故障排查

寻找产品相关问题的解决方案。对于未列出的问题,请联系当地的Cytiva服务代表。

Back pressure increases during operation

Unsatisfactory elution

Poor product recovery

Pressure curve issue

Worrying peak shape

Poor reproducibility

Columns Column leakage

General advice to achieve good performance

Before using the column make sure that:

  • Correct system has been selected in UNICORN System Control
  • Correct wavelength has been set for UV/UPC monitor
  • All tubing has been properly connected and tubing is not longer than needed
  • All connectors are free from leakage, verified by passing a leakage test
  • No tubing is folded or twisted
  • Online filter, if used, is changed on a regular basis
  • Correct buffers are used for the chosen columns and proteins
  • All inlet tubing has been immersed in correct buffer solutions
  • Enough buffer has been prepared
  • Buffers have been equilibrated to the environment temperature
  • Buffers/eluents have been degassed if necessary (e.g., in RPC runs)
  • Suitable columns have been selected for the target proteins
  • Column meets pressure requirements for selected medium
  • Columns have been cleaned and prepared according to column instructions
  • Samples have been clarified by centrifugation and/or filtration prior to sample loading
  • Samples have been adjusted to binding buffer conditions
  • Auto sampler (if used) has been prepared according to user manual
  • The fraction collector has been filled with appropriate number of microtiter plates or tubes
  • Appropriate arrangement for waste handling has been prepared

 

 

Unusual column appearance