FAQ

Packing procedures

Recommended flow velocities in the table below are for packing the column. 

Column Compression factor Slurry concentration (%) Bed height Flow velocity (cm/h) Step 1 Flow velocity (cm/h) Step 2
XK 16/40 1.15 50 - 70 10 - 301 100 600
XK 26/40 1.15 50 - 70 10 - 30 95 570
XK 50/30 1.15 50 - 70 10 - 30 90 540

1 For bed heights of 10cm and flow rates up to 300 cm/h we recommend prepacked HiPREP 16/10 columns

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Column Maximum operation pressure (bar)
HiScale 20
XK 16 5
XK 26 5
XK 50 3
BPG 100 8
BPG 140 6
BPG 200 6
BPG 300 4
BPG 450 2.5
BioProcess LPLC 6
BioProcess MPLC 20
Chromaflow 400 3
Chromaflow 600 3
Chromaflow 800 3
Chromeflow 1000 3
FineLINE Pilot 35 20
FineLINE 70 20
FineLINE 100 20
FineLINE 200 20
FineLINE 350 20

 

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Constant flow packing.

Refer to the document Constant flow packing in BPG columns, method description and practical example, which you will find under the Related Documents tab. The same method is valid for Blue Sepharose 6 Fast Flow, Butyl Sepharose 4 Fast Flow, Butyl-S Sepharose 6 Fast Flow, Chelating Sepharose Fast Flow, CM Sepharose Fast Flow, DEAE Sepharose Fast Flow, Heparin Sepharose 6 Fast Flow, IMAC Sepharose 6 Fast Flow, Ni Sepharose 6 Fast Flow, Octyl Sepharose 4 Fast Flow, Phenyl Sepharose 6 Fast Flow (low sub), Phenyl Sepharose 6 Fast Flow (high sub), PlasmidSelect, Q Sepharose Big Beads, Q Sepharose Fast Flow, Q Sepharose XL, Sephadex G-25, Sepharose 4 Fast Flow, Sepharose 6 Fast Flow, SP Sepharose Big Beads, SP Sepharose Fast Flow and SP Sepharose XL
in BPG columns.

Packing buffer: Water for Blue Sepharose 6 Fast Flow, Chelating Sepharose Fast Flow, CM Sepharose Fast Flow, DEAE Sepharose Fast Flow, Heparin Sepharose 6 Fast Flow, IMAC Sepharose 6 Fast Flow, Ni Sepharose 6 Fast Flow, PlasmidSelect, Q Sepharose Big Beads, Q Sepharose Fast Flow, Q Sepharose XL, Sepharose 4 Fast Flow, Sepharose 6 Fast Flow, SP Sepharose Big Beads, SP Sepharose Fast Flow and SP Sepharose XL.

Packing buffer: Water or 0.15 M NaCl for Sephadex G-25

Packing buffer: 20 % Ethanol for Butyl Sepharose 4 Fast Flow, Butyl-S Sepharose 6 Fast Flow, Octyl Sepharose 4 Fast Flow, Phenyl Sepharose 6 Fast Flow (low sub) and Phenyl Sepharose 6 Fast Flow (high sub).

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Recommended net porosity for Blue Sepharose 6 Fast Flow, Chelating Sepharose Fast Flow, CM Sepharose Fast Flow, DEAE Sepharose Fast Flow, Heparin Sepharose 6 Fast Flow, IMAC Sepharose 6 Fast Flow, Ni Sepharose 6 Fast Flow, Phenyl Sepharose 6 Fast Flow (low sub), Phenyl Sepharose 6 Fast Flow (high sub), Q Sepharose Fast Flow, Q Sepharose XL, SP Sepharose Fast Flow and SP Sepharose XL: 20 µm

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Packing in place technology 

Refer to the document Pack-in-place using Chromaflow columns, method description and practical example, which you will find under the Related Documents tab. The same method is valid for Blue Sepharose 6 Fast Flow, Butyl Sepharose 4 Fast Flow, Butyl-S Sepharose 6 Fast Flow, Chelating Sepharose Fast Flow, CM Sepharose Fast Flow, DEAE Sepharose Fast Flow, Heparin Sepharose 6 Fast Flow, IMAC Sepharose 6 Fast Flow, Ni Sepharose 6 Fast Flow, Octyl Sepharose 4 Fast Flow, Phenyl Sepharose 6 Fast Flow (low sub), Phenyl Sepharose 6 Fast Flow (high sub), Phenyl Sepharose High Performance, PlasmidSelect, PlasmidSelect Xtra, Q Sepharose Big Beads, Q Sepharose Fast Flow, Q Sepharose High Performance, Q Sepharose XL, Sepharose 4 Fast Flow, Sepharose 6 Fast Flow, SP Sepharose Big Beads, SP Sepharose Fast Flow, SP Sepharose High Performance and SP Sepharose XL in Chromaflow columns.

