FAQ

Flow path schemes

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

1. Flushed loop injection
With flushed loop injection means that the sample loop is completely filled with sample.



2. Partial loop injection
With partial loop injection means that the loop is partially filled with sample.


3. ul Pick-up injection
With ul Pick-up injection means that after aspiration from the vial, the sample volume
is transported into the loop with transport liquid (mobile phase) from another vial.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

For information of each component, place the cursor over the component of interest.

 

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Spare parts

Figure. ÄKTAmicro, front view.

Please note that item numbers are not necessarily in sequential order.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 Tubing Kit for Rinsing System 18111332 432.00 USD Add to cart
1 O-Ring, 1 × 3 mm, PFR 19003601 89.07 USD Add to cart
1 Check valve kit 28924679 907.00 USD Add to cart
1 Piston Kit 28924680 477.00 USD Add to cart
1 Seal kit 28924689 580.00 USD Add to cart
1 Purge valve 28924692 242.00 USD Add to cart
2 Fiber Detachment Tool 18111116 131.81 USD Add to cart
3 Monitor pH/C-900 18110776 11,500.00 USD Add to cart
4 Mounting Bracket for Valves 18110911 35.21 USD Add to cart
4 Injection Kit, INV-907 18111089 630.00 USD Add to cart
4 Fill Port, INV-907 18112766 266.00 USD Add to cart
4 Valve kit INV-917 18114756 790.00 USD Add to cart
5 Stop Plug 1/16" Male 18111252 59.00 USD Add to cart
5 Mixer M-925 for classic ÄKTA 18111889 2,698.00 USD Add to cart
6 On-line filter 18111801 723.00 USD Add to cart
6 On-line filter kit 18112094 137.66 USD Add to cart
7 Valve SV-903 18111449 788.00 USD Add to cart
8 Mixer M-925 Mixing Chamber, 0.6 ml 18111890 1,016.00 USD Add to cart
8 Mixer M-925 Mixing Chamber, 2 ml 18111891 590.00 USD Add to cart
8 Mixer M-925 Mixing Chamber, 90 µl 18114724 968.00 USD Add to cart
10 Conductivity Flow Cell, 0.2 µl 18114720 2,108.00 USD Add to cart
Cable UniNet, L = 0.7 m 18110974 298.00 USD Add to cart
Signal Cable, (6-pin miniDin-Open) 18111064 395.00 USD Add to cart

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 Connector 3/16" Male for Tubing o.d. 3/16" 18111249 314.00 USD Add to cart
2 Ferrules for 3/16" o.d. tubing connector 18111248 130.53 USD Add to cart
3 Stop Plug 5/16" Male 18111250 36.22 USD Add to cart
4 Connector 1/16" Male/Luer Female 18111251 109.24 USD Add to cart
5 Stop Plug 1/16" Male 18111252 59.00 USD Add to cart
6 Union 1/16" Female/M6 Male 18111257 128.73 USD Add to cart
7 Fingertight Union 1/16" Male/M6 Female 18111258 135.19 USD Add to cart
8 Connector for 1/8" tubing 18112117 199.76 USD Add to cart
9 Ferrule for o.d. 1/8" Tubing 18112118 145.94 USD Add to cart
10 Fingertight Connector 1/16" Male for Tubing o.d. 1/16" 18111255 123.16 USD Add to cart
11 Union 1/16" Male - 1/16" Male, i.d. 0.25 mm 18112092 38.29 USD Add to cart
12 Ferrules for o.d. 1/16" tubing connector 18112706 159.39 USD Add to cart
13 Connector 1/16" Male 18112707 316.00 USD Add to cart
14 Union 5/16" female - 1/16" male 18114208 56.92 USD Add to cart
15 Union 1/16" female-1/16" female 18385501 105.84 USD Add to cart
16 Tubing Connector Flangeless/M6 Male 18101798 79.78 USD Add to cart
17 Fingertight Narrow Head Connector, o.d. 1/16" 18114710 538.00 USD Add to cart
21 Union 1/16" Male - 1/16" Male, i.d. 0.13 mm 18112090 134.55 USD Add to cart

