FAQ
ÄKTAmicro
ÄKTAmicro Installation Guide
ÄKTAmicro System Manual
Getting started
Optional Configurations ÄKTAmicro User Manual
ÄKTAmicro Safety Handbook
ÄKTAmicro Pump P-905 User Manual
ÄKTAmicro Pump P-905 Short Instructions
ÄKTAmicro UV-900 User Manual
ÄKTAmicro UV-900 Short Instructions
ÄKTAmicro Monitor pH/C-900 User Manual
ÄKTAmicro Monitor pH/C-900 Short Instructions
Mixer M-925 Instructions
Valve INV-917 Instructions
Valve SV-903 Instructions
Getting Started with UNICORN
UNICORN User Reference Manual part 1
UNICORN User Reference Manual part 2
UNICORN Administration and Technical Manual
System certificate
Module certificates for
Pump P-905
Monitor UV-900
Monitor pH/C-900
Valve INV-917
How can I CLEAN my system?
ÄKTAmicro
Cleaning the system
Cleaning the instrument
- Wipe the surface regurlarly with a damp cloth. Do not allow spilt liquid to
dry on the instrument. - Remove dirt from the surface using a cloth and a mild cleaning agent or
20% alcohol. - Let the system dry completely before using it.
Cleaning the flow path
The column should be replaced by a bypass tubing when cleaning the flow path. If leaving the column in the flowpath, please observe the rated maximum flow and pressure for the column.
For column cleaning procedures and column storage instructions, please refer to the Instruction supplied with the column.
At the end of the day
If the system will be used with other buffers next day, rinse the pump and the flow path with distilled water using the PumpWash instruction as follows:
- Submerge the inlet tubings in distilled water.
- Run the PumpWash instruction.
Leaving for a few days
Perform a PumpWash with distilled water. Repeat with a bacteriostatic solution, 20% ethanol.
Monthly cleaning
Clean the flow path every month or when problems such as ghost peaks occur.
- Disconnect the column and replace it with the G4 tubing.
- Place all the inlet tubings in 100% acetonitrile.
- Manually, perform PumpWash for all the four inlet tubings.
- Flush the whole system with 100% acetonitrile for 5 minutes (1 ml/min).
- Immediately, repeat step 3 and 4 with distilled water to rinse the flow path from acetonitrile.
Why should I have REGULAR, PLANNED MAINTENANCE on my system?
With the pressure on producing sample or results, the condition of your ÄKTAdesign or Ettan system is critical and regular servicing will mean you can depend on your system to perform as expected. Planned maintenance can be part of a service agreement, scheduled to service your system before it is in need of attention. We can help you design a schedule and routine to allow you to maintain your system, please contact your local Cytiva service representative.
So what can you expect from a planned maintenance visit from Cytiva service representative?
- Thorough inspection and cleaning of system components
- Update of system firmware to ensure full compatibility of your system and UNICORN software
- Replacement of damaged or corroded seals, valve springs and solenoids
- Replacement of items that are reaching the end of their expected life – preventing future breakdowns
- Advice and guidance on proper daily use, cleaning and care of your system
- All work is documented and reported to help make any regulatory audits easier.
A complete overhaul, once a year, ensures that your instrument is running at peak performance so you can be confident of your scientific results. In addition, wear and tear on systems under constant use by multiple end users is minimized, giving the system a longer life and better value for money.
To find out more about service possibilities contact your local Cytiva service representative.
Purifying some samples at room temperature can lead to increased levels of degradation; performing your purifications at 4C can help. All our ÄKTA systems, Ettan LC, Ettan microLC, Ettan nanoLC and fraction collectors are suitable for use in the temperature range of 4-40C.
The computer systems are not however cold room compatible, and can be damaged by being placed at 4C.
When installing an ÄKTA system in a cold room the computer can be positioned up to 15 m away, allowing it to be positioned outside the cold room. Cold cabinets can provide an effective solution to running your ÂKTA system in the cold whilst protecting the PC. When moving a system to or from a cold room, time must be allowed for the system to adjust to its new temperature. You may also find that you need to tighten the connectors on your system slightly to prevent leaks when you bring a system out from the cold, and loosen them slightly before you put a system into the cold to prevent pressure build up.
Changes in temperature can also affect the viscosity of your buffers so keep a close eye on your back pressure.
ÄKTAbasic, ÄKTAexplorer, ÄKTAFPLC, ÄKTAmicro, ÄKTAprime, ÄKTAprime plus, ÄKTApurifier, Ettan LC, Ettan microLC, Ettan nanoLC and Ettan MDLC
Generally: The wetted parts are resistant to organic solvents and salt buffers commonly used in chromatography of biomolecules, except 100% Ethyl acetate, 100% Hexane and 100% Tetrahydrofuran.
Chemical resistance guide and chemical compatibility
The chemical resistance to some of the most commonly used chemicals in liquid chromatography is indicated in the table below.
The ratings are based on the following assumptions:
1. The synergistic effects of the chemical mixtures have not been taken into account.
2. Room temperature and limited over-pressure is assumed.
Note: Chemical influences are time and pressure dependent. Unless otherwise stated, all concentrations are 100%.
Chemical | Exposure < 1 day |
Exposure up to 2 months |
Comments |
Acetaldehyde | OK | OK | |
Acetic acid, < 5% | OK | OK | |
Acetic acid, 70% | OK | OK | |
Acetonitrile | OK | Avoid | FFKM, PP and PE swell. PEEK is affected by long term use |
Acetone, 10% | OK | Avoid | PVDF is affected by long term use |
Ammonia, 30% | OK | OK | Silicone is affected by long term use |
Ammonium chloride | OK | OK | |
Ammonium bicarbonate | OK | OK | |
Ammonium nitrate | OK | OK | |
Ammonium sulphate | OK | OK | |
1-Butanol | OK | OK | |
2-Butanol | OK | OK | |
Citric acid | OK | OK | |
Chloroform | OK | Avoid | ECTFE,CTFE, PP and PE are affected by long term use |
Cyclohexane | OK | OK | |
Detergents | OK | OK | |
Dimethyl sulphoxide | Avoid | Avoid | PVDF is affected by long term use |
1, 4-Dioxane | Avoid | Avoid | ETFE, CTFE, PP, PE and PVDF are affected by long term use |
Ethanol | OK | OK | |
Ethyl acetate | OK | Avoid | Silicone not resistant. Pressure limit for PEEK decreases. |
Ethylene glycol | OK | OK | |
Formic acid | OK | OK | Silicone not resistant |
Glycerol | OK | OK | |
Guanidinium hydrochloride | OK | OK | |
Hexane | OK | Avoid | Silicone not resistant. Pressure limit for PEEK decreases. |
Hydrochloric acid, 0.1 M | OK | OK | Silicone not resistant |
Hydrochloric acid, > 0.1 M | OK | Avoid | Silicone not resistant. Titanium is affected by long term use |
isopropanol | OK | OK | |
Methanol | OK | OK | |
Nitric acid, diluted | OK | Avoid | Silicone not resistant |
Nitric acid, 30% | Avoid | Avoid | Elgiloy is affected by long term use |
Phosphoric acid, 10% | OK | Avoid | Titanium and aluminium oxide are affected by long term use |
Potassium carbonate | OK | OK | |
Potassium chloride | OK | OK | |
Pyridine | Avoid | Avoid | ETFE, PP and PE not resistant |
Sodium acetate | OK | OK | |
Sodium bicarbonate | OK | OK | |
Sodium bisulphate | OK | OK | |
Sodium borate | OK | OK | |
Sodium carbonate | OK | OK | |
Sodium chloride | OK | OK | |
Sodium hydroxide, 2 M | OK | Avoid | Titanium, PVDF and borosilicate glass are affected by long term use |
Sodium sulphate | OK | OK | |
Sulphuric acid, diluted | OK | Avoid | PEEK and titanium are affected by long term use |
Sulphuric acid, medium concentration |
Avoid | Avoid | |
Tetrachloroethylene | Avoid | Avoid | Silicone, PP and PE are not resistant |
Tetrahydrofuran | Avoid | Avoid | Silicone, ETFE, CTFE, PP and PE are not resistant |
Toluene | OK | Avoid | Pressure limit for PEEK decreases |
Trichloroacetic acid, 1% | OK | OK | |
Trifluoroacetic acid, 1% | OK | OK | |
Urea | OK | OK | |
o-Xylene p-Xylene |
OK | Avoid | Silicone, PP and PE are affected by long term use |
We ship the most recent available manual and support documentation at the time
we ship the software to you. Retain this manual as it is the correct one for the version
of software you purchased. Early versions of UNICORN had paper manuals,
but more recent versions have manuals included on CD.