Packing buffer: Water for Blue Sepharose 6 Fast Flow, Chelating Sepharose Fast Flow, CM Sepharose Fast Flow, DEAE Sepharose Fast Flow, Heparin Sepharose 6 Fast Flow, IMAC Sepharose 6 Fast Flow, Ni Sepharose 6 Fast Flow, PlasmidSelect, PlasmidSelect Xtra, Q Sepharose Big Beads, Q Sepharose Fast Flow, Q Sepharose High Performance, Q Sepharose XL, Sepharose 4 Fast Flow, Sepharose 6 Fast Flow, SP Sepharose Big Beads, SP Sepharose Fast Flow, SP Sepharose High Performance and SP Sepharose XL.

Packing buffer: 20 % Ethanol for Butyl Sepharose 4 Fast Flow, Butyl-S Sepharose 6 Fast Flow, Octyl Sepharose 4 Fast Flow, Phenyl Sepharose 6 Fast Flow (low sub), Phenyl Sepharose 6 Fast Flow (high sub) and Phenyl Sepharose High Performance.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

XK 16/40
10 µm nets (5 pieces/pack) Code No: 18-8761-01
XK 26/40
10 µm nets (5 pieces/pack) Code No: 18-8760-01
XK 50/30
10 µm nets (5 pieces/pack)Code No: 18-8759-01
HiScale 16
20 µm nets (5 pieces/pack) Code No: 28-9663-79
HiScale 26
20 µm nets (5 pieces/pack) Code No: 28-9663-80
HiScale 50
20 µm nets (5 pieces/pack) Code No: 28-9663-81

 

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

BPG 100
23 µm nets
Adaptor net Code no: 18-1103-08
End piece net Code no: 18-9252-01
BPG 140
23 µm nets
Adaptor net Code no: 18-1113-01
End piece net Code no: 18-1113-00
BPG 200
23 µm nets
Adaptor net Code no: 18-9253-01
End piece net Code no: 18-9254-01
BPG 300
23 µm nets
Adaptor net Code no: 18-1012-54
End piece net Code no: 18-1012-34

 

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Column Compression factor Slurry concentration (%) Bed height (cm) Pressure (bar)
Chromaflow 400 1.15 50 - 70 10 - 30 3
Chromaflow 600 1.15 50 - 70 10 - 30 3
Chromaflow 800 1.15 50 - 70 10 - 30 3
Chromaflow 1000 1.15 50 - 70 15 - 30 3

 

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Recommended flow velocity intervals in the table below are for packing the column. The intervals are given in reverse order to indicate that the high flow velocity can be used for shorter beds whereas higher beds only allow lower flow velocity, due to the higher pressure drop across the bed.

Column Compression factor Slurry concentration (%) Bed height (cm) Flow velocity (cm/h)
BPG 100 1.15 50 - 70 10 - 20 500 - 375
BPG 100 1.15 50 - 70 20 - 30 375 - 250
BPG 140 1.15 50 - 70 10 - 20 500 - 375
BPG 140 1.15 50 - 70 20 - 30 375 - 250
BPG 200 1.15 50 - 70 10 - 20 500 - 375
BPG 200 1.15 50 - 70 20 - 30 375 - 250
BPG 300 1.15 50 - 70 10 - 20 500 - 375
BPG 300 1.15 50 - 70 20 - 30 375 - 250

 

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Column evaluation
The efficiency of a column depends on how well it is packed. A poorly packed column gives rise to uneven flow, resulting in zone broadening and reduced resolution. It is thus important to have a method by which the column can be tested before it is put into operation. Such a method should be simple, quantitative and should not introduce contaminating materials. It is also an advantage if the same method can be used to monitor column performance over its working life, so that it is easy to determine when the medium should be re-packed or replaced.

Avoid methods that use colored compounds such as Blue Dextran. They do not meet the above criteria and cannot be used with ion exchange and affinity chromatography media.

Experience has shown that the best method of expressing the efficiency of a column is in terms of the height equivalent to a theoretical plate, HETP, reduced plate number, h, and the peak asymmetry factor, As. These values can be determined easily by applying a NaCl or acetone solution, to the column (see below).

