Accessories

 

Housing for ProCell cartridges:

Housing assembly for 152 M cartridges (56-4106-35)

Housing assembly for 154 M cartridges (56-4106-36)

 

TC clamps and gaskets for the feed, retentate and permeate lines:

Clamp 50 mm TC (44-7134-01)

Gasket 50 mm i.d. 38 mm (1.5 inch) in silicone (56-4109-96) or nitrile (56-4109-97)

Clamp 64 mm TC (56-4113-12)

Gasket 64 mm i.d. 50 mm (2 inch) in silicone (56-4109-98), nitrile (11-0005-24) or EPDM (56-4113-18)

 

To blank off the feed, retentate and permeate ports:

Blank off cap 50 mm TC incl gasket (44-7135-01)

Clamp 50 mm TC (44-7134-01)

Blank off cap 64 mm TC (56-4109-92)

Clamp 64 mm TC (56-4113-12)

Gasket 64 mm i.d. 50 mm (2 inch) in silicone (56-4109-98), nitrile (11-0005-24) or EPDM (56-4113-18)

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
2 Gasket 50 mm (1.5") TC i.d. 38 mm, silicone 56410996 88.79 USD Add to cart
Get Quote
2 50 mm (21.5") TC-Gasket, Buna-N 56410997 75.77 USD Add to cart
Get Quote
4 Gasket 64 mm (2") TC i.d. 50 mm, silicone 56410998 105.37 USD Add to cart
Get Quote
4 Gasket 64 mm (2") TC, i.d. 51 mm, EPDM 56411318 55.64 USD Add to cart
Get Quote
5 Blind Flange, 51 mm TC 44713501 101.82 USD Add to cart
Get Quote
6 Clamp, 50 mm TC, SS 44713401 10.68 USD Add to cart
Get Quote
8 Clamp 6 inch 56410674 594.00 USD Add to cart
Get Quote
10 Housing Assembly for ProCell 152M Cartridges 56410635 10 400.00 USD Add to cart
Get Quote
11 Housing Assembly for ProCell 154M Cartridges 56410636 10 700.00 USD Add to cart
Get Quote

The use of a permeate flow control is strongly recommended during upstream clarification of target material. Permeate flow control rate is dependent on the process flux (LMH) and surface area.

For permeate flow control you need a permeate pump kit. Detailed information regarding permeate pump kit for GrandStand 450 (GSM450PERPRP) and for GrandStand 550 (GSM550PERPRP) available below.

 

Accessories for ProCell cartridges that have to be ordered separately.

Housing for ProCell cartridges:

Housing assembly for 152 M cartridges (56-4106-35)

Housing assembly for 154 M cartridges (56-4106-36)

To blank off the permeate port of the cartridge:

Blank off cap 50 mm TC (stainless steel) incl. gasket (44-7135-01)

Clamp 50 mm TC, stainless steel (12-4100-30)

To blank off retentate and feed ports of cartridge, when stored off the system:

Blank off cap 64 mm TC, plastic (56-4109-92)

Clamp 64 mm TC, stainless steel (12-4100-28)

Gasket o.d. 64 mm i.d. 51 mm, 2 inch (12-4100-33)

 

Please note that the item numbers are not necessarily in sequential order.


When stored off the system, store the cartridge with storage solution using blank off cap, gasket and clamp for each port.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.
# Product Name Product Code Price
1 Housing Assembly for ProCell 152M Cartridges 56410635 10 400.00 USD Add to cart
Get Quote
1 Housing Assembly for ProCell 154M Cartridges 56410636 10 700.00 USD Add to cart
Get Quote
3 Clamp 50 mm (1") TC 12410030 15.39 USD Add to cart
Get Quote
3 Clamp 50 mm TC 56410668 75.77 USD Add to cart
Get Quote
3 Clamp 50 mm (1.5") TC (Toggle Clamp) 56410669 75.77 USD Add to cart
Get Quote
6 TC-Gasket 51 mm i.d. 10 mm, EPDM 18101288 189.42 USD Add to cart
Get Quote
6 Blind Flange, 51 mm TC 44713501 101.82 USD Add to cart
Get Quote
7 50 mm (1.5") TC 13 mm (0.5") Barb 56410986 131.41 USD Add to cart
Get Quote
7 50 mm TC i.d. 38 mm (1.5") 19 mm (0.75"ch) Barb 56410987 131.41 USD Add to cart
Get Quote
Clamp 6 inch 56410674 594.00 USD Add to cart
Get Quote

Troubleshooting

Find solutions to product related issues. For unlisted issues please contact local Cytiva service representation.

General Advice for Troubleshooting when Using Cross Flow Filters

Please check the following items during troubleshooting:

Samples volume and system working volume have been used in calculations of concentration factor and diafiltration beforehand.

Process settings of Membrane Capacity, Flux, Shear (or volumetric cross flow rate) have been properly applied for both sample characteristics and membrane pore size or NMWC designation.

Filter is properly wetted before use (also make sure that there is no air lock in the filters created by condensation on microfiltration membranes as a result of incorrect autoclave settings).

Ensure CFF filters are properly wetted before starting processing. This may be tested against new cartridge “Clean Water Flux Test or against the water flux specifications in the Certificate of Quality. Microfitlration membranes are shipped dry and are rewetted with water or buffer. Hollow fiber ultrafiltration membranes are shipped in glycerol wetting preservative. These filters may require 15-20% alcohol/water solution to restore wetting or remove glycerol. Flat sheet cassette ultrafiltration devices (for example, KvickLab and KvickFlow Cassettes) are shipped in a solution of NaOH/Glycerol which must be flushed out before use.

Prior to using cassettes, make sure the single gasket is in the correct position in the holding device and that the cassette(s) within are oriented properly in the holder.

Check integrity of cassette system using water under pressure on retentate side to 1 barg to check for leaks in the seal of the cassette with the stainless steel holding device.

Check to see that cartridges or cassettes that have been previously used are checked for integrity (air diffusion or bubble point tests) and for cleanliness (normalized Clean Water Permeability before processing).