Due to numerous benefits, including absence of cross-contamination and the elimination of cleaning validation, single-use technology is widely adopted in the biopharmaceutical industry, even in very critical applications, such as formulation and filling. In these applications, the drug product is often high value and/or highly potent.
During manufacture of single-use systems (SUS), a strong quality by design (QbD) approach is applied, and several tests are performed. However, there is also a need to mitigate any risk of product loss or operator exposure due to the potential risk of a SUS leaking during drug product processing, particularly in critical applications (justified by a risk assessment).
As part of a risk assessment and mitigation strategy, one possible approach is to perform in situ point-of-use (POU) leak testing of single use assemblies using the Flowstar V LGR test instrument. This POU leak test was developed to detect leaks that are typically undetectable by visual inspection.
Fig 1. Life cycle of a SUS from manufacturing to its disposal (post-use).
Clearly, 100% of biocontainer bags are leak tested at our manufacturing facilities prior to release and although packaging is optimized to reduce the risk of leaks to SUS during transportation and storage, many SUS designs are complex and the risks of leaks to biocontainer bags may increase during the product life cycle (Fig 1), especially after it has been removed from the box (e.g., during cleanroom intermediate storage).
This application note presents this complementary POU test approach for SUS containing biocontainer bags and provides examples illustrating the use of the Flowstar V LGR test instrument.
Regulatory context
The 2022 revision to European Union (EU) ‘Guidelines for Good Manufacturing Practice (GMP) for Medicinal Products for Human and Veterinary Use Annex 1 Manufacture of Sterile Medicinal Products’ includes new content for closed systems (8.127-8.130) and single-use systems (8.131-8.139). Specifically, paragraph 8.130 emphasizes that the background environment for closed systems should be determined by their design, the nature of the processes involved, and the controls applied. For aseptic processing, or when there is any risk of compromising system integrity, the system must be located in a grade A environment. However, if the system can be shown to remain integral at every use a lower classified area such as grade C may be acceptable.
8.130 “The background environment in which closed systems are located should be based on their design and the processes undertaken. For aseptic processing and where there are any risks that system integrity may be compromised, the system should be located in grade A. If the system can be shown to remain integral at every usage (e.g. via pressure testing and/or monitoring) then a lower classified area may be used. Any transfer between classified areas should be thoroughly assessed (see paragraph 4.10). If the closed system is opened (e.g. for maintenance of a bulk manufacturing line) then this should be performed in a classified area appropriate to the materials (e.g. grade C for terminal sterilization processes, or grade A for aseptic processing) or be subject to further cleaning and disinfection (and sterilization in case of aseptic processes).”
This section of the regulatory guidance sets a strong expectation for 100% integrity control of SUS used in aseptic processing and located outside of grade A, either at the SUS supplier facility or at the end user POU. Integrity of such systems must be verified. Although leak testing does not constitute a formal integrity test (1,2) and therefore cannot provide absolute assurance of system integrity, it plays a valuable role in supporting SUS integrity assurance. In this context, leak testing using the Flowstar V LGR test instrument is a useful solution when applying a cGMP SUS integrity control strategy.
Testing SUS with the Flowstar V LGR test instrument
Test instrument working principle
The Flowstar V LGR test instrument applies a test pressure (between 20 to 50 mbar) to SUS containing Allegro™ biocontainer bags with a nominal volume of up to a 200 L. It measures the flow to be injected into the SUS which is equal to the expansion rate, plus the gas flow out of the tested SUS, if there is a leak. Typically, the test gas is air but other gases such as nitrogen can also be used. The final flow measurement is evaluated against a preset threshold value to determine whether a leak is present.
Considerations for the design of SUS
For a SUS to be leak tested using the Flowstar V LGR test instrument, the system design should meet the following requirements:
- The SUS must be designed so that it can be tested as a completely closed system. Each of the components in the system that will be subjected to the test should withstand the test pressure, 20-50 mbar for SUS including bag chamber(s), for the test duration.