It is important that the column is properly equilibrated ( >2 column volumes) before evaluating the packing. Ideally, run three test runs to see whether the values are stable. If an initially poor result improves during a later test, the reason can be that the column was not properly equilibrated. To check that the bed is stable, run the column at 70% packing pressure for 20 hours and test it again.

Note that pressure spikes may cause poor packing (cracking). If this happens, fit an air trap and a pressure relief valve between the pump and column. Locate the pressure relief valve between the air trap and the column.

Choice of test sample for columns
The most appropriate material for column testing is, of course, the sample that is to be run in the application, but this is not always practical or economical. As an alternative, a solution of either NaCl or acetone will give a good indication of the column packing quality. The eluate is monitored by measuring conductivity or UV absorption, and the resulting elution profile is used to calculate the HETP value.

The advantages of using NaCl are that it is readily available and can be used safely to test all columns. One disadvantage is that NaCl may interact with the medium matrix, especially ion exchanger matrices, and thus give erroneous results.

Acetone, in contrast, does not interact with the matrix and is detected by UV absorption at 280 nm. Alternatively, you can increase the running buffer concentration 10-fold and use it as test solution.

The figure below shows a UV trace for acetone in a typical BPG column application and gives calculated HETP and As values.

HETP calculation
The sample volume should be approximately 1% of the total bed volume and the concentration 1.0% v/v NaCl, or equivalent when using stronger buffer. Alternatively, use 1.0% v/v acetone. The flow velocity should be between 10 and 30 cm/h depending on the bead size of the chromatography medium. The high flow velocity could be used for small beads whereas large beads only allow low flow velocity. To avoid diluting the sample, apply it as close to the column inlet as possible. If an airtrap is included in the system, by-pass it during sample application to avoid back-mixing. Calculate the HETP value from the conductivity (or UV) curve as follows:

HETP, in its simplest terms, is expressed as:
HETP = L/N
where,
L = Bed height (cm)
N = Number of theoretical plates.

N is defined by the equation:
N = 5.54 (Ve /Wh)2
where,
Ve = Elution volume (ml)
Wh = Peak width at half height (ml)
Ve is measured as the volume passed through the column to the peak maximum.
Wh is measured as the peak width at half-peak height.
From the example in the figure, the HETP value can be calculated from the chromatogram as follows:

Ve (ml) Wh (ml) N N/m HETP cm
Acetone 18800 900 2417 4203 0.024


Well-packed columns have low HETP values. However, it is only possible to compare columns that have been packed with the same type of media and that have been tested under identical conditions.

As a general rule-of-thumb, a good HETP value is approximately two to four times the mean bead diameter of the medium in question, provided that the sample does not interact with the medium.

In practice, the correlation between HETP and column performance can only be assessed by the column operator. Once this has been established, a standard can be set to judge the acceptability of a column packing.

For example, the column operator may know from experience that a column packed with Sephadex G-25 gel filtration medium with HETP values above 0.05 cm does not give the required separation. Consequently, the operator will set this value as the maximum permissible i.e. the minimum acceptable quality.

Reduced plate number
Definition of reduced plate number: h = HETP/dp
h = reduced plate number.
HETP = above described height equivalent to a theoretical plate.
dp = mean particle diameter of the chromatography medium beads.

The reduced plate number should be in the range of 2-4 times the mean particle diameter of the chromatography medium beads.

Peak asymmetry factor calculation
The peak asymmetry factor should be as close as possible to 1, and the shape of the peak should be as symmetrical as possible. This is usually the case for gel filtration media, but for certain ion exchange and affinity media, the shape may be asymmetrical due to interaction with the media. A change in peak shape is usually the first indication of column deterioration.

The peak asymmetry factor, As, is calculated from the graph above:
As = b/a
where,
a = distance from peak apex to 10% of the peak height on the ascending side of the peak
b = distance from peak apex to 10% of the peak height on the descending part of the peak

Note: Measuring HETP, h and As values is the best way to judge the condition of the packed column. A packed column can look good, but still need repacking for optimal performance. Always check the column after packing and regularly between runs to ensure best performance.



Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Two step packing using constant flow in both steps. 

Packing buffer: water

For more information concerning the packing procedure, please read Packing Sepharose Fast Flow and Sepharose XL.

Packing Sepharose Fast Flow and XL media

1. Equilibrate all material to the temperature at which the chromatography will be performed.

2. Assemble the column (and packing reservoir if necessary).

3. Remove air from the column dead spaces by flushing the end-piece and adapter with water. Make sure no air has been trapped under the column bed support. Close the column outlet leaving the bed support covered with water.

