ÄKTA ready™ 450
Compact system for small-scale manufacturing with a low hold-up volume and flow rate, and a single-use flow path with a fully closed option.
ÄKTA ready™ 450 Compact system for small-scale manufacturing with a low hold-up volume and flow rate |
ÄKTA™ ready Robust single-use chromatography system with isocratic or gradient capability for process scale-up and manufacturing |
ÄKTA ready™ variable UV Single-use system with a three-wavelength UV sensor and a single-use pH sensor. Supports processing of viral vectors and nucleic acids |
ÄKTA ready™ extended Comprehensive single-use chromatography system providing flexibility in process scale-up and manufacturing applications |
ÄKTA ready™ XL High-capacity system to support your purification needs in manufacturing |
|
Footprint (W × H × D) | 640 × 940 × 558 mm | 1000 × 1650 × 800 mm (isocratic) and 1160 × 1650 × 800 mm (gradient) |
1140 × 1670 × 840 mm | 1060 × 1760 × 1000 mm | 1295 × 1920 × 1215 mm |
Weight | 110 kg | 230 kg (isocratic) and 250 kg (gradient) | 258 kg | 360 kg | 685 kg |
Operating max. pressure | 4 bar | ||||
Inlets | 8+4 | 6+1 | 6+6 | 6+4 | |
Outlets | 6 | ||||
UV | Simultaneous, 3-wave lengths | 280 nm | Simultaneous, 3-wave lengths | ||
pH | Single use, integrated | Optional to use traditional electrode | Single use, integrated | ||
Flow kit flow range | 5-450 mL/min | High flow kit: 7.5–510 L/h Low flow kit: 3–175 L/h |
¾ inch i.d.: Range: 45–1900 L/h 1-inch i.d. Range: 90-3500 L/h |
||
Flow rate accuracy | ± 6% of actual value, or ±12 mL/min, whichever is greater | High flow kit: ± 5% actual value at flow‡ ≥ 40 L/h ± 10% actual value at flow < 40 L/h Low flow kit: ± 5% actual value at flow‡ ≥ 20 L/h ± 10% actual value at flow < 20 L/h |
¾ inch i.d.: Accuracy, ±5 L/h or 2% whichever is greater 1-inch i.d.: Accuracy, ±5 L/h or 2% whichever is greater |
||
Connectors | Tri‑Clamp™ and AseptiQuik® | Tri‑Clamp™, ReadyMate™ and AseptiQuik® | Tri‑Clamp™ and AseptiQuik®. ReadyMate™ available on request | Tri‑Clamp™, ReadyMate™ and AseptiQuik® | Tri‑Clamp™ and AseptiQuik® |
BioPhorum (BPOG) Extractables Protocol | Yes |
||||
Pressure sensor | Range: 0.1–4 bar Accuracy: ± 0.1 bar, or ± 5% of actual value |
Range: 0–5 bar g Accuracy: ± 0.20 bar g |
Range: 0.1–4 bar Accuracy: ± 0.1 bar, or ± 5% of actual value |
||
Temperature sensor | Range: 2°C to 30°C. Accuracy: ± 1°C |
Range: 2°C to 40°C Accuracy: ± 3°C |
Range: 2°C to 30°C. Accuracy: ± 2°C |
||
Conductivity sensor | Range: 1–200 mS/cm Accuracy: ± 5% of actual value |
Range: 0–150 mS/cm Accuracy: ± 5% full scale |
Range: 1–200 mS/cm Accuracy: ± 0.5 mS/cm, or ± 7% of actual value |
||
UV sensor | Range: 0–2 AU* Accuracy: linearity ± 2% |
Range: 0.01–1.0 AU Accuracy: linearity ± 5% |
Range: 0–2 AU. Accuracy: linearity ± 2% |
||
pH sensor | Range: 3–10 Accuracy: ± 0.3 |
Range: pH 2–12 Accuracy: ± 0.2 pH at calibration temperature |
Range: 3–10 Accuracy: ± 0.3 |
||
Material compliance | USP <88> Class VI Test for biocompatibility Animal free or complies with the conditions in EMA/410/01 |
||||
Human-Machine Interface (HMI) | Stand alone |
Integrated with the system | |||
Control software | UNICORN™ 7.0.2 or higher | UNICORN™ 7.0.2 or higher, or Figurate powered with Emerson DeltaV™ |
* Valid for 250–700 nm, 2% linearity 0–1 AU for 206–250 nm.
‡ Valid for liquid with density of water and 1.0 to 1.3-fold the viscosity of water at the actual liquid temperature; and max temperature difference of 10°C between liquid and ambient temperature.