Fraction collector F9-R
Issues related to Fraction collector F9-R
Possible cause | Suggested remedy |
---|---|
Too many fraction collector instructions have been sent |
Wait for a while and try again. |
The fraction volume is too small |
Collect larger fractions. |
The flow rate is too high |
Reduce the flow rate. |
Possible cause | Suggested remedy |
---|---|
The fraction collector movement is blocked |
Make sure that the fraction collector can move and is free from obstructions. |
Possible cause | Suggested remedy |
---|---|
The tubing is not correctly mounted in the tubing holder nut |
Check that the tubing is correctly mounted, see Section 4.12 Install Fraction collector F9-R, on page 186 in the system handbook. |
Unclean drop sync sensor |
Clean the drop sensor photocell located above the tube sensor with a damp cloth. |
Air in the flow path |
Check the flow path for air. Fill system and purge pumps. If this error is recurrent, generate a System error report and contact service. |
To high flow rate for drop sync |
Decrease the flow rate or disable drop sync. |
Possible cause | Suggested remedy |
---|---|
Hardware error |
Switch off the instrument and wait until the temperature has decreased. Restart the instrument. If this error is recurrent, generate a System error report and contact service. |
Hot surroundings |
Decrease the room temperature. Maximum operating temperature is 35 C. |
Possible cause | Suggested remedy |
---|---|
Wrong node ID |
Check the Fraction collector's Node ID. If necessary, change the Node ID. See Section 4.12 Install Fraction collector F9-R, on page 186 in the system handbook. |
The cable between the Fraction collector and the ICU is not connected |
Make sure that the cable is connected. |