4. Resuspend the medium and pour the slurry into the column in a single continuous motion. Pouring the slurry down a glass rod held against the column wall will minimize the introduction of air bubbles.

5. If using a packing reservoir, immediately fill the remainder of the column and reservoir with packing buffer. Mount the adapter or lid of the packing reservoir and connect the column to a pump. Avoid trapping air bubbles under the adapter or in the inlet tubing.

6. Open the bottom outlet of the column and set the pump to run at the desired flow rate. Ideally, Sepharose Fast Flow media are packed in XK columns in a two-step procedure: Do not exceed 0.5 bar (0.05 MPa) in the first step and 1.5 bar (0.15 MPa) in the second step. If the packing equipment does not include a pressure gauge, use a packing flow rate of 100 cm/h (XK 16/20 column) in the first step, and 600 cm/h (XK 16/20 column) in the second step. If the recommended pressure or flow rate cannot be obtained, use the maximum flow rate the pump can deliver. This should also give a well-packed bed. 
Note: Do not exceed 75% of the packing flow velocity in subsequent chromatographic procedures using the same pump.

7. Maintain the packing flow for at least 3 bed volumes after a constant bed height is reached. Mark the bed height on the column.

8. Stop the pump and close the column outlet.

9. If using a packing reservoir, disconnect the reservoir and fit the adapter to the column.

10. With the adapter inlet disconnected, push the adapter down into the column until it reaches the mark, allowing the packing solution to flush the adapter inlet. Lock the adapter in position.

11. Connect the column to a pump or a chromatography system and start equilibration. Reposition the adapter if necessary.

Equilibration
Before starting a run, make sure that the bed has reached equilibrium. This is done by pumping starting buffer through the column until the conductivity and/or pH of the effluent is the same as for that of the in-going solution. The column is now equilibrated and ready for use.


Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

20 µm net
Stainless steel or plastic

For code no and detailed information see the spare part section.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Packing small-scale AxiChrom columns

AxiChrom™ 50, 70, 100, 140, and 200 columns are designed to simplify column handling and maintenance during process development.

  • Preprogrammed, verified packing methods that include Capto™, MabSelect™ media
  • Reproducible, predictable and scalable results.
  • Rotating pivot stand for easy emptying, maintenance and safer operation
  • Proven sanitary design with construction materials recognized for use in GMP environments.
  • Operates with ÄKTAexplorer™ and ÄKTApilot™ with UNICORN™ v 5.01 (and later).

AxiChrom columns are designed to simplify column handling and maintenance. Packing is facilitated by the Intelligent Packing methodology in the UNICORN software. The preprogrammed, verified packing methods shorten start-up time, eliminate errors and pack the column with optimal bed compression.

The pivot design allows the column tube to be tilted for convenient manual emptying as well as easy access to parts for maintenance.

Scalable and predictable performance over the entire AxiChrom platform is assured through the use of computational fluid dynamics in designing the liquid distribution system.

The AxiChrom column is sanitizable, and materials of construction are suitable for use in regulated environments. Comprehensive documentation, including IQ/OQ documents are available for all column sizes.

Packing large-scale AxiChrom columns

Large-scale AxiChrom™ columns are supported with Intelligent Packing and intuitive handling via the AxiChrom Master.The novel, swing-out-design facilitates maintenance and provides a safer working enivironment.

  • Preprogrammed verified packing methods that include Capto™ and MabSelect™ media
  • Reproducible and predictable results
  • Swing-out, hoist-free design for easy maintenance and safer operation
  • Proven, sanitary design meets stringent cGMP production standards
  • Key steps guided by AxiChrom Master

Intelligent Packing provides preprogrammed, verified axial compression packing methods that promote accurate and reproducible results. Intuitive handling supports preparation and maintenance procedures. Both are controlled by the AxiChrom Master, a separate unit with a touch-screen interface that guides the operator through critical steps for priming, packing, unpacking, and maintenance.

Packing large-scale AxiChrom columns (Video below)

Easy access to all column parts for maintenance in-situ is facilitated by the swing-out, hoist-free, column tube design.

The distribution system design has been developed using analytical methods and modern computational fluid dynamic modeling tools to ensure reproducible and predictable results over the entire range of scales.

Materials of construction meet the ASME Bioprocessing Equipment Standard for biopharmaceutical manufacture.
AxiChrom columns are supplied with comprehensive documentation to facilitate set-up, installation, and operation.

Something went wrong - we were not able to load your agreed MSA pricing. Please try refreshing the page.

Contact