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 Inlet filter holder kit 11000407 167.67 USD Add to cart
2 Screw lid kit 11000410 49.68 USD Add to cart
3 Inlet filter set 11000414 109.71 USD Add to cart
4 Inlet Filter Assembly Kit 18111315 190.44 USD Add to cart
5 Inlet Filter Set 18111442 207.00 USD Add to cart
6 Inlet manifold 18112080 156.73 USD Add to cart
7 Flow Restrictor FR-904 18111963 428.00 USD Add to cart
8 Cable UniNet, L = 0.18 m 18110972 144.58 USD Add to cart
9 Cable UniNet, L = 0.3 m 18110973 158.94 USD Add to cart
10 Cable UniNet, L = 0.7 m 18110974 298.00 USD Add to cart
11 Cable UniNet, L = 1.5 m 18111775 262.00 USD Add to cart
12 Cable UniNet, L = 3 m 18110975 386.00 USD Add to cart
13 Cable UniNet L=15 m 18111774 524.00 USD Add to cart
14 Mains cable, 120 V 19244701 71.49 USD Add to cart
15 Mains cable 220 V 19244801 110.80 USD Add to cart
16 Signal Cable, (6-pin miniDin-Open) 18111064 395.00 USD Add to cart
17 Short column holder, for columns 10 to 50 mm o.d. 18111317 220.00 USD Add to cart
18 Small Column Clamp 18114998 50.53 USD Add to cart
19 Lab Rod Holder 18111319 490.00 USD Add to cart
20 Tubing cutter, for PEEK, EFTE, and FEP tubing i.d. 0.25, 0.5, 0.75, 1 and 1.6 mm 18111246 99.55 USD Add to cart

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 Tubing cutter, for PEEK, EFTE, and FEP tubing i.d. 0.25, 0.5, 0.75, 1 and 1.6 mm 18111246 99.55 USD Add to cart
2 Connector 1/16" Male/Luer Female 18111251 109.24 USD Add to cart
3 Union 1/16" Male - 1/16" Male, i.d. 0.13 mm 18112090 134.55 USD Add to cart
4 Union 1/16" Male - 1/16" Male, i.d. 0.25 mm 18112092 38.29 USD Add to cart
5 Fingertight Narrow Head Connector, o.d. 1/16" 18114710 538.00 USD Add to cart
8 Stop Plug 1/16" Male 18111252 59.00 USD Add to cart
9 Tubing Kit, i.d. 0.10 mm 18114713 954.00 USD Add to cart
11 PEEK Tubing, 2 m, i.d. 0.15 mm, o.d.1/16" 18115659 446.00 USD Add to cart
12 Microfraction Collection Kit, ÄKTAmicro 28948780 545.00 USD Add to cart
13 Tubing i.d. 0.25 mm, o.d. 1/16" 18112095 66.24 USD Add to cart
14 PEEK Tubing, 2 m, i.d. 0.5 mm, o.d. 1/16" 18111368 73.48 USD Add to cart
17 New Purge Kit P-900 18112453 91.08 USD Add to cart
18 Inlet filter holder kit 11000407 167.67 USD Add to cart
19 Inlet filter set 11000414 109.71 USD Add to cart
20 On-line filter kit 18112094 137.66 USD Add to cart
21 Flow Restrictor FR-904 18111963 428.00 USD Add to cart
22 Mixer M-925 Mixing Chamber, 90 µl 18114724 968.00 USD Add to cart
23 Fiber Detachment Tool 18111116 131.81 USD Add to cart
24 Mains cable, 120 V 19244701 71.49 USD Add to cart
25 Mains cable 220 V 19244801 110.80 USD Add to cart
26 Cable UniNet, L = 3 m 18110975 386.00 USD Add to cart

Seals are considered as consumables and should be replaced annually or when needed. Please keep a set of each in your stock to prevent long downtime. For code numbers refer to the table below.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 MDLC Pump head 28923066 3,473.00 USD Add to cart
2 Inlet filter set 11000414 109.71 USD Add to cart
3 On-line filter kit 18112094 137.66 USD Add to cart
4 Flow Restrictor FR-904 18111963 428.00 USD Add to cart
5 Mixer M-925 Mixing Chamber, 0.6 ml 18111890 1,016.00 USD Add to cart
6 Mixer M-925 Mixing Chamber, 2 ml 18111891 590.00 USD Add to cart
7 Mixer M-925 Mixing Chamber, 90 µl 18114724 968.00 USD Add to cart
9 Valve SV-903 18111449 788.00 USD Add to cart
10 Valve kit INV-917 18114756 790.00 USD Add to cart
11 Check valve kit 28924679 907.00 USD Add to cart
12 Piston Kit 28924680 477.00 USD Add to cart
13 Seal kit 28924689 580.00 USD Add to cart
14 Purge valve 28924692 242.00 USD Add to cart