- There must be a connection point into the SUS to introduce the test gas. This SUS gas inlet is typically equipped with a hydrophobic sterilizing-grade filter, providing the sterile gas barrier for filling the SUS with the test gas. A quick connection is required to connect the Flowstar V LGR test instrument to the SUS. A ¾ inch tri clamp (TC) connector is the connection supplied on the Flowstar V LGR hose. A ¾ inch TC and gasket, as well as a ¾ inch adaptor to Staubli (F) and Staubli (M) to 1½ inch TC, are also supplied (Fig 2).
- The SUS should be designed so that the hydrophobic sterilizing-grade filter is appropriately sized to allow deflation or venting when the test is complete. For 2D biocontainer bags, the deflation can be assisted by the Cytiva air evacuation device (Fig 3).
Fig 2. Flowstar LGR SU connection point.
Fig 3. 2D biocontainer bag with connected tubing, showing an air evacuation device for assisted deflation.
Considerations for the test
As the expansion of gas is a function of temperature, any heat transfer could compromise the test result. Therefore, the complete test setup (Flowstar V LGR, SUS, and supporting hardware) must be brought to ambient temperature prior to testing. The test should not be conducted at a location where temperature variations are exacerbated (for example, under air conditioning flow or near a door). Temperature shifts should be avoided during the measurement. To avoid masking leaks, the use of a solution such as a mesh insert is recommended when leak testing 3D bag chambers in plastic or metal totes. It ensures that the 3D biocontainer bag does not directly press against the wall of the tote, which could result in a sealing effect during pressurization and mask a leak, giving a false-pass test result.
Test parameters
Determination of the test pressure
The determination of the test pressure for SUS must be approached carefully. The gas flow through a hole depends on the pressure drop across the hole, so with a higher differential pressure, leakage flow will be higher, and the test will be more sensitive. The test pressure should be high enough to achieve the required sensitivity while remaining below a level that could cause permanent deformation of the biocontainer bag and potentially compromise its integrity.
Input parameters into the Flowstar V LGR test instrument
The Flowstar V LGR test instrument enables POU leak testing of SUS with biocontainer bags of up to 200 L nominal volume using the “SUS leak test” function.
In addition, the Flowstar V LGR test instrument can also be used to test the integrity of sterilizing-grade filters or to leak test SUS without a biocontainer bag by selecting the appropriate test function.
Parameters for SUS leak test
To test SUS including one biocontainer bag or more, a specific “SUS leak test” function is available.
In addition to the information on operator, production area, product name, batch number, and SUS identification (ID), the input parameters for leak testing selected SUS assemblies with the Flowstar V LGR test instrument are as follows:
- The system volume range (< 50 mL, 50 mL to 5 L, and > 5 L). If volume > 5 L, the estimated actual system volume (typically corresponding to the sum of nominal volumes of the bag chambers in the system) should be entered.
- Fill pressure (up to 20% above relative test pressure is possible).
- Fill hold time (120-9999 s).
- Pressure hold time (120-9999 s).
- Test pressure (20-50 mbarg).
- Test time (300-9999 s).
- Maximum flow value (or threshold value) as determined in Table 2.
All parameters should be qualified by the user, taking into account the limits associated with all SUS components.
Results of the SUS leak test
The following test results can be displayed on the screen:
- ‘PASS: Flow within limits: x.xx mL/min’
- ‘FAIL: FLOW OUTSIDE LIMITS’
Determination of threshold values
A pragmatic approach was used to establish the threshold flow values. A minimum of 16 bag chambers of each size were tested to determine the range of measured flow rates. The threshold values were set to provide an adequate safety margin above the measured flow rates.
The values for different sizes of Allegro biocontainer bags with Allegro film are shown in Table 2.
It is important to note that the threshold values are dependent on the type of bag film and on the biocontainer bag design. Therefore, biocontainer bags made from a different film or with a different shape would require specific threshold value determination.
