What you will achieve

Recover purified plasmid DNA for storage or concentration measurement.

Starting material: 1-5 mL overnight culture of E. coli grown in rich medium.


1. Before you begin

  • Add the supplied RNase A solution to buffer P1 to a final concentration of 100 µg/mL. Mix, then store at 2-8°C.
  • Add ethanol, not denatured, to buffer PE. Follow the instructions on the bottle label.
  • Check buffers P2 and N3. If you see a precipitate, heat the buffer to 37°C and gently agitate it. Do not shake the buffer P2 bottle vigorously.
  • Close the buffer P2 lid immediately after use to prevent acidification by CO2 in the air.
  • Allow all buffers to reach room temperature.
  • Perform centrifugation at room temperature and 10 000-14 000 × g unless another setting is specified.
  • Use the three receiver tubes supplied with each NAB Nanosep filter insert. Do not substitute other receiver tubes.
  • After vortexing, briefly centrifuge the tubes to remove droplets from inside the lid.

If the sample does not pass through the membrane

Centrifuge it again until it has passed completely through. Full-speed centrifugation does not affect DNA yield or purity.


2. What you will need

Equipment

  • Microcentrifuge
  • Spectrophotometer
  • Vortex mixer

Consumables

Supplier Item Product code
Cytiva Nucleic acid binding (NAB) Nanosep centrifugal filter ODNABC33, ODNABC34
VWR Ethanol, not denatured 71001-866
VWR Spectrophotometer cuvettes, approximately 100 µL (260/280 nm) 47743-840
VWR 15 mL tubes, RNase- and DNase-free 89401-574
VWR 1.5 mL microcentrifuge tubes, RNase- and DNase-free 76005-210

Reagents

Supplier Reagent Product code
Qiagen Buffer PE 19065
Qiagen Buffer P1 19051
Qiagen Buffer P2 19052
Qiagen Buffer N3 19064
Qiagen Buffer PB 19066
VWR Tris buffer, pH 7.0 (1 M) 89500-584
Qiagen RNase A (100 mg/mL) 19101
Qiagen Nuclease-free water 129115

3. Process overview

Harvest the cells → Resuspend the pellet → Lyse the cells → Neutralize the lysate → Clarify the lysate → Bind the plasmid DNA → Wash with Buffer PB → Wash with Buffer PE → Remove residual buffer → Elute the plasmid DNA → Recover the purified DNA

4. Method

1. Harvest the cells

  • Centrifuge the bacterial culture at 3000 × g for 10 min at room temperature in a benchtop microcentrifuge.

Why it matters: This collects the bacterial cells into a pellet.

2. Resuspend the pellet

  • Discard the supernatant.
  • Resuspend the bacterial cell pellet in 250 µL Buffer P1 containing RNase A.
  • Transfer the suspension to a microcentrifuge tube.

Success check: No cell clumps remain.

3. Lyse the cells

  • Add 250 µL Buffer P2.
  • Invert the tube 4-6 times to mix.
  • If needed, continue inverting until the solution becomes viscous and slightly clear.

Important

Do not vortex because vortexing can shear the DNA.

Do not allow lysis to continue for more than 5 minutes.


4. Neutralize the lysate

  • Add 350 µL buffer N3.
  • Immediately invert the tube 4-6 times.
  • Mix thoroughly.

Why it matters: Thorough mixing prevents localized precipitation.

5. Clarify the lysate

  • Centrifuge for 5-10 min at 10 000-14 000 × g.

6. Bind the plasmid DNA

  • Transfer 500 µL supernatant at a time to the NAB Nanosep device.
  • Centrifuge for 45-60 s at 10 000-14 000 × g.
  • Discard the flow-through and retain the receiver tube.
  • Repeat with the remaining supernatant.

7. Wash with Buffer PB

  • Add 500 µL Buffer PB to the NAB Nanosep device.
  • Centrifuge for 60 s at 10 000-14 000 × g.
  • Discard the flow-through and retain the receiver tube.

8. Wash with Buffer PE

  • Add 500 µL Buffer PE.
  • Centrifuge for 60 s at 10 000-14 000 × g.

9. Remove residual buffer

  • Discard the flow-through.
  • Centrifuge again at full speed for 1 min.

10. Elute the plasmid DNA

  • Transfer the NAB Nanosep device to a clean supplied receiver tube.
  • Add 50 µL nuclease-free water or Buffer EB to the centre of the membrane.
  • Leave the device to stand for 1 min.

11. Recover the purified DNA

  • Centrifuge for 60 s at 10 000-14 000 × g.
  • Collect the purified plasmid DNA in the receiver tube.

5. Store the purified DNA

Storage condition

Store purified plasmid DNA in nuclease-free water at -20°C for several months.


6. Check your result

Measure DNA concentration

Measure absorbance at 260 nm (A260) with a spectrophotometer. Photometric measurement can be difficult at lower DNA concentrations. Use fluorometric quantification to measure smaller quantities accurately.

Interpret the reading

  • Use A260 readings greater than 0.10 and lower than 1.0 to support meaningful measurement.
  • An absorbance reading of 1.0 at 260 nm corresponds to 50 µg DNA/mL.
  • This relationship is valid only at neutral pH.
  • If dilution is needed, use a neutral-pH dilution buffer.