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 Inlet tubing kit, i.d. 1.6 mm, o.d. 3/16 in 18111926 161.46 USD Add to cart
2 FEP Tubing, i.d. 1/8 in (3/16 in) 18111247 107.23 USD Add to cart
3 Tubing i.d. 0.25 mm, o.d. 1/16" 18112095 66.24 USD Add to cart
4 ETFE Tubing, 2 m, i.d. 0.5 mm, o.d. 1/16 in 18112096 61.00 USD Add to cart
5 PEEK Tubing, 2 m, i.d. 0.75 mm, o.d. 1/16" 18111253 96.26 USD Add to cart
6 PEEK Tubing, 2 m, i.d. 0.5 mm, o.d. 1/16" 18111368 73.48 USD Add to cart
7 PEEK Tubing, 2 m, i.d. 1.0 mm, o.d. 1/16" 18111583 85.34 USD Add to cart
8 ETFE Tubing, 1.5 m, i.d. 0.25 mm, o.d. 1/16" 18112136 254.00 USD Add to cart
9 FEP Tubing, 3 m, i.d. 1/16", o.d. 1/8" 18112116 73.48 USD Add to cart
10 ETFE Tubing, 1.5 m, i.d. 0.75 mm, o.d. 1/16" 18111974 80.73 USD Add to cart
11 PEEK Tubing, 2 m, i.d. 0.15 mm, o.d.1/16" 18115659 446.00 USD Add to cart
12 Capillary tubing i.d. 0.35 mm 18114735 1,234.00 USD Add to cart
14 Tubing Kit, i.d. 0.10 mm 18114713 954.00 USD Add to cart
15 PEEK Tubing, 2 m, i.d. 0.15 mm, o.d.1/16" 18115659 446.00 USD Add to cart

Accessories

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
3 Short column holder, for columns 10 to 50 mm o.d. 18111317 220.00 USD Add to cart
4 Small Column Clamp 18114998 50.53 USD Add to cart
5 Lab Rod Holder 18111319 490.00 USD Add to cart
13 Microfraction Collection Kit, ÄKTAmicro 28948780 545.00 USD Add to cart

Troubleshooting

Find solutions to product related issues. For unlisted issues please contact local Cytiva service representation.

UV curve issues

Issues related to UV curve

Conductivity curve

Issues related to Conductivity curve

Pressure curve issues

Issues related to Pressure curve

Valve SV-903 unusual appearance

Issues related to Valve SV-903 unusual appearance

Monitor pH/C-900 unusual appearance

Issues related to Monitor pH/C-900 unusual appearance

General advice to achieve good performance

Before using the system make sure that:

  • The correct system has been selected in UNICORN System Control
  • The correct wavelength has been set for UV/UPC monitor
  • All tubing has been properly connected
  • All connectors are free from leakage
  • No tubing is folded or twisted
  • Online filter, if used, is changed on a regular basis
  • Correct buffers are used for the chosen columns and proteins
  • All inlet tubing has been immersed in correct buffer solutions
  • Enough buffer has been prepared
  • Buffers have been equilibrated to the environment temperature
  • Buffers/eluents have been degassed if necessary (e.g., in RPC runs)
  • Suitable columns have been selected for the target proteins
  • Columns have been cleaned and prepared according to column instructions
  • Samples have been clarified by centrifugation and/or filtration prior to sample loading
  • Samples have been adjusted to binding buffer conditions
  • Auto sampler (if used) has been prepared according to user manual
  • The fraction collector has been filled with appropriate number of microtiter plates or tubes
  • Appropriate arrangement for waste handling has been prepared

Valve INV-917 unusual appearance

Issues related to INV-917 unusual appearance

Pump P-905 unusual appearance

Issues related to Pump P-905 unusual appearance

Mixer unusual appearance

Issues related to Mixer unusual appearance

Monitor UV-900 unusual appearance

Issues related to Monitor UV-900 unusual appearance

Maintenance

Maintenance instructions and procedures.