The performance of all UNICORN controlled ÄKTA systems can be checked by using the installation test that is part of each systems strategy. The installation test checks the function of the liquid delivery and UV monitor systems and can be performed at any time to check the condition of the system.
Correct gradient formation is tested by producing a linear and step gradient of acetone and water. The UV monitoring system is tested by monitoring the absorbance of the acetone at 265, 254 and 280 nm.
Each installation test is designed to be run with the system under pressure as this ensures the efficient opening and closing of the check valves. For systems based on the pumps P-901 and P-903 this pressure is provided by internal restriction in the pumps.
For sanitary systems such as the ÄKTApilot™ the required pressure is provided by the column. Therefore, when performing the ÄKTApilot installation test you must provide pressure to the system, either by using a standard FR-902 flow restrictor or by fitting a length of capillary tubing that gives >0.1 MPa of back pressure.
Please read Method example in UNICORN Users Reference Manual.
How long should the UV LAMP on my system Last?
I am getting the error message ‛Error 71: WARNING low light intensity’.
The ÄKTAbasic, ÄKTApurifier, ÄKTAexplorer, ÄKTAmicro and Ettan LC systems contain a
UV-900 monitor, the light source for which is a xenon flash lamp. Xenon lamps
emit a high intensity continuous spectrum of light, your chosen wavelengths are
selected using a monochromator. The lamp is triggered only when needed,
this extends its lifespan, meaning the average lamp can withstand many years
of normal use.
The UV monitoring system of the ÄKTAprime, ÄKTAxpress, ÄKTAFPLC, Ettan
microLC, Ettan nanoLC and the Ettan MDLC uses a Zn lamp for monitoring at 214 nm and a Hg lamp
at all other wavelengths. Wavelengths are selected using a band pass filter. When
the system is operated at room temperature with a wavelength of 254 nm, the average
Hg lamp lifetime is 7000 hours. In the cold room the average lifetime is reduced to
2000 hours.
One of the most common causes for the error message
‛Error 71: WARNING low light intensity’ is not lamp failure but a dirty flow cell.
Therefore, the first thing that you should do if you get this message is give your
system a good clean. If you are using an ÄKTAFPLC, ÄKTAxpress or ÄKTAprime
this error can also be caused by incorrect positioning of the lamp. There are two
positions for aligning the Hg lamp with the filter housing, one for 280 nm
(marked by a filled white circle) and one for all other wavelengths
(marked by a white ring). Having the lamp and filter housings incorrectly aligned
can result in a low light intensity warning.
I want to know the FLOW and PRESSURE RANGE of my system.
System | Height (mm) |
Footprint (mm x mm) |
Weight (kg) |
Flow rate (ml/min) |
Pressure limit (MPa) |
---|---|---|---|---|---|
ÄKTA avant 25 | 660 | 860 x 710 | 116 | 0.001-25 | 20 |
ÄKTA avant 150 | 660 | 860 x 710 | 116 | 0.001-150 (normal range) 0.001-300 (column packing flow) |
5 |
ÄKTA pure | 630 | 535 x 470 | up to 53 kg | 0.001 to 25 (up to 50 during column packing) |
20 |
ÄKTA pure 150 | 630 | 535 x 470 | up to 53 kg | 0.01 to 25 (up to 300 during column packing) |
5 |
ÄKTAexplorer 10 | 620 | 500 x 460 | 75 | 0.001-10 | 25 |
ÄKTAexplorer 100 | 620 | 500 x 460 | 75 | 0.01-100 | 10 |
ÄKTAFPLC | 470 | 380 x 480 | 50 | 0.05-20 | 5 |
ÄKTAmicro | 610 | 480 x 450 | 55 | 0.001-2 | 35 |
ÄKTApilot | 900 | 750 x 540 | 114 | 4-400 (full gradients) 4-800 (limited gradients) |
2 |
ÄKTAprime plus | 530 | 400 x 450 | 13 | 0.1-50 | 1 |
ÄKTApurifier 10 | 620 | 500 x 460 | 75 | 0.001-10 | 25 |
ÄKTApurifier 100 | 620 | 500 x 460 | 75 | 0.01-100 | 10 |
ÄKTAxpress | 660 | 490 x 250 | 30 | 0.1-65 | 3 |
Ettan LC | 610 | 480 x 450 | 55 | 0.001-2 | 35 |
Ettan MDLC | 710 | 700 x 640 | 105 | 0.001-2 | 35 |
Ettan microLC | 1150 | 650 x 500 | 77 | 0.001-2 |
35
|
Ettan nanoLC | 1150 | 650 x 500 | 77 | 0.001-2 | 35 |
How can I determine the delay volume?
Three different methods are described below:
Method I - Determining the delay volume of your system by measuring different retention volumes
1) Check that the pump is delivering the correct flow at 1 ml/min. If the measured flow rate differs the retention volumes need to be corrected. 2) Mount a small loop, e.g. 100 µl, and prepare a solution of 5% acetone in water to use as sample. 3) Fill the system with clean water. Run manually or make your own method. Start the pump at 1 ml/min and inject the sample. The measured retention volume is called Volume 1. 4) Re-configure your system. Dismount the tubing from the UV flow cell and insert a low dead volume connector, e.g. a female-female Valve connector, in the flow path (replaces the UV flow cell). Mount the “tubing end” of the frac outlet into the UV flow cell and mount a waste-tubing to the bottom of the UV flow cell. 5) Set Frac size to a very large volume, e.g. 100 ml, so that the valve is in the Frac position during the entire run. Start the pump at 1 ml/min and inject the sample. The measured retention volume is called Volume 2. Delay volume = Volume 2 – Volume 1. |
Method II - Determining the delay volume of your system with the help of another ÄKTA system
This method is more accurate than the one described above, but demands one more system. 1) Check that the pump on your chosen ÄKTA system is delivering the correct flow at 1 ml/min. If the measured flow rate differs the retention volumes need to be corrected. 2) Mount a small loop, e.g. 100 µl, and prepare a solution of 2-5% acetone to use as sample. 3) Fill the system with clean water. Run manually or make your own method. Start the pump at 1 ml/min and inject the sample. The measured retention volume is called Volume 1. 4) Re-configure your system. In e.g. an ÄKTAprime system dismount the tubing from the UV flow cell to the Fraction collector including the restrictor, the Cond cell and the Frac valve. In the other ÄKTA system, e.g. ÄKTApurifier or ÄKTAexplorer, mount the “prime tubing kit” between the tubing to the UV flow cell and the UV flow cell. To do this a low dead volume connector, e.g. a female-female Valve connector, is needed. 5) In order to have the correct flow path the Frac valve in the “prime tubing kit” must be activated. This can be achieved be connecting it to the Valve B port of the P-900 pump in the ÄKTApurifier or ÄKTAexplorer system. The valve is activated by setting Pump B inlet to the B2 position. 6) Start the pump at 1 ml/min and inject the sample. The measured retention volume is called Volume 2. Delay volume = Volume 2 – Volume 1. |
Method III - Determining the delay volume by balancing eluted water
Manually set the flow path to the direction of the fraction collector. Unscrew the tubing that is connected to inlet of the UV flow cell and insert a luer adaptor instead. Fill a syringe with water and inject water into flow cell unless it drops at the outlet of the fraction collector (in which case you have likely exceeded the pressure in the tubing which might be more than 4 bar, depending on configuration and flow restrictor used). Now fill the syringe with air (at least 20 ml because of compression) and displace the water. Collect eluting water in a small cup. Determine the system delay volume by balancing the cup before and after elution. Repeat two times for calculation of a mean value. Enter the mean value in "system settings" in UNICORN. |
I want to know the DIMENSIONS and WEIGHT of my system.
