FAQ

Accessories

Running HiPrep, HisPrep and GSTPrep columns with ÄKTAexplorer, ÄKTAFPLC and ÄKTAprime plus, ÄKTApurifier and ÄKTAxpress.

 

Make sure you have all connectors and tubing needed for running the column.

Depending on your system configuration and column length you may need extra long tubing to connect your column.

Please do not extend the tubing unnecessarily which results in a lower resolution.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 PEEK Tubing, 2 m, i.d. 0.75 mm, o.d. 1/16" 18111253 88.79 USD Add to cart
Get Quote
2 Fingertight connector 1/16" male, narrow 28401081 101.51 USD Add to cart
Get Quote
3 Fingertight Connector 1/16" Male for Tubing o.d. 1/16" 18111255 113.65 USD Add to cart
Get Quote

Running HiPrep, HisPrep and GSTPrep columns with stand alone pump, for instance using HiLoad pump P-50 for step elution.

 

The figure below shows a complete set up which facilitates bypassing the column, running cleaning-in-place etc.

Make sure you have all connectors and tubing needed for running the column.

Depending on your system configuration and column length, you may need extra long tubing to connect your column.

Please do not extend the tubing unnecessarily which results in a lower resolution.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 Tubing cutter, for PEEK, EFTE, and FEP tubing i.d. 0.25, 0.5, 0.75, 1 and 1.6 mm 18111246 88.81 USD Add to cart
Get Quote
1 Unflanged tubing i.d. 1.9 mm, o.d. 2.7 mm 18820702 302.00 USD Add to cart
Get Quote
2 Tubing connectors o.d. 2.7 mm, M6 male 18465201 116.13 USD Add to cart
Get Quote
3 Tubing, 2 m i.d. 0.8 mm, o.d. 1.8 mm, ETFE 19743501 68.66 USD Add to cart
Get Quote
5 Pre-filter 6000 18458201 1,222.00 USD Add to cart
Get Quote
5 Filter kit 18458401 142.55 USD Add to cart
Get Quote
5 Holder Valve PSV-50 19199255 137.32 USD Add to cart
Get Quote
5 Solenoid Valve PSV-50 19199401 744.00 USD Add to cart
Get Quote
8 Fingertight Union 1/16" Male/M6 Female 18111258 122.88 USD Add to cart
Get Quote
9 UV Flow Cell, 2 mm 18111110 2,597.00 USD Add to cart
Get Quote
10 Filter 215 nm 11000733 744.00 USD Add to cart
Get Quote
10 Filter 260 nm 11000734 710.00 USD Add to cart
Get Quote
10 Filter 280 nm 11000735 571.00 USD Add to cart
Get Quote
10 Filter 405 nm 11000736 679.00 USD Add to cart
Get Quote
10 Empty Filter Holder 11000738 222.00 USD Add to cart
Get Quote
10 Monitor UVis-920 Excluding Filter and Flow Cell 11000754 9,825.00 USD Add to cart
Get Quote
10 Short optical fibre kit, 200 mm 18113485 867.00 USD Add to cart
Get Quote
10 Long optical fibre kit, 500 mm 18113486 640.00 USD Add to cart
Get Quote
10 UV flow cell, i.d. 8 mm 28966687 3,328.00 USD Add to cart
Get Quote
10 O-ring kit 28969705 147.21 USD Add to cart
Get Quote
12 Union M6 female - 1/16" male 18385801 275.00 USD Add to cart
Get Quote

Make sure you have all connectors and tubing needed for running the column.

Depending on your system configuration and column length, you may need extra long tubing to connect your column.

Please do not extend the tubing unnecessarily which results in a lower resolution.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 PEEK Tubing, 2 m, i.d. 0.75 mm, o.d. 1/16" 18111253 88.79 USD Add to cart
Get Quote
2 Fingertight connector 1/16" male, narrow 28401081 101.51 USD Add to cart
Get Quote
3 Fingertight Connector 1/16" Male for Tubing o.d. 1/16" 18111255 113.65 USD Add to cart
Get Quote
4 Union 1/16" Female/M6 Male 18111257 117.01 USD Add to cart
Get Quote
5 Union, M6 Female - 5/16" Male 18116916 116.02 USD Add to cart
Get Quote

Troubleshooting

Find solutions to product related issues. For unlisted issues please contact local Cytiva service representation.

Poor product recovery

Columns Column leakage

Back pressure increases during operation

General advice to achieve good performance

Before using the column make sure that:

  • Correct system has been selected in UNICORN System Control
  • Correct wavelength has been set for UV/UPC monitor
  • All tubing has been properly connected and tubing is not longer than needed
  • All connectors are free from leakage, verified by passing a leakage test
  • No tubing is folded or twisted
  • Online filter, if used, is changed on a regular basis
  • Correct buffers are used for the chosen columns and proteins
  • All inlet tubing has been immersed in correct buffer solutions
  • Enough buffer has been prepared
  • Buffers have been equilibrated to the environment temperature
  • Buffers/eluents have been degassed if necessary (e.g., in RPC runs)
  • Suitable columns have been selected for the target proteins
  • Column meets pressure requirements for selected medium
  • Columns have been cleaned and prepared according to column instructions
  • Samples have been clarified by centrifugation and/or filtration prior to sample loading
  • Samples have been adjusted to binding buffer conditions
  • Auto sampler (if used) has been prepared according to user manual
  • The fraction collector has been filled with appropriate number of microtiter plates or tubes
  • Appropriate arrangement for waste handling has been prepared

 

 

Unsatisfactory elution

Poor reproducibility

Unusual column appearance

Contact