ÄKTAmicro

Periodic maintenance

Regular maintenance is important for safe and trouble-free operation of your instrument. The user should perform daily and monthly maintenance. Preventive maintenance should be performed on a yearly basis by qualified service personnel.

For maintenance of a specific component, carefully read the component manual and follow the instructions. To avoid personal injury when performing maintenance on the ÄKTAmicro instrument, follow the instructions below.

Warning

Electrical shock hazard. All repairs should be done by service personnel authorized by Cytiva. Do not open any covers or replace parts unless specifically stated in the user documentation.

Warning

Disconnect power. Always disconnect power from the instrument before replacing any component on the instrument, unless stated otherwise in the user documentation.

Warning

Hazardous chemicals during maintenance. When using hazardous chemicals for system or column cleaning, wash the system or columns with a neutral solution in the last phase or step.

Warning

Do not perform any type of maintenance work while the system is powered electrically or when the piping system is pressurized. Note that the piping system can be pressurized even when the system is closed down.

Notice

When using hazardous chemicals, take all suitable protective measures, such as wearing protective glasses and gloves resistant to the chemicals used. Follow local regulations and instructions for safe operation and maintenance of the system.

Notice

Cleaning. Keep the instrument dry and clean. Wipe regularly with a soft damp tissue and, if necessary, a mild cleaning agent. Let the instrument dry completely before use.

Warning

Make sure that the piping system is completely leakage free before performing any CIP on the system.

Warning

NaOH is corrosive and therefore dangerous to health. When using hazardous chemicals, avoid spillage and wear protective glasses and other suitable personal protective equipment.

Warning

Before disassembly, check that there is no pressure in the piping system.

Warning

After assembly, the piping system must be tested for leakage at maximum pressure for continued protection against injury risks due to fluid jets, burst pipes or explosive atmosphere.

Warning

Disconnect power. Always disconnect power from the instrument before replacing fuses.

Warning

For continued protection from fire hazard, replace only with same type and rating fuse.

Warning

Incorrectly fitted tubing may loosen, causing a jet of liquid to spray out. This is especially dangerous if hazardous chemicals are in use. Connect the tubing by first inserting the tubing fully, then tightening the connector fingertight. PEEK nuts should be tightened a further 1/3-1/2 of a turn using the key supplied. Metal nuts should be tightened a further 1/4-1/3 a turn using the key supplied.

Caution

Handle the check valves with care when they have been removed from the pump heads to prevent loss of any internal components.

Caution

To protect the piston seals, the pump must never be run with air in the inlet tubing. Follow the procedure below to remove the air.

Caution

To protect the pump seals, always ensure that there is a constant supply of eluent. The pump should never be allowed to run dry.

Caution

To prevent precipitation of crystals when changing from a salt-containing buffer to organic solvent, always flush through the system with water as the intermediate liquid.

Recycle

This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of equipment.

Maintenance operations should be performed by the user at regular intervals

System
  • Inspect the complete system for eluent leakage.
  • The system can be left filled with buffer overnight. If you are not using the separation unit for a few days, wash the flow path with distilled water. Replace the column by a bypass capillary. Then wash the system with 20% ethanol and store it in 20% ethanol. Make sure that all tubing and all flow paths used are rinsed.
Pump P-905
  • Check for leakage. If there is a sign of liquid leaking between the pump head and the housing side panel or increased or decreased volume of rinsing solution, replace the pumphead.
  • When changing eluent, it is important to remove trapped air. Run the Purge the pump procedure below. If there is still air in the inlet tubing, stop and run the Removing trapped air bubbles from the pump procedure below.

 

Inlet filters

Check the inlet filters visually and replace them if necessary.

On-line filter

Replace the on-line filter.

