air naturally introduced as part of a filtration operating procedure, or that has been deliberately introduced into the single-use system prior to filtration. In these respects, the gas used to recover product is integral feature of the filtration system and process step, rather than an external aspect introduced afterwards.
The techniques are compatible with automated or manual systems and can minimize some of the CQA risks associated with the described recovery methods.
Like the “Downstream recovery with a gas filter” method described above, these methods offer a gentle means of recovering fluid entrained downstream of the membrane. They are mainly differentiated as they permit the maintenance of a closed manufacturing process downstream of the critical final sterilizing filters and do not necessitate the integrity testing (IT) of any additional filters connected downstream of the final sterilizing grade filter to confirm this state.
To employ any of the following techniques, we recommend that any biocontainers positioned for collection of filtered drug product are below the final sterilizing grade filter in a free-draining orientation. The final sterilizing filter should also be positioned in the vertical position with the filter outlet at the bottom.
Method 6: Air displacement using secondary tubing
This method requires a system design like that shown in Figure 1. The method features an additional tube connected to a port on the highest point of the fluid-receiving biocontainer to a connection point close to the filter outlet. This tube is closed with a valve before filtration.
Before, or upon operation of the system, a volume of sterile gas, at least as large as the downstream hold up volume, is transferred into part of the system downstream of the filter. The passage of gas is through the dry filter membrane of the final liquid sterilizing grade filter (prior to priming or wetting) at a pressure within the operating limits of the system.
The gas may be introduced by passing fluid through the filter during priming or filtration or added prior to the supply of any liquid to the system, in a controlled manner.
The gas is collected via the tubing connected to the filter’s outlet connected to the empty system downstream, including the biocontainer used to collecting the filtrate.
After filtration, the opening of the valve on the secondary tube allows the gas captured in the biocontainer to travel up to the filter core, recovering drug product into the downstream biocontainer.
This method is optimal for small batch volumes when the entire volume of fluid can be filtered into the collection bag.
Fig 9. Air displacement using secondary tubing.
Advantages
- Fast recovery.
- No downstream gas filters needed, negating the need to integrity test an extra downstream gas boundary filter either after installation or after use.
- No dilution.
- Low risk to single-use system component integrity.
- Avoids consideration of the effect of gas pressure on the retentive performance of the sterilizing filter.
Considerations to manage CQA risk/ for cGMP implementation
- Complexity of the SUS.
- Gas flow into system before filtration must be carefully controlled to avoid over-inflation.
- This method is optimal for small batch volumes when the entire volume of fluid can be filtered into the collection bag.
Method 7: Air displacement from a downstream biocontainer
This method requires a system design like that shown in Figure 1 with a biocontainer connected to the filter outlet tubing, located upstream of the collection biocontainer, with an appropriate valve arrangement (Fig 10).
Fig 10. Gas collection in a downstream gas collection bag.
A volume of sterile gas, at least as large as the downstream volume, is transferred through the dry membrane of the filter (prior to any filter priming or wetting) into a downstream gas collection bag. This gas is later to be released by opening a valve that permits it to move into the downstream hold-up volume to recover product.
Advantages
- Fast recovery
- No downstream gas filters needed, negating the need to integrity test an extra downstream gas boundary filter either after installation or after use.
- No dilution.
- Low risk to single-use system component integrity.
- Avoids consideration of the effect of gas pressure on the sterilizing filter.
Considerations to manage CQA risk/ for cGMP implementation
- Gas flow into bag may need to be carefully controlled to avoid over-pressurization of the bag.
- Potential to use flush bag as downstream recovery biocontainer.
- Consider for use with priming fluids to minimize dilution of filtrate.
Fig 11. Downstream biocontainer product recovery method in use on the Allegro Connect bulk fill system.
Recommendations
Where product recovery in closed systems is important, there are variety of methods that can be applied as part of standard operating procedures, although with risks associated with some methods that should be controlled.
The preferred technique should be established to account for the following factors, and accompanied with a risk assessment as part of the process design documentation.
- Simplicity: how easy is the technique to apply to the process? Does it require specific skills or operator training? Is it compatible with the existing SUS design or control hardware?
- Safety: are there any new process risks related to the recovery procedure such as pressure-related failure modes? Are there any procedural or design controls in place to prevent component failure or process deviations that may impact CQA?
- Reproducibility: can the process be defined in such a way as to eliminate batch-to-batch or operator-to-operator variability?
- Integration: can the method be integrated into an automated solution?