System | Height (mm) |
Footprint (mm x mm) |
Weight (kg) |
Flow rate (ml/min) |
Pressure limit (MPa) |
---|---|---|---|---|---|
ÄKTA avant 25 | 660 | 860 x 710 | 116 | 0.001-25 | 20 |
ÄKTA avant 150 | 660 | 860 x 710 | 116 | 0.001-150 (normal range) 0.001-300 (column packing flow) |
5 |
ÄKTA pure | 630 | 535 x 470 | up to 53 kg | 0.001 to 25 (up to 50 during column packing) |
20 |
ÄKTA pure 150 | 630 | 535 x 470 | up to 53 kg | 0.01 to 150 (up to 300 during column packing) |
5 |
ÄKTAexplorer 10 | 620 | 500 x 460 | 75 | 0.001-10 | 25 |
ÄKTAexplorer 100 | 620 | 500 x 460 | 75 | 0.01-100 | 10 |
ÄKTAFPLC | 470 | 380 x 480 | 50 | 0.05-20 | 5 |
ÄKTAmicro | 610 | 480 x 450 | 55 | 0.001-2 | 35 |
ÄKTApilot | 900 | 750 x 540 | 114 | 4-400 (full gradients) 4-800 (limited gradients) |
2 |
ÄKTAprime plus | 530 | 400 x 450 | 13 | 0.1-50 | 1 |
ÄKTApurifier 10 | 620 | 500 x 460 | 75 | 0.001-10 | 25 |
ÄKTApurifier 100 | 620 | 500 x 460 | 75 | 0.01-100 | 10 |
ÄKTAxpress | 660 | 490 x 250 | 30 | 0.1-65 | 3 |
Ettan LC | 610 | 480 x 450 | 55 | 0.001-2 | 35 |
Ettan MDLC | 710 | 700 x 640 | 105 | 0.001-2 | 35 |
Ettan microLC | 1150 | 650 x 500 | 77 | 0.001-2 |
35
|
Ettan nanoLC | 1150 | 650 x 500 | 77 | 0.001-2 | 35 |
Flow path schemes
1. Flushed loop injection
With flushed loop injection means that the sample loop is completely filled with sample.
2. Partial loop injection
With partial loop injection means that the loop is partially filled with sample.
3. ul Pick-up injection
With ul Pick-up injection means that after aspiration from the vial, the sample volume
is transported into the loop with transport liquid (mobile phase) from another vial.
Spare parts
Figure. ÄKTAmicro, front view.
Please note that item numbers are not necessarily in sequential order.
# | Product Name | Product Code | Price | |
---|---|---|---|---|
1 | Tubing Kit for Rinsing System | 18111332 | 432.00 USD |
Add to cart
|
1 | O-Ring, 1 × 3 mm, PFR | 19003601 | 89.07 USD |
Add to cart
|
1 | Check valve kit | 28924679 | 907.00 USD |
Add to cart
|
1 | Piston Kit | 28924680 | 477.00 USD |
Add to cart
|
1 | Seal kit | 28924689 | 580.00 USD |
Add to cart
|
1 | Purge valve | 28924692 | 242.00 USD |
Add to cart
|
2 | Fiber Detachment Tool | 18111116 | 131.81 USD |
Add to cart
|
3 | Monitor pH/C-900 | 18110776 | 11,500.00 USD |
Add to cart
|
4 | Mounting Bracket for Valves | 18110911 | 35.21 USD |
Add to cart
|
4 | Injection Kit, INV-907 | 18111089 | 630.00 USD |
Add to cart
|
4 | Fill Port, INV-907 | 18112766 | 266.00 USD |
Add to cart
|
4 | Valve kit INV-917 | 18114756 | 790.00 USD |
Add to cart
|
5 | Stop Plug 1/16" Male | 18111252 | 59.00 USD |
Add to cart
|
5 | Mixer M-925 for classic ÄKTA | 18111889 | 2,698.00 USD |
Add to cart
|
6 | On-line filter | 18111801 | 723.00 USD |
Add to cart
|
6 | On-line filter kit | 18112094 | 137.66 USD |
Add to cart
|
7 | Valve SV-903 | 18111449 | 788.00 USD |
Add to cart
|
8 | Mixer M-925 Mixing Chamber, 0.6 ml | 18111890 | 1,016.00 USD |
Add to cart
|
8 | Mixer M-925 Mixing Chamber, 2 ml | 18111891 | 590.00 USD |
Add to cart
|
8 | Mixer M-925 Mixing Chamber, 90 µl | 18114724 | 968.00 USD |
Add to cart
|
10 | Conductivity Flow Cell, 0.2 µl | 18114720 | 2,108.00 USD |
Add to cart
|
Cable UniNet, L = 0.7 m | 18110974 | 298.00 USD |
Add to cart
|
|
Signal Cable, (6-pin miniDin-Open) | 18111064 | 395.00 USD |
Add to cart
|
# | Product Name | Product Code | Price | |
---|---|---|---|---|
1 | Connector 3/16" Male for Tubing o.d. 3/16" | 18111249 | 314.00 USD |
Add to cart
|
2 | Ferrules for 3/16" o.d. tubing connector | 18111248 | 130.53 USD |
Add to cart
|
3 | Stop Plug 5/16" Male | 18111250 | 36.22 USD |
Add to cart
|
4 | Connector 1/16" Male/Luer Female | 18111251 | 109.24 USD |
Add to cart
|
5 | Stop Plug 1/16" Male | 18111252 | 59.00 USD |
Add to cart
|
6 | Union 1/16" Female/M6 Male | 18111257 | 128.73 USD |
Add to cart
|
7 | Fingertight Union 1/16" Male/M6 Female | 18111258 | 135.19 USD |
Add to cart
|
8 | Connector for 1/8" tubing | 18112117 | 199.76 USD |
Add to cart
|
9 | Ferrule for o.d. 1/8" Tubing | 18112118 | 145.94 USD |
Add to cart
|
10 | Fingertight Connector 1/16" Male for Tubing o.d. 1/16" | 18111255 | 123.16 USD |
Add to cart
|
11 | Union 1/16" Male - 1/16" Male, i.d. 0.25 mm | 18112092 | 38.29 USD |
Add to cart
|
12 | Ferrules for o.d. 1/16" tubing connector | 18112706 | 159.39 USD |
Add to cart
|
13 | Connector 1/16" Male | 18112707 | 316.00 USD |
Add to cart
|
14 | Union 5/16" female - 1/16" male | 18114208 | 56.92 USD |
Add to cart
|
15 | Union 1/16" female-1/16" female | 18385501 | 105.84 USD |
Add to cart
|
16 | Tubing Connector Flangeless/M6 Male | 18101798 | 79.78 USD |
Add to cart
|
17 | Fingertight Narrow Head Connector, o.d. 1/16" | 18114710 | 538.00 USD |
Add to cart
|
21 | Union 1/16" Male - 1/16" Male, i.d. 0.13 mm | 18112090 | 134.55 USD |
Add to cart
|
# | Product Name | Product Code | Price | |
---|---|---|---|---|
1 | Inlet filter holder kit | 11000407 | 167.67 USD |
Add to cart
|
2 | Screw lid kit | 11000410 | 49.68 USD |
Add to cart
|
3 | Inlet filter set | 11000414 | 109.71 USD |
Add to cart
|
4 | Inlet Filter Assembly Kit | 18111315 | 190.44 USD |
Add to cart
|
5 | Inlet Filter Set | 18111442 | 207.00 USD |
Add to cart
|
6 | Inlet manifold | 18112080 | 156.73 USD |
Add to cart
|
7 | Flow Restrictor FR-904 | 18111963 | 428.00 USD |
Add to cart
|
8 | Cable UniNet, L = 0.18 m | 18110972 | 144.58 USD |
