Different strategies can be applied to avoid contamination of the environment, the instrument and cross-contamination during a filter integrity test. For low and medium risks, the installation of a vent valve controlled by the filter integrity test instrument is the preferred option. For high risks due to aggressive or hazardous components, strict avoidance of venting in the direction of the filter integrity test instrument is the recommended strategy.
Introduction
To protect operators, the environment, and the filter integrity test instrument from undesirable exposure to liquids, hazardous substances, particulates, or biological components like microorganisms or viruses during a filter integrity test, protective strategies are needed. This document describes strategies to avoid contamination in different possible scenarios.
In the course of a filter integrity test, the most critical step where contamination can occur is the depressurization step at the end of the test. During pressurization at the beginning and during the filter test itself, the direction of flow is from the instrument to the filter, however during venting a significant amount of gas flows into the direction of the instrument. This venting gas could contain residues of the wetting fluid and/or residues of the product if the test is performed after filtration, or if the product is used as a wetting agent for the filter integrity test. If the process liquid used for wetting the filter has hazardous effects in small concentrations, a form of protection of the environment and the test equipment must be applied.
However, even if a liquid with no harmful contents is present on the filter during the integrity test, the risk of contamination of the test instrument should be avoided. Filter integrity test instruments are sensitive instruments and the measurement can be disturbed by contamination of the instrument, either by the wetting fluid or by product residues. Contamination of the instrument or other test equipment (like tubes with product) can also be the source of cross-contamination.
In addition, depending on the wetting agent or product, the venting cannot be carried out in the environment but a controlled venting has to be installed. Different strategies are possible in order to reduce or avoid the risk of contamination of the instrument and of the environment. The most suitable approach applicable to a given installation should be evaluated on a case by case basis and is dependent on the relative level of risk.
- Low risk: If the filter is wetted with a standard wetting fluid like water, the risk caused by contamination is low because these fluids can easily be removed.
- Medium risk: Product residues (or product used as a wetting agent) could cause problems after venting if the product causes a malfunction of the instrument due to corrosion or jamming of valves. Further, this scenario could potentially include the danger of cross-contamination due to presence of liquids or particles in the test system.
- High risk: If the product or parts of it can be hazardous to the operator or the environment, the venting into the environment and contamination of the instrument must be avoided. Cross-contamination could also be considered as high risk for the product.
These different risks mentioned above can be handled by two key strategies in principle:
- Cleaning after contamination.
- Avoiding contamination.
Cleaning after contamination
Cleaning can be an acceptable solution for low-risk situations and after unpredicted contamination. Choosing an adequate cleaning procedure requires adapting to the contamination. Different product contamination, for example, might require different cleaning procedures which must be qualified beforehand.
Avoiding contamination
For higher risks, a cleaning strategy alone is not acceptable and avoidance is the preferable strategy. This basically means that the filter cannot be depressurized by the integrity test instrument but must be vented under controlled conditions, (e.g., into a kill tank) or appropriate measures to protect the instrument and the environment (e.g., installation of an additional filter) must be taken. As the filter must be vented also in other situations during use in the filtration process, a controlled venting is required for critical products.
There are different options and components which can also be combined if necessary to avoid contamination of the instrument, including the risk of cross-contamination and contamination of the environment during the integrity test.
Installation of the integrity test instrument
The integrity test instrument should be connected to the top (vent) of the filter housing (Fig 1). Connections to the drain must be avoided because fluid might collect at the drain and can be forced into the direction of the instrument when the test is completed and the housing is depressurized.
Fig 1. Installation of a filter integrity test instrument. (A) Incorrect installation with the instrument connected to the drain. (B) Correct installation with the instrument connected to the top of the filter.
Installation of an external vent valve
Even when connected to the vent, there is the potential risk of small droplets being pushed into the instrument during depressurization. These droplets might be wetting fluids like water or alcohol solvent mixtures, but also product residues when testing after filtration or with product when used as a wetting agent. Particularly, product residues which can be sticky or aggressive could attack the instrument and for example could cause failures of the internal valves. An external 3-way valve between the instrument and the filter, which is controlled by the instrument, allows venting of the filter outside of the instrument, ensuring that no fluid comes into contact with the test instrument (Fig 2). The same 3-way valve can also be used to direct the gas coming out of the filter into a controlled area.
Fig 2. Installation of an external vent valve: (A) During the test. (B) After the test.
Installation of a hydrophobic filter
A hydrophobic filter can be installed between the external vent valve and the product filter to avoid potential contamination of the environment with bacteria or viruses (Fig 3). This is relevant for testing after filtration where such contamination might be on the upstream side of the filter or for product-wet pre-use testing. Such a protective filter would also retain liquid contaminants with high surface tension. However, contaminants with low surface tension could wet, block, or even pass such a filter. A second benefit of the installation of a hydrophobic filter is that the upstream side of the product filter is also protected from possible contamination originating from the integrity test system or the pressurized air supply respectively. Such protection may be desirable for pre-use integrity testing.
Fig 3. Installation of a filter between the instrument and the filter under test.
Avoidance of venting in the direction of the integrity test instrument
The preferred method to reduce the risk, especially if the filter housing can contain chemical or biological hazardous components, is to avoid venting of the filter at the end of the filter test in the direction of the integrity test instrument. To accomplish this objective an external 2-way valve can be installed between the test unit and the filter housing which is controlled by the integrity test instrument and closes the connection between the instrument and the filter housing before the automatic venting process is started by the test instrument (Fig 4). The venting of the filter housing can be performed under controlled conditions, for example into a kill tank.
Fig 4. Installation to avoid automatic venting after the filter integrity test: (A) During the test. (B) After the test.
Alternatively, a pneumatic valve can be installed on the filter housing which is closing at the end of the test controlled by the test instrument (Fig 5). Such an installation can effectively further reduce the risk of cross-contamination. Also, in this case, venting is performed, for example, over a kill tank.
Fig 5. Alternative installation to avoid automatic venting and reduce the risk of contamination: (A) During the test. (B) After the test.
In these cases, as shown in these figures, the filter housing remains pressurized at the end of the integrity test and must be vented before further use. However, this should be viable for high-risk filtration systems, as the capability of controlled venting is required in any case for the filtration of hazardous products.