Pump rinsing solution
  • Change rinsing solution. Always use 20% ethanol as rinsing solution.
  • If the volume of rinsing solution in the storage bottle has increased, it can be an indication of internal pump leakage. Replace the pumphead.
  • If the volume of rinsing solution in the storage bottle has decreased significantly, check if the rinsing system connectors are mounted properly.
  • If the rinsing system connectors are not leaking, the rinsing membranes or piston seals may be leaking. Replace the pumphead.

 

System

Run the Cleaning the system procedure.

Monitor UV-900

Run the Checking the instrument procedure below.

Monitor UV-900

Run the Cleaning the flow cell and optical fiber connectors procedure below.

Monitor pH/C-900

Run the Cleaninging the conductivity flow cell procedure below. Cleaning the flow cell may be required more often if crude samples are regularly used.

Mixer M-925

Check that the mixer chamber is clean and without damage. Check the tubing connectors. Replace if required.

Valve INV-917

Check for external and/or internal leakage. Run the Changing channel plate and distribution plate procedure below yearly or when required.

Mixer M-925

Replace the complete mixing chamber on a regular basis.

Pump P-905

Replace the pumphead.

Monitor pH/C-900

Run the Cleaninging the conductivity flow cell procedure below. Cleaning the flow cell may be required more often if crude samples are regularly used.

Purge the pump

1
Fill a reservoir flask with distilled water. Immerse the inlet tubings of both pump modules, with filters, in the water.

Note: Never place the reservoir flask below the level of the pump inlet.

2
Connect a male Luer syringe of about 30 ml to the open end of the purge tubing.

3
Connect the male Luer connector at the other end of the purge tubing to the left purge valve at pump module module A.

4
Turn the purge valve counterclockwise half a turn to open it and slowly draw eluent to the syringe.

5
When fluid starts to enter the syringe continue to draw a few milliliters before closing the purge valve. Check that there is no air left in the inlet tubing.

6
Repeat steps 3 to 5 for pump module B, if fitted.

Removing trapped air bubbles from the pump

During routine operation, the presence of air bubbles in the pump heads is seen as an erratic flow, a noisy detector signal or an irregular pressure recording. With air bubbles trapped in the pump, the pressure reading on the display will fluctuate considerably.

To remove the air from the pump:

1.
If the air has accumulated because of a leaking connector, correct the fault.

2.
Ensure that the inlet filter is fully immersed in eluent.

3.
Run the Purge the pump procedure.

If the problem remains, try to remove the air bubbles in accordance with the method described below. Use 100% methanol.

1.
Run at 2 ml/min.

2.
Continue to run for 10 minutes, or until the air bubble is removed.

3.
Run the Changing eluent procedure.

Changing eluent

When changing from one eluent to another, it is extremely important that the two eluents are totally miscible with one another. If the two eluents are immiscible, the pump should be flushed first with an intermediate liquid, which is miscible with both eluents. Failure to do this will cause a wrong flow of eluent from the pump.

When changing from a salt-containing buffer to an organic solvent, use water as the intermediate liquid to prevent precipitation.

1
Stop the pump by setting it in End mode.

2
Transfer the inlet tubing into the new eluent or into the intermediate liquid.

3
Run the pump at a flow rate and time as specified in the table.

Pump
Flow rate
Time
P-905
2 ml/min
20 minutes

4
Stop the pump. If an intermediate liquid is being used, transfer the inlet tubing into the final eluent and repeat step 3 with the new eluent.

In UNICORN, select instruction PumpWash in System Control:Manual:Pump.

Cleaning the system

Clean the flow path every month or when problems such as ghost peaks occur.

1
Disconnect the column and replace it with the G4 tubing.

2
Place all the inlet tubing in 100% acetonitrile.

3
Manually, perform PumpWash for all the four inlet tubings.

4
Flush the whole system with 100% acetonitrile for 5 minutes (1 ml/min).

5
Immediately, repeat step 3 and 4 with distilled water to rinse the flow path from acetonitrile.

Checking the instrument

1
Select menu Check, press OK.

2
Select menu Check Lamp Intensity.

Check lamp Intensity
210nm 85%
300nm 95%

If the lamp intensity is < 20%, contact your local GE Healthcare service representative for lamp replacement or change of internal optical fiber.

1
Select menu Check, press OK.

2
Select menu Check Lamp Run Time.

Check Lamp Run Time 200h

If the lamp on time is > 4000 hours, contact your local GE Healthcare service representative for lamp replacement.