Add to cart
|
9 | Cable UniNet, L = 0.3 m | 18110973 | 158.94 USD |
Add to cart
|
10 | Cable UniNet, L = 0.7 m | 18110974 | 298.00 USD |
Add to cart
|
11 | Cable UniNet, L = 1.5 m | 18111775 | 262.00 USD |
Add to cart
|
12 | Cable UniNet, L = 3 m | 18110975 | 386.00 USD |
Add to cart
|
13 | Cable UniNet L=15 m | 18111774 | 524.00 USD |
Add to cart
|
14 | Mains cable, 120 V | 19244701 | 71.49 USD |
Add to cart
|
15 | Mains cable 220 V | 19244801 | 110.80 USD |
Add to cart
|
16 | Signal Cable, (6-pin miniDin-Open) | 18111064 | 395.00 USD |
Add to cart
|
17 | Short column holder, for columns 10 to 50 mm o.d. | 18111317 | 220.00 USD |
Add to cart
|
18 | Small Column Clamp | 18114998 | 50.53 USD |
Add to cart
|
19 | Lab Rod Holder | 18111319 | 490.00 USD |
Add to cart
|
20 | Tubing cutter, for PEEK, EFTE, and FEP tubing i.d. 0.25, 0.5, 0.75, 1 and 1.6 mm | 18111246 | 99.55 USD |
Add to cart
|
# | Product Name | Product Code | Price | |
---|---|---|---|---|
1 | Tubing cutter, for PEEK, EFTE, and FEP tubing i.d. 0.25, 0.5, 0.75, 1 and 1.6 mm | 18111246 | 99.55 USD |
Add to cart
|
2 | Connector 1/16" Male/Luer Female | 18111251 | 109.24 USD |
Add to cart
|
3 | Union 1/16" Male - 1/16" Male, i.d. 0.13 mm | 18112090 | 134.55 USD |
Add to cart
|
4 | Union 1/16" Male - 1/16" Male, i.d. 0.25 mm | 18112092 | 38.29 USD |
Add to cart
|
5 | Fingertight Narrow Head Connector, o.d. 1/16" | 18114710 | 538.00 USD |
Add to cart
|
8 | Stop Plug 1/16" Male | 18111252 | 59.00 USD |
Add to cart
|
9 | Tubing Kit, i.d. 0.10 mm | 18114713 | 954.00 USD |
Add to cart
|
11 | PEEK Tubing, 2 m, i.d. 0.15 mm, o.d.1/16" | 18115659 | 446.00 USD |
Add to cart
|
12 | Microfraction Collection Kit, ÄKTAmicro | 28948780 | 545.00 USD |
Add to cart
|
13 | Tubing i.d. 0.25 mm, o.d. 1/16" | 18112095 | 66.24 USD |
Add to cart
|
14 | PEEK Tubing, 2 m, i.d. 0.5 mm, o.d. 1/16" | 18111368 | 73.48 USD |
Add to cart
|
17 | New Purge Kit P-900 | 18112453 | 91.08 USD |
Add to cart
|
18 | Inlet filter holder kit | 11000407 | 167.67 USD |
Add to cart
|
19 | Inlet filter set | 11000414 | 109.71 USD |
Add to cart
|
20 | On-line filter kit | 18112094 | 137.66 USD |
Add to cart
|
21 | Flow Restrictor FR-904 | 18111963 | 428.00 USD |
Add to cart
|
22 | Mixer M-925 Mixing Chamber, 90 µl | 18114724 | 968.00 USD |
Add to cart
|
23 | Fiber Detachment Tool | 18111116 | 131.81 USD |
Add to cart
|
24 | Mains cable, 120 V | 19244701 | 71.49 USD |
Add to cart
|
25 | Mains cable 220 V | 19244801 | 110.80 USD |
Add to cart
|
26 | Cable UniNet, L = 3 m | 18110975 | 386.00 USD |
Add to cart
|
Seals are considered as consumables and should be replaced annually or when needed. Please keep a set of each in your stock to prevent long downtime. For code numbers refer to the table below.
# | Product Name | Product Code | Price | |
---|---|---|---|---|
1 | MDLC Pump head | 28923066 | 3,473.00 USD |
Add to cart
|
2 | Inlet filter set | 11000414 | 109.71 USD |
Add to cart
|
3 | On-line filter kit | 18112094 | 137.66 USD |
Add to cart
|
4 | Flow Restrictor FR-904 | 18111963 | 428.00 USD |
Add to cart
|
5 | Mixer M-925 Mixing Chamber, 0.6 ml | 18111890 | 1,016.00 USD |
Add to cart
|
6 | Mixer M-925 Mixing Chamber, 2 ml | 18111891 | 590.00 USD |
Add to cart
|
7 | Mixer M-925 Mixing Chamber, 90 µl | 18114724 | 968.00 USD |
Add to cart
|
9 | Valve SV-903 | 18111449 | 788.00 USD |
Add to cart
|
10 | Valve kit INV-917 | 18114756 | 790.00 USD |
Add to cart
|
11 | Check valve kit | 28924679 | 907.00 USD |
Add to cart
|
12 | Piston Kit | 28924680 | 477.00 USD |
Add to cart
|
13 | Seal kit | 28924689 | 580.00 USD |
Add to cart
|
14 | Purge valve | 28924692 | 242.00 USD |
Add to cart
|
# | Product Name | Product Code | Price | |
---|---|---|---|---|
1 | Inlet tubing kit, i.d. 1.6 mm, o.d. 3/16 in | 18111926 | 161.46 USD |
Add to cart
|
2 | FEP Tubing, i.d. 1/8 in (3/16 in) | 18111247 | 107.23 USD |
Add to cart
|
3 | Tubing i.d. 0.25 mm, o.d. 1/16" | 18112095 | 66.24 USD |
Add to cart
|
4 | ETFE Tubing, 2 m, i.d. 0.5 mm, o.d. 1/16 in | 18112096 | 61.00 USD |
Add to cart
|
5 | PEEK Tubing, 2 m, i.d. 0.75 mm, o.d. 1/16" | 18111253 | 96.26 USD |
Add to cart
|
6 | PEEK Tubing, 2 m, i.d. 0.5 mm, o.d. 1/16" | 18111368 | 73.48 USD |
Add to cart
|
7 | PEEK Tubing, 2 m, i.d. 1.0 mm, o.d. 1/16" | 18111583 | 85.34 USD |
Add to cart
|
8 | ETFE Tubing, 1.5 m, i.d. 0.25 mm, o.d. 1/16" | 18112136 | 254.00 USD |
Add to cart
|
9 | FEP Tubing, 3 m, i.d. 1/16", o.d. 1/8" | 18112116 | 73.48 USD |
Add to cart
|
10 | ETFE Tubing, 1.5 m, i.d. 0.75 mm, o.d. 1/16" | 18111974 | 80.73 USD |
Add to cart
|
11 | PEEK Tubing, 2 m, i.d. 0.15 mm, o.d.1/16" | 18115659 | 446.00 USD |
Add to cart
|
12 | Capillary tubing i.d. 0.35 mm | 18114735 | 1,234.00 USD |
Add to cart
|
14 | Tubing Kit, i.d. 0.10 mm | 18114713 | 954.00 USD |
Add to cart
|
15 | PEEK Tubing, 2 m, i.d. 0.15 mm, o.d.1/16" | 18115659 | 446.00 USD |
Add to cart
|
Accessories
# | Product Name | Product Code | Price | |
---|---|---|---|---|
3 | Short column holder, for columns 10 to 50 mm o.d. | 18111317 | 220.00 USD |
Add to cart
|
4 | Small Column Clamp | 18114998 | 50.53 USD |
Add to cart
|
5 | Lab Rod Holder | 18111319 | 490.00 USD |
Add to cart
|
13 | Microfraction Collection Kit, ÄKTAmicro | 28948780 | 545.00 USD |
Add to cart
|
Troubleshooting
Find solutions to product related issues. For unlisted issues please contact local Cytiva service representation.