Table 2. Test conditions, threshold values and corresponding theoretical sensitivity in detecting a defect size for SUS with an Allegro biocontainer bag
| Allegro bag size | Fill pressure (mbar) | Fill hold time (s) | Pressure hold time (s) | Test pressure (mbar) | Test time (s) | Threshold value (mL/min) | Approximate leak size detection (µm)* |
| 5 L (2D) | - | - | 125 | 50 | 600 | 3 | ≥ 30 |
| 10 L (2D) | - | - | 125 | 50 | 600 | 6.5 | ≥ 40 |
| 20 L (2D) | - | - | 125 | 50 | 600 | 10 | ≥ 50 |
| 50 L (2D) | - | - | 300 | 50 | 600 | 48 | ≥ 100 |
| 100 L (3D) | 24 | 120 | 120 | 20 | 600 | 19 | ≥ 100 |
| 200 L (3D) | 24 | 120 | 120 | 20 | 900 | 25 | ≥ 125 |
*Assuming a perfect leak.
The expansion of gas is a function of temperature. A temperature shift could compromise the test result, therefore, the complete test setup (Flowstar V LGR, SUS, and supporting hardware) was brought to ambient temperature before testing. The test was not conducted at a location where temperature variations are exacerbated (under air conditioning flow, near door etc.).
These thresholds were determined by selecting values at least three standard deviations above the mean flow value and are also applicable for retesting.
The mean value was calculated from the leak tests performed on newly installed bags (3D biocontainer bags were installed in their specific tote before running the test).
A screenshot of the Flowstar V LGR test instrument screen during a test is shown in Figure 4.
To avoid re-entering the test parameters for a specific single-use system, the instrument software allows you to create, save, and recall test programs, either by selecting the corresponding test program or by scanning its barcode.
The Flowstar V LGR test instrument includes the features required for operating in accordance with U.S Code of Federal Regulations Title 21 Food and Drugs, Part 11: Electronic Records; Electronic Signatures (21 CFR Part 11) and EU GMP Annex 11: Computerized Systems.
It is designed for use in a cleanroom environment.
Fig 4. Screenshot of the Flowstar V LGR test instrument screen with input parameters and test result.
Examples of SUS tested with the Flowstar V LGR test instrument
Examples of tested SUS are shown in Figure 5 to Figure 7.
Fig 5. 100 L Allegro biocontainer bag SUS for product filtration and storage.
Fig 6. SUS leak test of the pre-use post sterilization integrity test (PUPSIT) system including a 20 L flush bag on the manual drug product filtration system.
Fig 7. Leak testing of a final fill design connected to a PUPSIT design with Supor™ prime filter including the critical sterile connection to the beta bag and the beta bag inlet.
Conclusion
The Flowstar V LGR test instrument can be used for leak testing of SUS to detect leaks that may, for example, arise between supplier assembly of the SUS and its actual use. The Flowstar V LGR test instrument performs true POU leak tests after installation of the SUS. No further handling of the SUS is required between the pre-use leak test and drug processing. The sensitivity of the test depends on the system volume, design and applied test pressure.
Leak testing can be selected to reinforce assurance of SUS used with high value products, in critical processing steps outside of grade A, or with highly potent drugs and is a consideration based on your risk assessment. This approach has been designed to be easily implemented in a GMP manufacturing environment and is now successfully implemented in the industry, especially in formulation and filling steps (3).
In addition, the Flowstar V LGR test instrument can be used to perform integrity testing of sterilizing-grade filters, as such, it is a versatile test instrument, suitable for a wide range of applications.
References
- Bio-Process Systems Alliance (BPSA). Design, Control & Monitoring of SUS for Integrity Assurance, Volume 2. 2023. (Section 5.1: What is a difference between a leak test & an integrity test?)
- BPSA Integrity Assurance FAQs - Bio-Process Systems Alliance. Bio-Process Systems Alliance. Published March 31, 2025. Accessed November 12, 2025. https://bpsalliance.org/bpsa-integrity-assurance-faqs/
- Rutigliani, N. (PharmaCongress, Day 2 – Facilities & Technology). Aseptic Sterile Filtration Using Single Use and Isolator Technologies. PharmaCongress Presentation. Retrieved from: https://www.pharma-congress.com/files/userFiles/Presentations/Day2/Facilities-Technology/06_Rutigliani_Aseptic_Sterile_Filtration_using_Single_Use_and_Isolator_Technologies.pdf