The flip time is the time the monochromator stepper motor has been in operation.

1
Select menu Check, press OK.

2
Select menu Check Flip Time.

Check Flip Time 50h

If the flip time is > 2000 hours, contact your local GE Healthcare service representative for maintenance.

Cleaning the flow cell and optical fiber connectors

Cleaning the flow cell

A clean flow cell and optical connectors are essential for ensuring the correct operation of the UV-monitor.

1
Connect a syringe to the inlet of the flow cell and squirt distilled water through the cell in small amounts. Then fill the syringe with a 10% surface active detergent solution like Decon 90, Deconex 11, RBS 25 or equivalent, and squirt five times.

2
After five squirts, leave the detergent solution in the flow cell for at least 20 minutes.

3
Pump the remaining detergent solution through the flow cell.

4
Rinse the syringe and flush the cell with distilled water (10 ml).

Cleaning the optical fiber connectors

Wipe with 30% isopropanol on lens paper.

Instrument housing

Wipe the instrument housing regularly with a damp cloth. Let the instrument dry completely before use.

Cleaning the conductivity flow cell

If the conductivity measurements are not comparable to previous results, the electrodes in the flow cell may be contaminated and require cleaning.

To clean the flow cell:

1
Pump 15 ml of 1 M NaOH at 1 ml/min through the flow cell either by using a pump or a syringe.

2
Leave it for 15 minutes.

3
Rinse thoroughly with 50 ml de-ionised water.

Note: If the flow cell is totally blocked, the blockage can be broken using a thin needle or a piece of string with a diameter less than 0.3 mm.

Changing channel plate and distribution plate

A replacement kit, Valve Kit INV-907, is available, see Spare parts recommended to keep on site in the Spare Parts section under the Related Products tab for code no.

1
Ensure that the valve is in position 1 and then disconnect it from the pump.

2
Remove the 4 screws on the front using the supplied 3 mm Allen key. Loosen each one equally in turn so the distribution plate comes off parallel to the valve body.

3
Slide the screws out.

4
Remove the distribution plate containing the ports.

5
Remove the old channel plate and insert a new one.

6
Remount a new distribution plate so that the text 3 is horizontal and to the right of the central tubing connection. Using the Allen key, tighten the 4 screws in turn, a little at a time, until the distribution plate is fixed to the valve body.

Replacing the piston seal, membranes and piston

If there are signs of liquid leaking between the pump head and the housing side panel or the volume of the rinsing solution has increased or decreased, replace the piston seal of the leaking pump head.

CAUTION! Do not disassemble the pump head unless there is good reason to believe that the seal is leaking. Always ensure that sufficient spare components are available before attempting to replace the piston seal. It is not possible to reinstall a used piston seal after removal.

Note: The power must be switched OFF when removing and refitting the pump heads.

Note: Always replace the piston seals on both pump heads at the same time. An even better practice is to replace all four piston seals.

Spare parts and tools required:
Seal kit containing:
2 or 4 piston seals
2 or 4 rinse membranes
1/4 inch wrench (supplied with the pump)
3 mm allen key (supplied with the pump)
Screwdriver (supplied with the pump)

Note: After a new seal has been installed, the pump should be run in. Run the Running-in a new piston seal procedure.

Note: Before disassembling the pump heads move all input buffers bottles below the level of the pump heads to prevent siphoning.

CAUTION! Read the following instructions carefully. The individual parts of the pump head can be assembled incorrectly. Take care to ensure that the orientation of each part is correct before continuing with the next instruction.

1
Switch off the pump at the mains power switch on the back panel.

2
Remove the piston seal rinsing system. The connectors are simple plugin fittings.

3
Completely loosen the tubing connector on the outlet valve.

4
Remove the Switch valves and disconnect the tubing. Unscrew the two white knurled screws under the pump to release the support bracket locking the inlet manifold into the inlet valve. Gently lower and remove the complete manifold.

5
Using the Allen key, unscrew and completely remove one of the two Allen screws locking the pump head in position.

6
When unscrewing the second locking screw, push firmly on the front face of the pump head to compensate for the pressure of the piston return spring. Hold the pump head firmly to prevent it from twisting. Remove the second screw and, without allowing the pump head to twist sideways, carefully pull it out.