UV curve issues
Issues related to UV curve
Possible cause | Suggested remedy |
---|---|
Air in the eluents |
Degas if necessary. We recommend helium sparging. |
Dirt and residues in the flow path from previous runs |
Clean the system according to instructions in the ÄKTAmicro System Manual. |
Poor mixing of the buffers |
Check the mixer function by placing a stirrer bar on top of the mixer housing. The stirrer bar should rotate when the system is activated. |
Residues in the column from previous run |
Clean the column in accordance to the column instruction. |
Possible cause | Suggested remedy |
---|---|
Air in the eluents |
Degas if necessary. We recommend helium sparging. |
Air in the flow cell |
|
Contaminated UV-cell |
Clean the UV-cell according to instructions in the ÄKTAmicro Monitor UV-900 User Manual. |
Impure buffer |
Check if the signal is still noisy in water. |
Leaking connections of the UV cell optical fibres |
Tighten the connectors. If necessary replace the connectors. Please refer to order information for connectors. |
Conductivity curve
Issues related to Conductivity curve
Possible cause | Suggested remedy |
---|---|
Calibration of the conductivity cell is incorrect |
Calibrate the conductivity cell according to instructions in the ÄKTAmicro Monitor pH/C-900 User Manual. |
Calibration solution 1.00 M NaCl not correct prepared |
Prepare a new calibration solution and recalibrate the conductivity cell according to instructions in the ÄKTAmicro Monitor pH/C-900 User Manual. |
Possible cause | Suggested remedy |
---|---|
Contaminated conductivity flow cell |
Clean the flow cell according to instructions in the ÄKTAmicro Monitor pH/C-900 User Manual. |
The ambient temperature may have decreased. |
The conductivity of the solution changes with temperature. Use a temperature compensation factor according to instructions in the ÄKTAmicro Monitor pH/C-900 User Manual. |
Possible cause | Suggested remedy |
---|---|
Air bubbles are passing through the flow cell |
|
Charged sample has been detected |
Prepare the sample so that charge particles are eliminated. |
Possible cause | Suggested remedy |
---|---|
Column is not equilibrated. |
Check that the column is equilibrated. If necessary clean the column. |
Conductivity flow cell cable is not correctly connected. |
Connect the conductivity cell cable to the rear panel of the instrument. |
If temperature compensation is being used, the temperature sensor might not be calibrated. |
Please calibrate the temperature sensor according to instructions in the ÄKTAmicro Monitor pH/C-900 User Manual. |
If temperature compensation is being used, incorrect temperature compensation factor might be in use. |
Please adjust the temperature compensation factor according to the instructions in the ÄKTAmicro Monitor pH/C-900 User Manual. |
Poor mixing of the buffers. |
Check the mixer function by placing a stirrer bar on top of the mixer housing. The stirrer bar should rotate when the system is in Run mode. The mixer function can also be checked by running the installation test. |
System pump doesn’t operate properly. |
Check the operation of the pump according to the ÄKTAmicro Pump P-905 User Manual. |
Valve doesn’t operate correctly. |
Check the valves according to Valve INV-917 Instructions or Valve SV-903 Instructions. |
Possible cause | Suggested remedy |
---|---|
Air in the flow cell |
Remove the air by flushing the flow cell. |
Column is not equilibrated |
Check that the column is equlibrated. If necessary clean the column. |
Contaminated conductivity flow cell |
Clean the flow cell according to instructions in the ÄKTAmicro Monitor pH/C-900 User Manual. |
Leaking tubing connections |
Tighten the connectors. If necessary replace the connectors. Please refer to the order information for connector. |
Poor mixing of the buffers |
Check the mixer function by placing a stirrer bar on top of the mixer housing. The stirrer bar should rotate when the system is in Run mode. The mixer function can also be checked by running the installation test. |
System pump doesn't operate properly |
Check the operation of the pump according to the ÄKTAmicro Pump P-905 User Manual. |
Possible cause | Suggested remedy |
---|---|
Dirt or residues in the flow path from previous run |
Clean the system according to instructions in the ÄKTAmicro System Manual. Make sure that the tubing has been washed properly. |
The system pump doesn’t operate properly |
Check the operation of the pump according to the ÄKTAmicro Pump P-905 User Manual. |
The volume of the mixing chamber is too large |
Replace the mixing chamber to a chamber with smaller volume according to the Mixer M-925 Instructions. |
Possible cause | Suggested remedy |
---|---|
The mixing chamber contains particles or other impurities |
Clean or replace the mixing chamber according to the Mixer M-925 Instructions. |
The mixer motor doesn’t operate resulting in poor mixing |
Check the motor operation. Place the hand/finger on the mixer and start it by starting the pump at low flow rate. You should both hear and feel the mixer motor and stirrer when they are spinning. |
The system pump doesn’t work properly |
Check that the pump is operating and is programmed correctly according to the ÄKTAmicro Pump P-905 User Manual. |
The volume of the mixing chamber is too small |
Replace the mixing chamber to a chamber with larger volume according to Mixer M-925 Instructions. |
Pressure curve issues
Issues related to Pressure curve
Possible cause | Suggested remedy |
---|---|
Air bubbles passing through or trapped in the pump |
|
Blockage or partial blockage of flow path |
|
Inlet and outlet checkvalves not functioning correctly |
There might be dirt in the checkvalves. |
Piston seal is leaking |
Replace the pumphead. |
Valve SV-903 unusual appearance
Issues related to Valve SV-903 unusual appearance
Possible cause | Suggested remedy |
---|---|
Connector incorrectly fitted or worn |
Tighten or replace the connector if necessary. |
Possible cause | Suggested remedy |
---|---|
Damage of the valve interior |
Replace the valve according to instructions in the Valve SV-903 Instructions. |
Possible cause | Suggested remedy |
---|---|
Connector incorrectly fitted or worn |
Tighten or replace the connector if necessary. |
Pump is not operating correctly |
Check the pump |
Pump is incorrectly programmed |
Check the programming of the pump |
Monitor pH/C-900 unusual appearance
Issues related to Monitor pH/C-900 unusual appearance
Possible cause | Suggested remedy |
---|---|
Bad scaling |
Adjust the conductivity and pH scaling according to instructions in the ÄKTAmicro Monitor pH/C-900 User Manual. Select conductivity scaling and Using an external chart recorder. |
Chart recorder not properly set up |
Check the chart recorder in accordance with its instructions. |
Input voltage is wrong |
Make sure that the input voltage is 1 V. |
Possible cause | Suggested remedy |
---|---|
No power to the monitor |
Check that the mains cable is connected and the power switch is in ON-position 1. |
General advice to achieve good performance
Before using the system make sure that:
- The correct system has been selected in UNICORN System Control
- The correct wavelength has been set for UV/UPC monitor
- All tubing has been properly connected
- All connectors are free from leakage
- No tubing is folded or twisted
- Online filter, if used, is changed on a regular basis
- Correct buffers are used for the chosen columns and proteins
- All inlet tubing has been immersed in correct buffer solutions
- Enough buffer has been prepared
- Buffers have been equilibrated to the environment temperature
- Buffers/eluents have been degassed if necessary (e.g., in RPC runs)
- Suitable columns have been selected for the target proteins
- Columns have been cleaned and prepared according to column instructions
- Samples have been clarified by centrifugation and/or filtration prior to sample loading
- Samples have been adjusted to binding buffer conditions
- Auto sampler (if used) has been prepared according to user manual
- The fraction collector has been filled with appropriate number of microtiter plates or tubes
- Appropriate arrangement for waste handling has been prepared
Valve INV-917 unusual appearance
Issues related to INV-917 unusual appearance
Possible cause | Suggested remedy |
---|---|
Connector incorrectly fitted or worn |
Tighten or replace the connector if necessary. Please refer to order information for connectors. |
