Preventive maintenance supports performance and compliance

Scheduled maintenance tunes your instrument to deliver high-quality products with PM procedures that are documented in a service report.

INCLUDED TESTS
Extensive set of 40 tests and inspections
Valve manifold Mass flow controller test
Pressure verification Pump speed test
Vessel temperature verification Filter temperature test
Condenser temperature test Agitator alignment verification and speed test
pH signal test and DO signal test Vessel weight test
UPS test Emergency stop test
Light stack test Cabinet air filter test (n/a for XDR 10)
Mechanical inspections