7
Place the pump head face down on the bench. Pull out the piston together with the return spring.

8
Inspect the piston and return spring for sign of damage. If damaged, they should be replaced.

9
Wipe the piston with a clean cloth. If salt solutions have been used the piston may be slightly corroded. This corrosion can be removed with a rubber eraser. If it cannot be wiped or rubbed clean, scrape off any deposits with a scalpel or razor blade. Inspect the piston with a magnifying glass for scratches. Replace with a new piston if any scratches or cracks are found.

10
Remove the two screws securing the drain plate and the rinse chamber. Remove and discard the rinsing membrane. Remove the rinse chamber. Remove also the support washer.

11
Gently withdraw the piston seal. Discard the used seal.

12
The pump head, rinse chamber and drain plate should be carefully rinsed or cleaned in an ultrasonic bath, if available. If dirt can be seen on any surfaces, the inlet and outlet check valves should be removed and cleaned separately. Run the Cleaning the inlet and outlet check valves procedure.

13
Wet the new seal slightly and place it in the hole on the pump head and press it down into position with a hard flat object. Refit the support washer on top of the new seal.

14
With the pump head still facing downwards on the bench, place the rinse chamber onto the head with the rinse ports in line with the inlet and outlet check-valves. The conical depression in the rinse chamber should be facing upwards, ready to accept the new rinsing membrane. Fit the rinsing membrane with the conical face downwards.

15
Place the drain plate on top of the assembly. Use the two screws to lock the complete assembly together.

Note: Align the drainage hole in the drainage plate with the inlet check valve (the opposite side of the pump head marked OUT/UP).

16
Wipe clean the piston and remove all finger prints. Wet the piston and then insert it into the return spring. With the pump head facing downwards on the bench, insert the piston into the pump head by pushing it gently but firmly vertically downwards into the seal.

CAUTION! Do not push the piston at an angle to the head and DO NOT twist the piston.

17
Turn the head so that the inlet valve and drainage hole are facing downwards and the text UP/OUT on the pump head is facing upwards. Mount the complete pump head over the locating pins on the front panel. Locate the metal end of the piston and the spring towards the drive cam.

Hold the pump head firmly against the side panel of the housing with one hand. Do not allow the assembly to twist under pressure from the return spring. Using the Allen key, fit and tighten one of the Allen screws.

Fit and tighten the remaining Allen screw.

18
Reconnect the outlet tubing to the outlet check valve and the manifold block as described above.

19
Reconnect the inlet manifold and, if used, the switch valves.

20
Refit the tubing of the piston seal rinse system.

21
The pump should now be purged and the new piston seal carefully run the Running-in a new piston seal procedure.

Running-in a new piston seal

The piston seal should be run-in using 100% methanol.

1
Ensure that the reservoir is filled with sufficient eluent. Immerse the inlet tubing in the eluent. The reservoir should be placed at least 30 cm above the pump inlet.

2
Connect a male Luer syringe of about 30 ml to the open end of the purge tubing.

3
Connect the male Luer connector at the other end of the purge tubing to the left purge valve at pump module module A.

4
Turn the purge valve counter-clockwise half a turn to open it and slowly draw eluent to the syringe.

5
When fluid starts to enter the syringe continue to draw a few milliliters before closing the purge valve. Check that there is no air left in the inlet tubing.

6
Repeat steps 3 to 5 for pump module B, if fitted.

7
Check that the outlet tubings are not blocked.

8
Connect a thin capillary or a column that will give sufficient back pressure.

9
Run the pump at 1 ml/min (or 2 ml/min at 50%B) at a back-pressure of 5-10 MPa for 2 hours, or longer if possible (e.g. overnight).

10
Finally, Change the eluent according to the procedure given below.

Replacing a damaged piston

Typical symptoms of a damaged piston are observed as excessive piston seal wear, unstable pressure, a reduction in the flow or, in som cases, noise as the piston moves. The piston should be removed, examined for damage or salt precipitation and the replaced with a new piston if necessary.

If a damaged piston has been in operation, the piston seal will be destroyed and should be replaced. Run the Replacing the piston seal, membranes and piston procedure.