Possible cause | Suggested remedy |
---|---|
Dirt in the flow path |
|
Possible cause | Suggested remedy |
---|---|
Internal parts are worn (The internal leakage can be detected at the small hole on the underside of the valve body) |
Change channel plate and distribution plate according to Valve INV-917 Instructions. |
Possible cause | Suggested remedy |
---|---|
Wrong electrical connection |
Check the connection to the pump. The valve should be connected to the UniNet 2 socket, not to the UniNet 1 socket. |
Wrong ID set on the switch |
Set the ID number 1 as indicated in the flow scheme in UNICORN. |
UniNet cable is worn |
Check the UniNet cable and replace if required. |
Possible cause | Suggested remedy |
---|---|
Valve parts incorrectly reassembled after replacement |
Check that the distribution plate markings 1-7 are horizontal. |
Pump P-905 unusual appearance
Issues related to Pump P-905 unusual appearance
Possible cause | Suggested remedy |
---|---|
There could be several causes of abnormal pressure recording, such as trapped air in pump heads, partially blocked solvent filters, leaking connectors, piston seal leakage, check valve malfunction or piston damage. |
To check the pump function, a recording of the pressure can be made, or by checking the pressure in UNICORN: By observing the piston stroke indicator in the Check menu together with the pressure trace, the pump head which is functioning abnormally can be identified. Some examples of normal and abnormal pressure traces together with comments are shown in the ÄKTAmicro Pump P-905 User Manual. |
Possible cause | Suggested remedy |
---|---|
Chart recorder not properly set up |
Check the chart recorder in accordance with its instructions. |
Possible cause | Suggested remedy |
---|---|
Connector incorrectly fitted or worn |
|
Possible cause | Suggested remedy |
---|---|
Piston seal or rinsing membrane incorrectly fitted or worn |
Replace the pumphead. |
Possible cause | Suggested remedy |
---|---|
Piston is damaged |
Replace the pumphead. |
Spring is corroded |
Replace the pumphead. |
Possible cause | Suggested remedy |
---|---|
No power to the pump |
Check that the mains cable is connected and the power switch is in ON-position 1. |
Mixer unusual appearance
Issues related to Mixer unusual appearance
Possible cause | Suggested remedy |
---|---|
Mixer not functioning |
Check the mixer function by placing a stirrer bar on top of the mixer housing. The stirrer bar should rotate when the system is in Run mode. The mixer function can also be checked by running the installation test. |
Monitor UV-900 unusual appearance
Issues related to Monitor UV-900 unusual appearance
Possible cause | Suggested remedy |
---|---|
Monitor UV-900 not properly set up |
|
Recorder not properly set up |
Check the chart recorder in accordance with its instructions. |
Possible cause | Suggested remedy |
---|---|
No power to the monitor |
Check that the mains cable is connected and the power switch is in ON-position 1. |
Possible cause | Suggested remedy |
---|---|
Mixer chamber volume too small |
Replace the mixing chamber to a chamber with larger volume according to the Mixer M-925 Instructions. |
Maintenance
Maintenance instructions and procedures.
ÄKTAmicro
Periodic maintenance
Regular maintenance is important for safe and trouble-free operation of your instrument. The user should perform daily and monthly maintenance. Preventive maintenance should be performed on a yearly basis by qualified service personnel.
For maintenance of a specific component, carefully read the component manual and follow the instructions. To avoid personal injury when performing maintenance on the ÄKTAmicro instrument, follow the instructions below.
Electrical shock hazard. All repairs should be done by service personnel authorized by Cytiva. Do not open any covers or replace parts unless specifically stated in the user documentation.
Disconnect power. Always disconnect power from the instrument before replacing any component on the instrument, unless stated otherwise in the user documentation.
Hazardous chemicals during maintenance. When using hazardous chemicals for system or column cleaning, wash the system or columns with a neutral solution in the last phase or step.
Do not perform any type of maintenance work while the system is powered electrically or when the piping system is pressurized. Note that the piping system can be pressurized even when the system is closed down.
When using hazardous chemicals, take all suitable protective measures, such as wearing protective glasses and gloves resistant to the chemicals used. Follow local regulations and instructions for safe operation and maintenance of the system.
Cleaning. Keep the instrument dry and clean. Wipe regularly with a soft damp tissue and, if necessary, a mild cleaning agent. Let the instrument dry completely before use.
Make sure that the piping system is completely leakage free before performing any CIP on the system.
NaOH is corrosive and therefore dangerous to health. When using hazardous chemicals, avoid spillage and wear protective glasses and other suitable personal protective equipment.
Before disassembly, check that there is no pressure in the piping system.
After assembly, the piping system must be tested for leakage at maximum pressure for continued protection against injury risks due to fluid jets, burst pipes or explosive atmosphere.
Disconnect power. Always disconnect power from the instrument before replacing fuses.
For continued protection from fire hazard, replace only with same type and rating fuse.
Incorrectly fitted tubing may loosen, causing a jet of liquid to spray out. This is especially dangerous if hazardous chemicals are in use. Connect the tubing by first inserting the tubing fully, then tightening the connector fingertight. PEEK nuts should be tightened a further 1/3-1/2 of a turn using the key supplied. Metal nuts should be tightened a further 1/4-1/3 a turn using the key supplied.
Handle the check valves with care when they have been removed from the pump heads to prevent loss of any internal components.
To protect the piston seals, the pump must never be run with air in the inlet tubing. Follow the procedure below to remove the air.
To protect the pump seals, always ensure that there is a constant supply of eluent. The pump should never be allowed to run dry.
To prevent precipitation of crystals when changing from a salt-containing buffer to organic solvent, always flush through the system with water as the intermediate liquid.
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of equipment.
Maintenance operations should be performed by the user at regular intervals
- Inspect the complete system for eluent leakage.
- The system can be left filled with buffer overnight. If you are not using the separation unit for a few days, wash the flow path with distilled water. Replace the column by a bypass capillary. Then wash the system with 20% ethanol and store it in 20% ethanol. Make sure that all tubing and all flow paths used are rinsed.
- Check for leakage. If there is a sign of liquid leaking between the pump head and the housing side panel or increased or decreased volume of rinsing solution, replace the pumphead.
- When changing eluent, it is important to remove trapped air. Run the Purge the pump procedure below. If there is still air in the inlet tubing, stop and run the Removing trapped air bubbles from the pump procedure below.
Check the inlet filters visually and replace them if necessary.
On-line filterReplace the on-line filter.
Pump rinsing solution- Change rinsing solution. Always use 20% ethanol as rinsing solution.
- If the volume of rinsing solution in the storage bottle has increased, it can be an indication of internal pump leakage. Replace the pumphead.
- If the volume of rinsing solution in the storage bottle has decreased significantly, check if the rinsing system connectors are mounted properly.
- If the rinsing system connectors are not leaking, the rinsing membranes or piston seals may be leaking. Replace the pumphead.
Run the Cleaning the system procedure.
Run the Checking the instrument procedure below.
Run the Cleaning the flow cell and optical fiber connectors procedure below.
Monitor pH/C-900Run the Cleaninging the conductivity flow cell procedure below. Cleaning the flow cell may be required more often if crude samples are regularly used.
