Troubleshooting
Find solutions to product related issues. For unlisted issues please contact local Cytiva service representation.
Superose 6 Increase 3.2/300
Possible cause | Suggested remedy |
---|---|
- |
Confirm that the column is the cause (see below). If so, clean it according to the procedure described in section Cleaning-in-place (CIP) in the Instructions. To confirm that the high back-pressure in the system is caused by the column, disconnect one piece of equipment at a time (starting at the fraction collector) with the pumps running. Check the pressure reading after each piece has been disconnected to determine the source of the back-pressure. |
Possible cause | Suggested remedy |
---|---|
- |
Note that small amounts of air will normally not affect the performance of the column. Run 3 to 4 CV well de-gassed eluent in an upflow direction at a flow rate of 0.04 ml/min at room temperature. |
Possible cause | Suggested remedy |
---|---|
- |
Carefully tighten the black end cap in both ends of the column by hand. |
Possible cause | Suggested remedy |
---|---|
- |
Do not use the column in a system if the black end caps are loose. This is because the liquid could leak into the space between the inner column and the outer shell. Carefully tighten the black end caps by hand. |