Spare parts and tools required:

  • Seal kit
  • 1/4 inch wrench
  • 3 mm Allen key
  • Screwdriver
  • Piston kit

Cleaning the inlet and outlet check valves

Faulty operation of the check valves is usually indicated by irregular flow, very low flow or unstable pressure traces. Probable causes of this are air or dirt in a check valve preventing it from closing to seal and hold the pressure.

Record the pressure according to instructions, checking the pump pressure at the end of this document and identify the faulty check valve by observing which pump head is delivering the flow according to check menu at the end of this document.

Try to clean the check valves in–place on the pump head by pumping distilled water at 2 ml/min for 2 minutes. This also prevents precipitation of crystals. The pump 100% methanol for approximately 10 minutes. If this does not correct the problem, follow the instructions for removing and then cleaning the valves.

Note: Change solvent to distilled water and flush out all salt before removing the check valves.

Tools required: A 13 mm, a 18 mm and a 1/4 inch wrench.

Before removing the check valves, move all input buffers bottles below the level of the pump heads, to prevent siphoning.

1
If the condition of the check valve is not improved by in–place cleaning, disconnect and remove the inlet manifold and outlet tubing.

2
Use the 13 mm wrench (inlet check valve) or the 18 mm wrench (outlet check valves) to remove the valve from the pump head.

3
Immerse the complete valve in methanol and place in an ultrasonic bath for some minutes.

Then repeat the ultrasonic bath with distilled water.

4
Refit the check valves. The inlet check valve (with a lip for the manifold and a larger diameter opening) is fitted to the side marked IN of the pump head. Tighten the valves until fully finger-tight and then use the wrench to tighten a further 1/3rd (110o) of a turn. Do not overtighten the valves since damage to the internal components can occur.

5
Refit the outlet tubing and the inlet manifold.

6
Purge the pump carefully and check that the pumping action has been corrected.

Note: Check valves have precision matched components and should only be disassembled further by a trained service engineer. If the problem cannot be corrected, the valve should be replaced completely.

To check the pump function, a recording of the pressure can be made or by checking the pressure in UNICORN. This pressure recording is more sensitive than the reading on the display. By observing the piston stroke indicator in the Check menu together with the pressure trace, the pump head which is functioning abnormally can be identified.

There can be several causes of an abnormal pressure recording, for example:

  • air trapped in the pump heads
  • partially blocked solvent filters
  • leaking connections
  • piston seal leakage
  • check valve malfunction
  • piston damaged

Check menu

To enable trouble shooting it is possible to check which pump module head that delivers flow.

1
Select main menu Check, press OK.

2
The display shows the status of the pistons for both pump modules.

A: Left means that the left pump head is delivering flow in the A pump and

B: Right means that the right pump head in delivering flow in the B pump. At the changing point both are displayed. The A pump is closest to the front panel.

Note: 
Check Heads 2.30MPa
A: Left
B: Right

1
Select main menu Check, press OK.

2
Select sub menu Check Total Run Time. The display shows the accumulated operation time (Run and Hold mode) for the A and B pumps.

Note:
Check Total Run Time
A: 50 h
B: 20 h

1
Select main menu Check, press OK.

2
Select sub menu Check Piston Strokes. The display shows the accumulated number of piston strokes for the A and B pumps.

Note:
Check Piston Strokes
11374, 2398

1
Ensure that the pump is at zero pressure.

2
Select main menu Check, press OK.

3
Select sub menu Calib Press Offset, press OK. “Working, please wait ...” is displayed until the calibration is completed.

Service information relevant to the instrument can be checked. Information may not be available in all menus.

1
Select main menu Check, press OK.

2
Select sub menu Check Service Mode, press OK.

3
The service telephone number is displayed1. Turn the selection dial clockwise to select next sub menu.

4
The service contract number is displayed1. Turn the selection dial clockwise to select next sub menu.

5
The instrument serial number is displayed. Turn the selection dial clockwise to select next sub menu.

6
Instrument name and software version are displayed. Turn the selection dial clockwise to select next sub menu.

7
The date of the last service is displayed. Turn the selection dial clockwise to select next sub menu.

8
A test of the instrument buzzer is performed, press OK.

1Not always pre-programmed. Depends on customer-specific contract.

Contact