Mixer M-925Check that the mixer chamber is clean and without damage. Check the tubing connectors. Replace if required.
Check for external and/or internal leakage. Run the Changing channel plate and distribution plate procedure below yearly or when required.
Replace the complete mixing chamber on a regular basis.
Replace the pumphead.
Monitor pH/C-900Run the Cleaninging the conductivity flow cell procedure below. Cleaning the flow cell may be required more often if crude samples are regularly used.
Purge the pump
1
Fill a reservoir flask with distilled water. Immerse the inlet tubings of both pump modules, with filters, in the water.
Note: Never place the reservoir flask below the level of the pump inlet.
2
Connect a male Luer syringe of about 30 ml to the open end of the purge tubing.
3
Connect the male Luer connector at the other end of the purge tubing to the left purge valve at pump module module A.
4
Turn the purge valve counterclockwise half a turn to open it and slowly draw eluent to the syringe.
5
When fluid starts to enter the syringe continue to draw a few milliliters before closing the purge valve. Check that there is no air left in the inlet tubing.
6
Repeat steps 3 to 5 for pump module B, if fitted.
Removing trapped air bubbles from the pump
During routine operation, the presence of air bubbles in the pump heads is seen as an erratic flow, a noisy detector signal or an irregular pressure recording. With air bubbles trapped in the pump, the pressure reading on the display will fluctuate considerably.
To remove the air from the pump:
1.
If the air has accumulated because of a leaking connector, correct the fault.
2.
Ensure that the inlet filter is fully immersed in eluent.
3.
Run the Purge the pump procedure.
If the problem remains, try to remove the air bubbles in accordance with the method described below. Use 100% methanol.
1.
Run at 2 ml/min.
2.
Continue to run for 10 minutes, or until the air bubble is removed.
3.
Run the Changing eluent procedure.
Changing eluent
When changing from one eluent to another, it is extremely important that the two eluents are totally miscible with one another. If the two eluents are immiscible, the pump should be flushed first with an intermediate liquid, which is miscible with both eluents. Failure to do this will cause a wrong flow of eluent from the pump.
When changing from a salt-containing buffer to an organic solvent, use water as the intermediate liquid to prevent precipitation.
1
Stop the pump by setting it in End mode.
2
Transfer the inlet tubing into the new eluent or into the intermediate liquid.
3
Run the pump at a flow rate and time as specified in the table.
Pump
|
Flow rate
|
Time
|
P-905
|
2 ml/min
|
20 minutes
|
4
Stop the pump. If an intermediate liquid is being used, transfer the inlet tubing into the final eluent and repeat step 3 with the new eluent.
In UNICORN, select instruction PumpWash in System Control:Manual:Pump.
Cleaning the system
Clean the flow path every month or when problems such as ghost peaks occur.
1
Disconnect the column and replace it with the G4 tubing.
2
Place all the inlet tubing in 100% acetonitrile.
3
Manually, perform PumpWash for all the four inlet tubings.
4
Flush the whole system with 100% acetonitrile for 5 minutes (1 ml/min).
5
Immediately, repeat step 3 and 4 with distilled water to rinse the flow path from acetonitrile.
Checking the instrument
1
Select menu Check, press OK.
2
Select menu Check Lamp Intensity.
Check lamp Intensity
210nm 85%
300nm 95%
If the lamp intensity is < 20%, contact your local GE Healthcare service representative for lamp replacement or change of internal optical fiber.
1
Select menu Check, press OK.
2
Select menu Check Lamp Run Time.
Check Lamp Run Time 200h
If the lamp on time is > 4000 hours, contact your local GE Healthcare service representative for lamp replacement.
Cleaning the flow cell and optical fiber connectors
Cleaning the flow cell
A clean flow cell and optical connectors are essential for ensuring the correct operation of the UV-monitor.
1
Connect a syringe to the inlet of the flow cell and squirt distilled water through the cell in small amounts. Then fill the syringe with a 10% surface active detergent solution like Decon 90, Deconex 11, RBS 25 or equivalent, and squirt five times.
2
After five squirts, leave the detergent solution in the flow cell for at least 20 minutes.
3
Pump the remaining detergent solution through the flow cell.
4
Rinse the syringe and flush the cell with distilled water (10 ml).
Cleaning the optical fiber connectors
Wipe with 30% isopropanol on lens paper.
Instrument housing
Wipe the instrument housing regularly with a damp cloth. Let the instrument dry completely before use.
Cleaning the conductivity flow cell
If the conductivity measurements are not comparable to previous results, the electrodes in the flow cell may be contaminated and require cleaning.
To clean the flow cell:
1
Pump 15 ml of 1 M NaOH at 1 ml/min through the flow cell either by using a pump or a syringe.
2
Leave it for 15 minutes.
3
Rinse thoroughly with 50 ml de-ionised water.
Note: If the flow cell is totally blocked, the blockage can be broken using a thin needle or a piece of string with a diameter less than 0.3 mm.
Changing channel plate and distribution plate
A replacement kit, Valve Kit INV-907, is available, see Spare parts recommended to keep on site in the Spare Parts section under the Related Products tab for code no.
1
Ensure that the valve is in position 1 and then disconnect it from the pump.
2
Remove the 4 screws on the front using the supplied 3 mm Allen key. Loosen each one equally in turn so the distribution plate comes off parallel to the valve body.
3
Slide the screws out.
4
Remove the distribution plate containing the ports.
5
Remove the old channel plate and insert a new one.
6
Remount a new distribution plate so that the text 3 is horizontal and to the right of the central tubing connection. Using the Allen key, tighten the 4 screws in turn, a little at a time, until the distribution plate is fixed to the valve body.
Replacing the piston seal, membranes and piston
If there are signs of liquid leaking between the pump head and the housing side panel or the volume of the rinsing solution has increased or decreased, replace the piston seal of the leaking pump head.
CAUTION! Do not disassemble the pump head unless there is good reason to believe that the seal is leaking. Always ensure that sufficient spare components are available before attempting to replace the piston seal. It is not possible to reinstall a used piston seal after removal.
Note: The power must be switched OFF when removing and refitting the pump heads.
Note: Always replace the piston seals on both pump heads at the same time. An even better practice is to replace all four piston seals.
Spare parts and tools required:
Seal kit containing:
2 or 4 piston seals
2 or 4 rinse membranes
1/4 inch wrench (supplied with the pump)
3 mm allen key (supplied with the pump)
Screwdriver (supplied with the pump)
Note: After a new seal has been installed, the pump should be run in. Run the Running-in a new piston seal procedure.
Note: Before disassembling the pump heads move all input buffers bottles below the level of the pump heads to prevent siphoning.
CAUTION! Read the following instructions carefully. The individual parts of the pump head can be assembled incorrectly. Take care to ensure that the orientation of each part is correct before continuing with the next instruction.
1
Switch off the pump at the mains power switch on the back panel.
2
Remove the piston seal rinsing system. The connectors are simple plugin fittings.
3
Completely loosen the tubing connector on the outlet valve.
4
Remove the Switch valves and disconnect the tubing. Unscrew the two white knurled screws under the pump to release the support bracket locking the inlet manifold into the inlet valve. Gently lower and remove the complete manifold.
5
Using the Allen key, unscrew and completely remove one of the two Allen screws locking the pump head in position.
6
When unscrewing the second locking screw, push firmly on the front face of the pump head to compensate for the pressure of the piston return spring. Hold the pump head firmly to prevent it from twisting. Remove the second screw and, without allowing the pump head to twist sideways, carefully pull it out.
7
Place the pump head face down on the bench. Pull out the piston together with the return spring.
8
Inspect the piston and return spring for sign of damage. If damaged, they should be replaced.
9
Wipe the piston with a clean cloth. If salt solutions have been used the piston may be slightly corroded. This corrosion can be removed with a rubber eraser. If it cannot be wiped or rubbed clean, scrape off any deposits with a scalpel or razor blade. Inspect the piston with a magnifying glass for scratches. Replace with a new piston if any scratches or cracks are found.
10
Remove the two screws securing the drain plate and the rinse chamber. Remove and discard the rinsing membrane. Remove the rinse chamber. Remove also the support washer.
11
Gently withdraw the piston seal. Discard the used seal.
12
The pump head, rinse chamber and drain plate should be carefully rinsed or cleaned in an ultrasonic bath, if available. If dirt can be seen on any surfaces, the inlet and outlet check valves should be removed and cleaned separately. Run the Cleaning the inlet and outlet check valves procedure.
13
Wet the new seal slightly and place it in the hole on the pump head and press it down into position with a hard flat object. Refit the support washer on top of the new seal.
14
With the pump head still facing downwards on the bench, place the rinse chamber onto the head with the rinse ports in line with the inlet and outlet check-valves. The conical depression in the rinse chamber should be facing upwards, ready to accept the new rinsing membrane. Fit the rinsing membrane with the conical face downwards.
15
Place the drain plate on top of the assembly. Use the two screws to lock the complete assembly together.
Note: Align the drainage hole in the drainage plate with the inlet check valve (the opposite side of the pump head marked OUT/UP).
16
Wipe clean the piston and remove all finger prints. Wet the piston and then insert it into the return spring. With the pump head facing downwards on the bench, insert the piston into the pump head by pushing it gently but firmly vertically downwards into the seal.
CAUTION! Do not push the piston at an angle to the head and DO NOT twist the piston.
17
Turn the head so that the inlet valve and drainage hole are facing downwards and the text UP/OUT on the pump head is facing upwards. Mount the complete pump head over the locating pins on the front panel. Locate the metal end of the piston and the spring towards the drive cam.
Hold the pump head firmly against the side panel of the housing with one hand. Do not allow the assembly to twist under pressure from the return spring. Using the Allen key, fit and tighten one of the Allen screws.
Fit and tighten the remaining Allen screw.
18
Reconnect the outlet tubing to the outlet check valve and the manifold block as described above.
19
Reconnect the inlet manifold and, if used, the switch valves.
20
Refit the tubing of the piston seal rinse system.
21
The pump should now be purged and the new piston seal carefully run the Running-in a new piston seal procedure.
Running-in a new piston seal
The piston seal should be run-in using 100% methanol.
1
Ensure that the reservoir is filled with sufficient eluent. Immerse the inlet tubing in the eluent. The reservoir should be placed at least 30 cm above the pump inlet.
2
Connect a male Luer syringe of about 30 ml to the open end of the purge tubing.
3
Connect the male Luer connector at the other end of the purge tubing to the left purge valve at pump module module A.
4
Turn the purge valve counter-clockwise half a turn to open it and slowly draw eluent to the syringe.
5
When fluid starts to enter the syringe continue to draw a few milliliters before closing the purge valve. Check that there is no air left in the inlet tubing.
6
Repeat steps 3 to 5 for pump module B, if fitted.
7
Check that the outlet tubings are not blocked.
8
Connect a thin capillary or a column that will give sufficient back pressure.
9
Run the pump at 1 ml/min (or 2 ml/min at 50%B) at a back-pressure of 5-10 MPa for 2 hours, or longer if possible (e.g. overnight).
10
Finally, Change the eluent according to the procedure given below.
Replacing a damaged piston
Typical symptoms of a damaged piston are observed as excessive piston seal wear, unstable pressure, a reduction in the flow or, in som cases, noise as the piston moves. The piston should be removed, examined for damage or salt precipitation and the replaced with a new piston if necessary.
If a damaged piston has been in operation, the piston seal will be destroyed and should be replaced. Run the Replacing the piston seal, membranes and piston procedure.
Spare parts and tools required:
- Seal kit
- 1/4 inch wrench
- 3 mm Allen key
- Screwdriver
- Piston kit
Cleaning the inlet and outlet check valves
Faulty operation of the check valves is usually indicated by irregular flow, very low flow or unstable pressure traces. Probable causes of this are air or dirt in a check valve preventing it from closing to seal and hold the pressure.
Record the pressure according to instructions, checking the pump pressure at the end of this document and identify the faulty check valve by observing which pump head is delivering the flow according to check menu at the end of this document.
Try to clean the check valves in–place on the pump head by pumping distilled water at 2 ml/min for 2 minutes. This also prevents precipitation of crystals. The pump 100% methanol for approximately 10 minutes. If this does not correct the problem, follow the instructions for removing and then cleaning the valves.
Note: Change solvent to distilled water and flush out all salt before removing the check valves.
Tools required: A 13 mm, a 18 mm and a 1/4 inch wrench.
Before removing the check valves, move all input buffers bottles below the level of the pump heads, to prevent siphoning.
1
If the condition of the check valve is not improved by in–place cleaning, disconnect and remove the inlet manifold and outlet tubing.
2
Use the 13 mm wrench (inlet check valve) or the 18 mm wrench (outlet check valves) to remove the valve from the pump head.
3
Immerse the complete valve in methanol and place in an ultrasonic bath for some minutes.
Then repeat the ultrasonic bath with distilled water.
4
Refit the check valves. The inlet check valve (with a lip for the manifold and a larger diameter opening) is fitted to the side marked IN of the pump head. Tighten the valves until fully finger-tight and then use the wrench to tighten a further 1/3rd (110o) of a turn. Do not overtighten the valves since damage to the internal components can occur.
5
Refit the outlet tubing and the inlet manifold.
6
Purge the pump carefully and check that the pumping action has been corrected.
Note: Check valves have precision matched components and should only be disassembled further by a trained service engineer. If the problem cannot be corrected, the valve should be replaced completely.
To check the pump function, a recording of the pressure can be made or by checking the pressure in UNICORN. This pressure recording is more sensitive than the reading on the display. By observing the piston stroke indicator in the Check menu together with the pressure trace, the pump head which is functioning abnormally can be identified.
There can be several causes of an abnormal pressure recording, for example:
- air trapped in the pump heads
- partially blocked solvent filters
- leaking connections
- piston seal leakage
- check valve malfunction
- piston damaged
Check menu
To enable trouble shooting it is possible to check which pump module head that delivers flow.
1
Select main menu Check, press OK.
2
The display shows the status of the pistons for both pump modules.
A: Left means that the left pump head is delivering flow in the A pump and
B: Right means that the right pump head in delivering flow in the B pump. At the changing point both are displayed. The A pump is closest to the front panel.
Note:
Check Heads 2.30MPa
A: Left
B: Right
1
Select main menu Check, press OK.
2
Select sub menu Check Total Run Time. The display shows the accumulated operation time (Run and Hold mode) for the A and B pumps.
Note:
Check Total Run Time
A: 50 h
B: 20 h
1
Select main menu Check, press OK.
2
Select sub menu Check Piston Strokes. The display shows the accumulated number of piston strokes for the A and B pumps.
Note:
Check Piston Strokes
11374, 2398
1
Ensure that the pump is at zero pressure.
2
Select main menu Check, press OK.
3
Select sub menu Calib Press Offset, press OK. “Working, please wait ...” is displayed until the calibration is completed.
Service information relevant to the instrument can be checked. Information may not be available in all menus.
1
Select main menu Check, press OK.
2
Select sub menu Check Service Mode, press OK.
3
The service telephone number is displayed1. Turn the selection dial clockwise to select next sub menu.
4
The service contract number is displayed1. Turn the selection dial clockwise to select next sub menu.
5
The instrument serial number is displayed. Turn the selection dial clockwise to select next sub menu.
6
Instrument name and software version are displayed. Turn the selection dial clockwise to select next sub menu.
7
The date of the last service is displayed. Turn the selection dial clockwise to select next sub menu.
8
A test of the instrument buzzer is performed, press OK.
1Not always pre-programmed. Depends on customer-specific contract.